Varian Medical Systems, Inc: FDA-registered device maker in Palo Alto, CA, United States, 37 510(k) clearances

Varian Medical Systems, Inc is registered with the FDA (registration 2916710) in Palo Alto, CA, United States. It lists 5 device types. The FDA files 37 510(k) clearances under its listings, the newest dated January 21, 2026. 123 recall events name it as the recalling firm. Registered through 2027. FDA data exported November 3, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
2916710
Address
911 Hansen Way, Palo Alto, CA, United States, 94304
Owner or operator
Varian Medical Systems, Inc (2914292)
Roles
Complaint file establishment, Manufacturer, Specification developer
First device listed
April 26, 2000
Registered through
2027
Brand names listed
4D Integrated Treatment Console v13 MR4, 4D Treatment Console V.13, Acuity with Conebeam Computed Tomography, Anchored Beacon Transponder, Beacon Care Package - Soft Tissue (17G), Calypso Permanent Beacon Transponders, Calypso Prostate Beacon Transponders, Calypso Soft Tissue Beacon Transponders and more

Devices it lists

DeviceProduct codeClassRoleSource
Accelerator, Linear, MedicalIYE2Complaint file establishment, Manufacturer, Specification developeraccessdata.fda.gov
Instrument, Quality-Assurance, RadiologicLHO1Manufactureraccessdata.fda.gov
System, Planning, Radiation Therapy TreatmentMUJ2Complaint file establishment, Manufacturer, Specification developeraccessdata.fda.gov
System, Radiation Therapy, Charged-Particle, MedicalLHN2Manufactureraccessdata.fda.gov
System, Simulation, Radiation TherapyKPQ2Complaint file establishmentaccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
January 21, 2026K252977Halcyon, Ethos Radiotherapy System (5.0)IYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
October 7, 2025K252948TrueBeam, TrueBeam STX, Edge and VitalBeamIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
May 16, 2025K250099Mobius3D (4.1)IYESubstantially EquivalentVarian Medical Systemsaccessdata.fda.gov
December 21, 2023K232870TrueBeam, TrueBeam STx, EDGE and VitalBeam (4.1)IYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
November 3, 2023K232113Halcyon, Ethos Radiotherapy SystemIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
September 12, 2023K231317TrueBeam, TrueBeam STx, Edge, VitalBeamIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
May 22, 2023K223839TrueBeam, TrueBeam STx, Edge, VitalBeamIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
January 17, 2023K222941Halcyon, Ethos Radiotherapy SystemIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
June 3, 2022K213977TrueBeam, TrueBeam STx, Edge, VitalBeamIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
February 12, 2021K2037664D Integrated Treatment ConsoleMUJSubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
January 14, 2021K203669Mobius3DIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
February 10, 2020K192377Ethos Treatment Management, Ethos Treatment Planning, Ethos Radiotherapy System, HalcyonIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
June 28, 2018K181404Varian High Energy Linear Accelerator, UNIQUE, Varian Multileaf CollimatorIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
April 6, 2018K170570Anchored Beacon(R) transponder, Anchored transponder, Beacon Care Package - LungIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
April 2, 2018K180586Varian Head FrameIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
July 28, 2017K172013VitalBeamIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
July 12, 2017K171733TrueBeam, TrueBeam STx, EdgeIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
January 19, 2017K162472TrueBeam-TrueBeam STx-EdgeIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
December 12, 2016K162476Varian High Energy Linear AcceleratorIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
November 25, 2015K152761Permanent Beacon Transponder, Soft Tissue Beacon Transponder, Beacon Care Package - Soft Tissue (17G)IYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
October 26, 2015K151882Varian Treatment (VTx)IYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
May 8, 2015K150706VitalBeamIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
December 19, 2014K143224TrueBeam-TrueBeam STx-EdgeIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
December 11, 2014K142560Varian Head FrameIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
October 2, 2014K142268Varian Verification System (VVS)IYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
August 27, 2014K141230DOSELAB PROIYESubstantially EquivalentMobius Medical Systems, LPaccessdata.fda.gov
January 31, 2014K133240VARIAN MULTILEAF COLLIMATORIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
January 27, 2014K1333314D INTEGRATED TREATMENT CONSOLEMUJSubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
October 11, 2013K131807HIGH ENERGY LINEAR ACCELERATORIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
February 21, 2013K124002VERIFICATION CONSOLELHNSubstantially EquivalentVarian Medical Systemsaccessdata.fda.gov
January 15, 2013K123788INTEGRATED COLLIMATOR VERIFICATION AND INTERLOCK SYSTEM (ICVI)IYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
December 5, 2012K123137CALYPSO SYSTEMIYESubstantially EquivalentVarian Medical Systemsaccessdata.fda.gov
February 16, 2012K112841SURFACE BEACON TRANSPONDERSIYESubstantially EquivalentCalypso Medical Technologies, Inc.accessdata.fda.gov
October 15, 2010K102024RPM RESPIRATORY GATING SYSTEMIYESubstantially EquivalentVarian Medical Systemsaccessdata.fda.gov
August 15, 2007K071992HIGH-DEFINITION 120 MULTILEAF COLLIMATOR (HD120 MLC)IYESubstantially EquivalentVarian Medical Systemsaccessdata.fda.gov
December 30, 2003K033339ACUITY WITH CONEBEAM COMPUTED TOMOGRAPHYKPQSubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov
December 23, 2003K033343TRILOGY RADIOTHERAPY DELIVERY SYSTEMIYESubstantially EquivalentVarian Medical Systems, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI. Newest 100 of 123.

InitiatedEventClassStatusProductReasonSource
August 12, 201985269Class IITerminatedVarian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High Definition 120 Multileaf Collimator Model Number(s): product code: H54 and H56; MLC software version 8.5 - Product Usage: is inteAfter a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary- secondary interlock due to carriage fault on the MLC that immediately preceded the initiation of the arc treatment. The issue will occur only witaccessdata.fda.gov
December 26, 201881912Class IITerminatedEclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant oThere is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples theaccessdata.fda.gov
December 2, 201676018Class IITerminatedEclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotheraModifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.accessdata.fda.gov
September 13, 201675296Class IITerminatedVarian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head Frame System is for use with a computed tomography scanner to perform imaging for treatment planning and Varian Medical Systems has received a report that a user was able to easily rotate the head frame posts when attached to the mounting cam on the metal head ring. There was no report of serious injury due to this issueaccessdata.fda.gov
June 23, 201674585Class IITerminatedOptical Surface Monitoring System{OSMS, Varian Private Label} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed trVarian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch rotation while performing patient alignment with Optical Surface Monitoring System [OSMS].accessdata.fda.gov
January 11, 201673092Class IITerminatedEclipse Treatment Planning System versions 11, 13.0, 13.5 and 13.6; Model number: H48; Product Usage: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignanWhen using PBC 11.0.31 to calculate the dose for a conventional arc field with more than 100 segments for Eclipse versions 11.0, 13.0, 13.5 or 13.6, the displayed dose does not correspond to the calculated Monitor Units (MU). Potential for unintended radiation exposure.accessdata.fda.gov
September 28, 201572374Class IITerminatedARIA Radiation Oncology, versions 10, 11, 13.0 and 13.5 with Clinical Assessment License. Model number HIT Product Usage: ARIA Radiation Oncology is a radiation treatment plan and image management application.An anomaly was identified with the ARIA for Radiation Oncology software with a Clinical Assessment license. Admin Instructions may not print on drug order prescriptions or be transmitted with e-Rx.accessdata.fda.gov
August 18, 201469126Class IITerminated4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatment delivery by monitoring linear accelerator parameters and preventing the radiation therapy device from commenciDue to user error, the 4DITC can improperly allow users to clear a safety interlock imposed by the 4DITC on behalf of Varian's Bar Code Conical Verification system from the BrainLab's ExacTract console when it is used with BCCV.accessdata.fda.gov
June 17, 201468627Class IITerminatedC-Series High Energy Linear Accelerator - Models: Novalis Tx, Trilogy, Trilogy Tx, Clinac iX, Clinac 2100 C/D, Clinac 2300C/D, Clinac 21EX, Clinac 23EX with versions 7, 8, 9 and 6 MV configuration and maximum dose rate oTrend in reports of unexpected decrease in beam output in C-series High Energy Linear Accelerators for 6MV photon treatment mode.accessdata.fda.gov
April 30, 201469637Class IITerminatedVarian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.accessdata.fda.gov
March 20, 201467896Class IITerminatedOn-Board Imager - Exact Arm Elbow Motor Shaft (kVD, kVS, and MV) with ''Type 01" elbow motors. Product Usage: The On-Board Imager device is an x-ray device used for evaluation of correct patient position in relation to iVarian has identified a failure with the shaft of the Type 01 elbow motor for Exact arms used in the On-Board Imager (OBI) device due to belt over-tensioning.accessdata.fda.gov
November 18, 201366956Class IITerminatedPlan Parameters workspace within Eclipse Treatment Planning System Version 11 and 13; ARIA Radiation Oncology versions 11 and 13 Acuity, versions 11 and 13. Manufactured by Varian Medical Systems Palo Alto, CA. Used in rAn anomaly has been traced to a software deficiency within the Plan Parameters workspace. When a dose relevant parameter is changed and "do no clear MUs and Reference Point dose" is selected, the Monitor units may change.accessdata.fda.gov
October 24, 201366764Class IITerminatedC-Series: Clinac, Trilogy, Trilogy Tx., Novalis high energy linear accelerators and UNIQUE single energy linear accelerator; Versions 7, 8 and 9. The UNIQUE is not sold in the US. Product Usage: The Varian High Energy LiThis correction is to notify users that a solution to a previous correction has been developed and Varians reps will be contacting locations to schedule installation.accessdata.fda.gov
March 26, 201364828Class IITerminatedVarian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage: The RPM device is to be used to characterize the patients respiratory motion information to synchronize theirVarian has identified an anomaly with the RPM Respiratory Gating System version I.7.5 with 3D Option. While operating with the 6-dot marker block, this version of RPM Gating may produce an inverted respiratory waveform in amplitude and phase-gating sessions.accessdata.fda.gov
January 28, 201364319Class IITerminatedVarian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.accessdata.fda.gov
November 5, 201263713Class IITerminatedVarian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Affected Versions: ARIA version 11.0 below 11.0.55, Build 11.0.28 and 11.0.34, RefeVarian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality where a 360 degree arc field may convert to static field during DICOM export or import.accessdata.fda.gov
September 24, 201263471Class IITerminatedARIA Radiation Oncology, Version 11. Varian Medical Systems, Palo Alto, CA. The ARIA Radiation Oncology product is a treatment plan and image management application.An anomaly has been identified with the ARIA Oncology Information System Version 11 where Monitor Units of one fraction may be incorrect, when partial treatment is being continued.accessdata.fda.gov
August 17, 201263093Class IITerminatedVarian brand Eclipse Treatment Planning System, Model Number: H48; Eclipse 8.9 and 10.0 Reference/FSCA Identifier: CP-08922; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The EclipVarian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.accessdata.fda.gov
August 14, 201263033Class IITerminatedVarian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-daccessdata.fda.gov
July 2, 201262562Class IITerminatedVarian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., PalA software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.accessdata.fda.gov
June 18, 201262420Class IITerminatedVarian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (EcliVarian has identified an anomaly with the Eclipse Treatment Planning (for radiotherapy treatments) where changing the prescription in a calculated VMAT or RapidArc plan may lead to Monitor Units which do not reflect the new prescription. If the dose prescription of a VMAT or RapidArc plan is changed after the initial Dose Volume Optimization, the Monitor Uniaccessdata.fda.gov
April 9, 201261694Not yet classifiedTerminatedTrueBeam and TrueBeam STx V1.0, 1.5, 1.6.95 and below. TrueBeam Radiotherapy Delivery System is intended to provide stereotactic radiosurgery and precision radiotherapy anywhere in the body where radiation treatment is iVarian received a report involving a Gantry collision and is sending a notification to remind users of the collision protection tools available with TrueBeam and TrueBeam STx, including actions the operator should be aware of.accessdata.fda.gov
April 9, 201261693Not yet classifiedTerminatedTrue Beam; Versions 1.0, 1.5.12.0 and below, 1.6.9.5 and below. Varian Medical Systems, Palo Alto, CA. Product Usage: TrueBeam Radiotherapy delivery system is intended to provide sterotactic radiosurgery and precision raA software anomaly has been identified in TrueBeam when using respiratory gating with patients in the prone position. The gating curve is inverted when a prone patient plan is loaded into TrueBeam.accessdata.fda.gov
January 30, 201261198Not yet classifiedTerminatedEclipse Treatment Planning System, versions 7.3, 8.0, 8.1, 8.2, 8.6, 8.9 and 10.0; Model H48 Varian Medical Systems, Palo Alto, CA. Product Usage: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radioVarian has been identified an anomaly with the Eclipse Treatment Planning System where the calculated dose distribution is not updated or invalidated when a recalculation or reload is performed while the compensator or fluence editor is open.accessdata.fda.gov
January 30, 201261196Not yet classifiedTerminatedTrueBeam and True Beam STx versions 1.0 through 1.5, Model number H19; Varian Medical Systems, Palo, Alto, CA 94304. TrueBeam Radiotherapy Delivery System is intended to provide stereotactic radiosurgery and precision raAn anomaly has been identified with the respiratory gating software of TrueBeam. When importing breath-hold gating protocols, the gating thresholds can be reset to default values, rather than retaining the thresholds established during planning.accessdata.fda.gov
January 30, 201261193Not yet classifiedTerminatedFastPlan Treatment Planning System, Model HZ3, Varian Medical Systems, Palo Alto, CA. Varian FastPlan Treatment Planning System is a radiation treatment planning system composed of a high resolution UNIX graphics computeAn anomaly has been identified with the CT Localization module of the FastPlan treatment planning system. If the module reports a significant difference in tilt or spin on a slice and the use elects to "Discard" that slice, an incorrect calculation of the isocenter location may result.accessdata.fda.gov
January 16, 201261074Not yet classifiedTerminatedVarian brand Varian 52-leaf Standard MLC multileaf collimator; Reference/FSCA Identifier: CP-04792; Model Number(s): All models, including H52, H54, H56, HHM; Product is manufactured and distributed by Varian Medical SysWhen using the Standard and Millennium 52-leaf MLC multi-leaf collimators for beam shaping, under specific treatment conditions, dose may be delivered outside of the intended treatment field that exceed the expected transmission dose.accessdata.fda.gov
December 16, 201160830Not yet classifiedTerminatedVarian brand ARIA Radiation Oncology, ARIA versions 10.0.39 and below, Reference/FSCA Identifier: CP-06070, Model Number: HIT, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA ProductVarian received customer complaints involving ARIA Oncology Information Systems, where the Planned Total Dose did not calculate correctly in Physicians Intent. It was identified that incorrect values were being stored for Planned No. Fractions, leading to an unexpected calculation.accessdata.fda.gov
December 12, 201160837Not yet classifiedTerminatedVarian brand Eclipse Treatment Planning System, 7.3, 8.0, 8.1, 8.2, 8.6, 8.9, 10.0; Model Number: H48; Reference/FSCA Identifier: CP-07007, Product is manufactured and distributed by Varian Medical Systems Inc., Palo AltAn anomaly has been identified with the Eclipse Treatment Planning System when planning for the Siemens Multileaf Collimator (MLC) where a virtual jaw can be positioned inside the MLC aperture and then used as the beam limiting device for the dose calculation. This will result in an incorrect dose distribution within Eclipse.accessdata.fda.gov
December 12, 201160829Not yet classifiedTerminated4D Integrated Treatment Console (4DITC) TrueBeam Linear Accelerator Varian Medical Systems, Palo Alto, CA. The 4DITC is designed to interface to the Varian Clinac Treatment Control System for the purpose of providing verVarian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should not be added according to the system specifications.accessdata.fda.gov
December 12, 201160826Not yet classifiedTerminatedVarian brand 4D Integrated Treatment Console (4DITC) v10.2.3, Reference/FSCA Identifier: CP-06873, Model Number: H51, distributed by and manufactured by: Varian Medical Systems Inc., Palo Alto, CA The 4D Integrated TreatWhen multiple conical collimator treatment plans are scheduled to be treated in a session, the Plan Label should be displayed on the BCCV dialog box should match the Plan Label on the 4DITC. However, the name of the plan is not being updated correctly.accessdata.fda.gov
November 28, 201160659Not yet classifiedTerminatedVarian Unique Single Energy Linear Accelerator, C-Series Clinac or Trilogy, versions 7.x and 8.x, Reference/FSCA Identifier: CP-0661; Model Numbers: H14, H18, H27, H29; Product is manufactured and distributed by Varian MAn event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.accessdata.fda.gov
November 23, 201160684Not yet classifiedTerminatedPortal Vision, R-Arm with 4D Integrated Treatment Console; Varian Medical Systems. Palo Alto, CA. The Portal Vision device is used for verification of the treatment field and shielding block in relation to anatomical lanAn anomaly was discovered when using the 4DITC in combination with an R-Arm to acquire double exposure MV images.accessdata.fda.gov
November 10, 201161012Not yet classifiedTerminatedVarian brand Varian Clinac Accelerators with Mark, Millennium MLC 120 (Multileaf Collimator) and BrainLAB m3 MCL, Reference/FSCA CP-05591, Model Number: H53, Product is manufactured and distributed by Varian Medical SystFailure to retract or park the Varian MLC leaves prior to use of the demountable BrainLAB m3 could result in the Varian MLC leaves interfering with the treatment field defined by the BrainLAB m3 without an interlock being asserted.accessdata.fda.gov
October 31, 201160441Not yet classifiedTerminatedVarian brand Eclipse Release 10 with RT Chart Release 10 (Client Build 10.0.28 or Build 10.0.34), An Eclipse Treatment Planning System used with ARIA Radiation Oncology (RT Chart module), Reference/FSCA Identifier: CP-05An anomaly has been identified with Varian Eclipse and RT Chart where adding a reference point in RT Chart and editing the field dose contributions may change the Prescribed Dose Per Fraction in Eclipse.accessdata.fda.gov
October 31, 201160411Not yet classifiedTerminatedTrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiAn anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.accessdata.fda.gov
October 6, 201160231Not yet classifiedTerminatedVarian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i; Ref: BT-00264, Model 1-H63; Product is distributed by Varian Medical Systems Inc., Palo Alto, CA and ManufactuThe Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.accessdata.fda.gov
October 4, 201160238Not yet classifiedTerminatedHead Ring Posts with part number - re-usable components of the Frame Array Module (of the Optical Guidance Platform and Floorstand devices. Varian Medical Systems, Palo Alto, CA 94304. The Optical Guidance Platform is foAn anomaly has been identified with the Head Ring posts used by both the Optical Guidance Platform FrameArray module and the Floorstand where the head ring posts may be damaged due to excessive mechanical stress resulting in possible failure during usage.accessdata.fda.gov
September 9, 201160022Not yet classifiedTerminatedVarian Clinac, Trilogy, Trilogy Tx and Novalis linear accelerators. Manufactured by Varian Medical System, Palo Alto, CA. Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radVarian has received reports in which a user has remotely rotated the gantry into contact with the couch or with the patient, in both the manual mode and the automate mode.accessdata.fda.gov
July 21, 201159586Not yet classifiedTerminated4D Integrated Treatment Console v 8.6 with ARIA RT Chart v8.5, all of 4DITC v 8.3; Varian Medical Systems, Palo Alto, CA Designed to assist the operator of a radiation therapy device in providing the radiation therapy deThe 4D Integrated Treatment Console may erratically fail to load the planned physical wedge accessory without warning or notification when receiving an approved plan from the oncology information system.accessdata.fda.gov
July 20, 201159572Not yet classifiedTerminatedVarian Treatment version 6.6.5052. Varian Medical Systems, Palo Alto, Ca Intended to provide accurate treatment set-ups for each patient by monitoring linear accelerator set-up parameters and by preventing the radiation Potential for Varian Treatment application failing to load the physical wedge filter for a treatment field on a Siemens Mevatron linear accelerator. Delivery of a treatment field without the planned wedge would deliver more dose than intended and would change the shape of the dose distribution for that field.accessdata.fda.gov
July 13, 201159404Not yet classifiedTerminatedTitanium Fletcher-style Applicator Set - Defined Geometry - CT Compatible with Titanium intrauterine probe with stopper length 40 mm, 30 degree angle. Varian Medical Systems, Palo Alto, CA; Manufactured by Varian MedicalPotential of weld failure of the Titanium intrauterine probe at the cervical stopper.accessdata.fda.gov
July 13, 201159402Not yet classifiedTerminatedFlexible Probe with Blocking Washer (part of Segmented Cylinder Set) The Segmented Cylinder Applicator Set was developed to treat cancer of the vagina, vaginal stump and rectum.Blocking washers on the Flexible Probe with Blocking washers may shift and cause potential for treatment length error.accessdata.fda.gov
July 13, 201159401Not yet classifiedTerminatedGammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 113 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany Product Usage: ThNeedle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal.accessdata.fda.gov
July 13, 201159388Not yet classifiedTerminatedGammaMed Flexible Applicator probe; 3.2mm diameter, braced PVDF. Varian Medical Systems, Palo Alto, CA. Mfg by Varian Medical Systems, Haan GmbH. Flexible Applicator Robes are designed for brachytherapy intracavitary treCertain GammaMed Flexible applicator probes may not have been heat annealed during manufacturing to prevent shrinkage after autoclave.accessdata.fda.gov
July 7, 201159394Not yet classifiedTerminatedEclipse Treatment Planning System, 4D Structure Merge Tool Option in Eclipse Release 8.5 -10 (Client Build 8.2.2x through 10.0.34) with 4D Planning; Model Number: H48; Reference/FSCA Identifier: CP-04931; Product is manuAn anomaly has been identified in the 4D Structure Merge option of certain versions of Varian's Eclipse Treatment Planning System. The treatment field energy may change from the current selection to a previous selection without user intervention and without invalidating the MU or calculated dose distribution.accessdata.fda.gov
June 30, 201159378Not yet classifiedTerminatedEclipse Treatment Planning System; Varian Medical Systems, Palo Alto, CA Product Usage: Eclipse Treatment Planning System is used to plan radiotherapy treatments for patients with malignant or benign diseases. It is usedIn affected Eclipse versions, the treatment field energy may change from the current selection to a previous selection without use intervention and without invalidating the MU or calculated dose distribution.accessdata.fda.gov
June 27, 201159351Not yet classifiedTerminatedVarian brand 4D Integrated Treatment Console (4DITC) Radiation Therapy, Interpretation of a RapidArc Treatment Plan as a 3D Conformal Arc Treatment Plan; version v8.6 or 8.8 in combination with treatment planning systemsIn certain configurations, the 4DITC can interpret a RapidArc or VMAT treatment field as 3D Conformal Arc Treatment Plan. If a RapidArc or VMAT treatment field is incorrectly interpreted as Conformal Arc, the field will be delivered with a constant dose rate and gantry speed for the entire duration of the delivery, which can lead to delivered dose being diffaccessdata.fda.gov
June 23, 201159350Not yet classifiedTerminatedVarian brand 4D Integrated Treatment Console (4DITC) Radiation Therapy, version 8.1 SP2 to 10.0 SP1; Model Number: H51; Reference/PSCA Identifier: CP-03940; Product is manufactured and distributed by Varian Medical SysteVarian has scheduled for an upgrade of the device from current version to 10.2. There are circumstances in which the 4DITC may be involved in a failure to correctly record treatment history information.accessdata.fda.gov
June 21, 201159262Not yet classifiedTerminatedOn-Board Imager (OBI) 1.3, 1.4, 1.5; and Trilogy Mx, TrueBeam 1.0, Offline Review 1.0 - 2.0; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The On-Board Imager device is used for veVarian has identified a software anomaly in the 3D Match environment, regarding a slight mismatch between CT image and Structures in On-Board Imager w/s.accessdata.fda.gov
June 6, 201159226Not yet classifiedTerminatedHigh-Definition 120 Multileaf Collimator HD 120 MLC is an accessory x-ray collimator designed to be mounted on Varian Trilogy Tx and Trilogy linear accelerators and is intended to shape s-ray field perimeter. Field shapeWhen using the HD120 Multi Leaf Collimator for beam shaping, and under specific treatment conditions, dose may be delivered outside of the intended treatment field that exceeds the expected transmission dose.accessdata.fda.gov
May 31, 201164997Class IITerminatedVarian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Varian High Energy Accelerator is intended to provide stereotactic radiosurgery and pUnder certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slaccessdata.fda.gov
May 26, 201159087Not yet classifiedTerminatedClinac, Trilogy. Trilogy Tx and Novalis Tx Linear Accelerators. Varian Medical Systems, Inc. Oncology Systems Intended for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions in the boOn Varian Linear Accelerators where the customer may calibrate the collimator angle position readout in reverse, switching the 90 degree and 270 degree positions.accessdata.fda.gov
May 24, 201159088Not yet classifiedTerminatedClinac Linear Accelerators; Varian Medical Systems, Palo Alto, CA. Radiation Therapy intended to deliver megavoltage x-ray treatments for conventional radiotherapy and stereotactic radiosurgery and radiotherapy.The throat cover on the High Energy Clinac may detach if not properly installed and possibly strike a patient.accessdata.fda.gov
May 20, 201159006Not yet classifiedTerminatedEclipse Treatment Planning System; Varian Medical Systems, Palo Alto, CA Radiotherapy treatment for patients with malignant and benign diseases.A software anomaly has been identified with the Eclipse Monte Carlo (eMC) algorithm where not-equidistant data points in the open beam profile used during configuration may lead to inaccurate results.accessdata.fda.gov
May 19, 201158994Not yet classifiedTerminatedTrueBeam Linear Accelerators (aka Trilogy Mx) Varian Medical Systems, Palo Alto, CA Product Usage: Intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in Imaging arms of the TrueBeam Accelerator may have loose encoder pulleys that could lead to inaccurate readout of arm geometry.accessdata.fda.gov
May 18, 201158976Not yet classifiedTerminatedClinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anyThe Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.accessdata.fda.gov
May 17, 201158974Not yet classifiedTerminatedARIA, versions 8.0 and 8.1; Varian Medical Systems, Palo Alto, CA. Product Usage: Aria Radiation Oncology product is a treatment and image management application. It enables the authorized user to enter, access, modify, An issue has been detected when using the RT Chart function in ARIA versions 8.0 and 8.1. When users select the "Convert to Setup Field" option, a Set-up field is created, that removes the DMLC plan components in order to create an open collimator field for setup purposes.accessdata.fda.gov
May 13, 201158970Not yet classifiedTerminatedClinac iX, 23EX, CX, DHXC, Novalis Tx, Trilogy and TrueBeam Linear Accelerators Drive Stands Indicated for sterostatic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when Some drive stand bolts may fail to meet mechanical specifications.accessdata.fda.gov
May 13, 201158969Not yet classifiedTerminatedEclipse Treatment Planning System; Release 6.5 through 10 (build 7.3.10 through 10.0.34) Varian Medical Systems, Inc. Used to plan radiotherapy treatments for patients with Malignant or benign diseases. Plans beam irradiIn affected Eclipse versions, user editing of the MLC Leaf positions may not automatically invalidate calculated dose distributions.accessdata.fda.gov
May 3, 201158836Not yet classifiedTerminatedEclipse Radiation Treatment Planning System, model H48; Varian Medical Systems Inc..In affected Analytical Anisotropy Algorithm versions when modeling the second source, the collimator jaw positions are read from the first control point of the field. This can cause the delivered dose to be higher than planned locally by up to 10%.accessdata.fda.gov
April 5, 201158590Not yet classifiedTerminatedVarian Medical System TrueBeam system for stereotactic radiosurgery and radiotherapy. Model number H19Position sensor failure mode may result in an inaccurate position calculation. 1. The video returned by one of the two cameras inside the Spectra is all white or all black, and the Spectra stops tracking. Or 2. Certain video intensities are not available on one of the two cameras. The image has abnormally abrupt transitions from dark to light portions, withoaccessdata.fda.gov
April 4, 201158581Not yet classifiedTerminatedEclipse Treatment Planning System, Client software versions 8.0 through 8.6; Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System An anomaly in Eclipse Treatment Planning System, versions 8.0 to 8.6, whereby Actual SSD (source surface distance) value may not correspond to the Planned SSD value after Planning Approval.accessdata.fda.gov
March 31, 201158558Not yet classifiedTerminatedVarian brand Clinac and TrueBeam, High Energy Linear Accelerator, Model Numbers: H14, H19, H29; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Indicated for stereotactic radiosurgerThe coolant may leak. The resulting coolant leak presents a risk of electrical shock to any person working within the protective housing.accessdata.fda.gov
March 30, 201158505Not yet classifiedTerminatedOptical Guidance Platform, Version 2.6 and 2.6.1, Model Number: HZl, Manufactured and Distributed by: Varian Medical Systems Inc., Palo Alto, CA. For use with a charged particle accelerator to perform precise positioningA software anomaly has been identified with the Optical Guidance Platform (OGP) Software v2.6 and v2.6.1 where the transfer of datasets from treatment planning systems other than the FastPlan system result in a lateral offset error.accessdata.fda.gov
March 28, 201158438Not yet classifiedTerminatedVarian brand Clinac, Medical Linear Accelerator, All Varian Clinac with Model Numbers: H14, H18, H26, H27, H28, H29, Hcx; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Trilogy Varian has identified an anomaly whereby, following prolonged use, the screw fastener holding the wedge body to the tray may fail.accessdata.fda.gov
March 17, 201158474Not yet classifiedTerminatedVarian Medical Systems Inc., Millennium and HD-120 MultiLeaf Collimators (MLC), a component of the Clinac, All Varian Clinac using the collimator cover P/N 1106900-xx, Model Numbers: H14, H18, H27, H29, HCX; part of the An improperly installed collimator cover may detach and fall from the machine, potentially striking the patient or nearby personnel.accessdata.fda.gov
March 9, 201158424Not yet classifiedTerminatedARIA, RT Chart Application v 8.0 through 8.9 Varian Medical Systems, Palo Alto, CA. Aria Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, mAn issue was identified with the gantry rotation direction when creating opposing field within the RT Chart function of ARIA, or possibly with third party planning systems and delivering the plan at 4DITC.accessdata.fda.gov
March 3, 201158252Not yet classifiedTerminatedOptical Guidance Platform, version 2.6 and 2.6.1; Reference/FSCA Identifier: CP-03976; Model Number: HZ1; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Optical Guidance Platform isThe anomaly that has been identified with the Optical Guidance Platform (OGP) software may not be always displaying the correct transfer date on the patient file.accessdata.fda.gov
March 3, 201158246Not yet classifiedTerminatedOptical Guidance Platform, version 2.6 and 2.6.1; Reference/FSCA Identifier: CP-03899; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Optical Guidance Platform is for use with a chaThe optical guidance platform may not be properly enforcing a 24 hour time limit between optical camera recalibrations.accessdata.fda.gov
February 28, 201158230Not yet classifiedTerminatedThe 4D Integrated Treatment Console; Model Number: H51; Reference: 4DITC versions 8.6 and below; File Reference: CP-2007-0288; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The 4D A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.accessdata.fda.gov
February 24, 201158240Not yet classifiedTerminatedClinac High Energy Medical Linear Accelerator, a Trilogy Radiotherapy Delivery System; Model #s: H14, H18, H27, H29, HCX; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Trilogy The wedge angle labeling on the wedge tray may not correctly match the wedge body. However, the wedge body is labeled with the correct wedge angle.accessdata.fda.gov
February 18, 201158112Not yet classifiedTerminatedARIA versions 8.0 and 8.1 Varian Medical Systems, Palo Alto, CA 94304 The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, stoAn anomaly was identified in the ARIA version 8.0 and 8.1, when an image sequence template is added to a field in RT chart and a CIAO (Completely Irradiated Aperture Opening) is calculated, the resulting CIAO display at 4DITC may be incorrect.accessdata.fda.gov
February 7, 201158072Not yet classifiedTerminatedVarian Medical Systems, Varis Vision - RT Chart version 7.3.10. For radiation therapy.An issue with the Varis Vision software where treatment can occur without the dynamic Multileaf Collimator component and may deliver a higher than expected dose of radiation.accessdata.fda.gov
December 14, 201057527Not yet classifiedTerminated4D Integrated Treatment Console (4DITC); All versions up to and including 8.8.17 Varian Medial Systems, Palo Alto, CA Designed to assist the operator of a radiation device in providing accurate treatment set-ups for eachMultiple isocenter plans will load on the 4DITC and are available for imaging and treatment, but the 4DITC software does not recognize that mulitplex isocenter is present.accessdata.fda.gov
December 10, 201057610Not yet classifiedTerminatedVarian High Energy Linear Accelerator: Varian Medical Systems, Palo Alto, CA 94304 Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiationAn anomaly was identified whereby the bolts used to fasten the counterweight to the gantry may not, in some cases, be fully tightened to the required torque specification.accessdata.fda.gov
July 20, 201056774Not yet classifiedTerminatedVarian brand Trilogy System with RapidArc, Model Numbers: H14, H26, H27, H29; Distributed and Manufactured by: Varian Medical Systems Inc., Palo Alto, CAThe system may deliver an inaccurate dose that does not match the treatment plan. As a result of a system configuration problem, the VMAT (Volumetric Modulated Arc Therapy) plans may be delivered as Dynamic Arc therapy plans. If this occurs, the VMAT plan will not be recognized. This issue affects Trilogy machines with the Rapid Arc Treatment delivery optionaccessdata.fda.gov
June 23, 201056182Not yet classifiedTerminatedVarian Radiotherapy delivery system (C-Series Clinac, Trilogy and Novalis Tx), model numbers H14, H17, H18, H23, H26, H29, HCXPosition errors may result in misdelivery of radiation.accessdata.fda.gov
June 10, 201056347Not yet classifiedTerminatedVarian brand C-series Clinac, Trilogy and Novalis Tx, Software Versions 6.X and 7.X, Model Numbers: H14, H27, H29, HCX, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Trilogy RaThe Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch.accessdata.fda.gov
May 24, 201056346Not yet classifiedTerminatedVarian brand TrueBeam and TrueBeam STx, Model Number:TMX- H19, Distributed by and Manufactured by: Varian Medical Systems Inc., Palo Alto, CADue to a software anomaly, guidance-based couch shift values may not be applied as expected when the operator selects "Apply Shift" and presses Motion Enable buttons. This can result in the actual patient position differing from that position indicated by the imaging application affecting poistion accuracy and precision.accessdata.fda.gov
May 13, 201056012Not yet classifiedTerminatedVarian Medical Systems, Inc. Eclipse/VARiS Vision, an Eclipse radiotherapy Treatment Planning System (Eclipse TPS); Eclipse Version 7.3.10 (multiple workstations) VARiS Vision System 7. x; Model Numbers: H48/H83.Incorrect treatment plan or record: A software design deficiency has the potential for corrupted treatment plan records or treatment records in cases where multiple users editing a record simultaneously can result in a corrupted (but accessible) treatment plan record.accessdata.fda.gov
April 19, 201056349Not yet classifiedTerminatedVarian brand C-Series Clinacs and Acuity, Exact Couch Top, Model Numbers: H69, Product is manufactured and distributed by: Varian Medical Systems Inc., Palo Alto, CAThe product has been shipped since 2005 with the wrong label regarding max weight to be used on accessory rails.accessdata.fda.gov
March 23, 201056024Not yet classifiedTerminatedVarian Medical Systems FastPlan software, Versions 5.5 and 5.5.1, Model Number HZ3, manufactured by Varian Medical Systems, Palo Alto, CA Treatment Planning System is intended as aid in quickly and precisely planning theDue to software Issue, an error occurs when performing cone planning using MRI, which may result in the patient receiving a higher radiation dose than expected, potentially to healthy tissue.accessdata.fda.gov
January 19, 201056350Not yet classifiedTerminatedVarian brand C-Series Ximatron Radiotherapy Treatment Stimulator, Model H72, Product is manufactured and distributed by Varian Oncology Systems, Palo Alto, CA The product is indicated for use in radiation therapy simulatThere is a potential for failure of a lead screw drive system or image intensifier lead screw drive assembly, if normal maintenance inspections are not followed correctly.accessdata.fda.gov
December 9, 200954648Not yet classifiedTerminatedVarian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds 8.2.x). For radiation therapy.Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan.accessdata.fda.gov
December 1, 200956354Not yet classifiedTerminatedVarian brand C Series Clinics (Includes Trilogy, and Novalis Tx), Model Numbers: H14, H26, H27, H29, Distributed by and/or Manufactured by: Varian Medical Systems Inc., Palo Alto, CA The system consists of two major compA circuit design anomaly whereby a faulty balance potentiometer can lead to an undetected electron beam asymmetry.accessdata.fda.gov
October 16, 200954146Not yet classifiedTerminated4D Integrated Treatment Console, versions: 8.0.12, 8.0.23 SP2, 8.1 SP2, 8.1.1.3, 8.1.1.4, 8.1.2, 8.1.3, 8.1.4, 8.3, 8.6, 8.6.1, 8.6.2, 8.8.0; Model Number: H51; Distributed and manufactured by: Varian Medical Systems IncSoftware anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field.accessdata.fda.gov
October 8, 200953778Not yet classifiedTerminatedC-Series Clinac, Trilogy and Novalis Tx used for SRS treatments, Model Numbers H14, H18, H27, H29, HCX, manufactured by Varian Medical Systems, Palo Alto, CA. Device is a radiation Therapy accelerator intended to deliverProduct may deliver radiation treatment to areas larger than intended,to healthy tissue.accessdata.fda.gov
September 15, 200953461Not yet classifiedTerminatedVARiS 6.2, 6.1 and 6.0 when used with LINK, Model numbers H88 and H85. Manufactured by Varian Medical Systems, Palo Alto, CA.Software issue: Under certain circumstances, when a therapy treatment field is transferred to VARiS Vision 6.2, 6.1 and 6.0 via LINK procedures, the rotation direction will be reversed.accessdata.fda.gov
July 17, 200953045Not yet classifiedTerminatedVarian Medical Systems, Eclipse Proton Convolution Superposition DCF, version 8.6.15, Model H48, radiation therapy treatment planning system.Software Issue/Mistreatment Potential-- Due to software issues, dose distribution may be incorrect and unintended radiation doses may be delivered.accessdata.fda.gov
May 29, 200952636Not yet classifiedTerminatedClinac with Version 7.x Software, manufactured by Varian Medical Systems, Palo Alto, CA. Part of the Trilogy Radiotherapy Delivery System for radiation therapy intended to delivery megavoltage x-ray treatments for convenUnexpected Movement: if the stereotactic motion disable function is turned on, the couch can be moved via float mode unexpectedly.accessdata.fda.gov
May 8, 200952654Not yet classifiedTerminatedVaris, Aria Radiation Oncology - Version 8.1.15, 8.5.11, and 8.6.07, Model number HIT-INFSYS, Manufactured by Varian Medical Systems Inc., Palo Alto, CA The Varis RV function is designed to assist the operator of a radiaPotential mistreatment-- due to software issues, retired plans may be used and treated repeatedly.accessdata.fda.gov
January 26, 200951217Not yet classifiedTerminatedVarian Medical Systems Floorstand. An accessory of the Optical Guidance Platform, Model number HZ5 for use with a charged particle accelerator (such as the Clinac 600C).Potential patient impact -- limit switches for Floorstand may be inadequate for some models and allow for potential impact.accessdata.fda.gov
November 21, 200850496Not yet classifiedTerminatedVarian Medical Systems Eclipse treatment planning system, Model Number H48, versions 6.5/7.x and Inspiration 8.0.Possible incorrect wedge orientation in treatment: Software labeling was issued without warning message relating to preservation of user selected values for Siemens wedge orientation, wedge limits and SAD values for Elekta machines, and collimator operating limits for Saturne Machines when the software version is upgraded.accessdata.fda.gov
October 2, 200850143Not yet classifiedTerminatedEclipse 8.5 (build 8.2.2X) systems with distributed Calculation Framework (DCF) 8.2.22, model number H48, manufactured by Varian Medical Systems, Inc., Palo Alto, CA. Indicated for planning proton treatment of neoplasms When using the multiple static segment option, the device may alter dose delivery.accessdata.fda.gov
August 11, 200849399Not yet classifiedTerminatedExact Couch with Exact Couch Top, a component of the C-Series Clinacs, Acuity and Ximatron treatment systems. Model numbers H14, H18, H27, H72, H77. A powered radiation therapy patient support assembly is an electricallyDuring couch movement, a patient's fingers may get pinched at certain points which could result in broken, disengaged or pinched fingers.accessdata.fda.gov
June 11, 200848730Not yet classifiedTerminatedVarian Medical Systems Eclipse device, model H48, Version 6.5 is used to plan photon and electron radiation therapy treatments.Software anomaly in which swapping IDs of planning fields can produce inconsistencies between dose matrix and field parameters.accessdata.fda.gov
June 2, 200848661Not yet classifiedTerminatedVarian RV Software Varian Treatment for Non-Varian Linacs Version 6.6.5042 and 6.6.5043; Model Number: H46; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Designed to assist the opeA malfunction within the Varis software might result in a misadministration (underdose). The software does not correctly interpret the number, causing field parameters having decimal values to be incorrect.accessdata.fda.gov
May 29, 200848654Not yet classifiedTerminatedVarian Medical Systems FastPlan, versions 5.5 and 5.5.1, model number HZ3, manufactured by Varian Medical Systems, Palo Alto, CA The product is a radiation treatment planning system.Due to issues with image orientation verification, a patient may receive all or some of prescribed high dose to healthy tissue instead of intended area.accessdata.fda.gov
May 21, 200848652Not yet classifiedTerminatedVaris Aria Radiation Oncology Version 8.1, manufactured by Varian Medical Systems, Palo Alto, CA The device is designed to assist the operator in providing accurate treatment setups for each patient by monitoring parametDose delivery may be altered after import from a third party planning system during radiation therapy.accessdata.fda.gov

Compare with other makers

Updated January 21, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.