Skip to the page

Data / Medical devices / United States / Tas Medical, Inc.

Tas Medical, Inc.: FDA-registered device maker in Windermere, FL, United States, 2 510(k) clearances

Tas Medical, Inc. is registered with the FDA (registration 3029835193) in Windermere, FL, United States. It lists 1 device type. The FDA files 2 510(k) clearances under its listings, the newest dated June 3, 2026. 1 recall event names it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3029835193
Address
9100 Conroy Windermere Rd Suite 200, Windermere, FL, United States, 34786
Owner or operator
TAS Medical, Inc. (10089073)
Roles
Complaint file establishment, Specification developer
First device listed
November 3, 2023
Registered through
2026
Brand names listed
TAS Tissue Approximation System, Tissue Approximation System

Devices it lists

DeviceProduct codeClassRoleSource
Suture, Nonabsorbable, Synthetic, PolyamideGAR2Complaint file establishment, Specification developeraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
June 3, 2026K261484Tissue Approximation SystemGARSubstantially EquivalentTas Medical, Inc.accessdata.fda.gov
July 5, 2023K220980Tissue Approximation System (TAS)GARSubstantially EquivalentTas Medical, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
January 27, 202596274Class IIOngoingTISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAPTissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrentaccessdata.fda.gov

Compare with other makers

Updated June 3, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.