Skip to the page

Data / Medical devices / United States / Tandem Diabetes Care, Inc.

Tandem Diabetes Care, Inc.: FDA-registered device maker in San Diego, CA, United States, 21 510(k) clearances and 2 PMA approvals

Tandem Diabetes Care, Inc. is registered with the FDA (registration 3007981285) in San Diego, CA, United States. It lists 14 device types. The FDA files 21 510(k) clearances and 2 PMA approvals under its listings, the newest dated April 24, 2026. 13 recall events name it as the recalling firm (5 Class I). Registered through 2027. FDA data exported November 3, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3007981285
Address
12400 High Bluff Drive, San Diego, CA, United States, 92130
Owner or operator
Tandem Diabetes Care, Inc. (10039506)
Roles
Complaint file establishment, Manufacturer, Remanufacturer, Repackager or relabeler, Specification developer
First device listed
January 16, 2012
Registered through
2027
Brand names listed
Alternate Controller Enabled Insulin Infusion Pump, Control-IQ Technology, Interoperable Automated Glycemic Controller, Sugarmate Platform, Sugarmate Platform; Sugarmate, Alexa Skill Application; Sugarmate, Android Mobile Application; Sugarmate, iOS Mobile App, T:slim X2 Insulin Pump, Tandem Device Updater, Tandem Mobi Adhesive Sleeve and more

Devices it lists

DeviceProduct codeClassRoleSource
Accessories, Pump, InfusionMRZ2Complaint file establishment, Specification developeraccessdata.fda.gov
Alternate Controller Enabled Insulin Infusion PumpQFG2Complaint file establishment, Manufacturer, Remanufacturer, Repackager or relabeler, Specification developeraccessdata.fda.gov
Automated Insulin Dosing , Threshold SuspendOZO3Manufacturer, Remanufacturer, Repackager or relabeleraccessdata.fda.gov
Calculator, Drug DoseNDC2Manufacturer, Remanufactureraccessdata.fda.gov
Continuous Glucose Monitor Retrospective Data Analysis SoftwarePHV1Manufactureraccessdata.fda.gov
Continuous Glucose Monitor Secondary DisplayPJT2Manufactureraccessdata.fda.gov
Insulin Pump Secondary DisplayPKU2Manufactureraccessdata.fda.gov
Interoperable Automated Glycemic ControllerQJI2Complaint file establishment, Manufacturer, Specification developeraccessdata.fda.gov
Interoperable Automated Glycemic Controller, Insulin SuspendQJS2Manufacturer, Remanufactureraccessdata.fda.gov
Pump, Infusion, InsulinLZG2Complaint file establishment, Specification developeraccessdata.fda.gov
Pump, Infusion, Insulin, To Be Used With Invasive Glucose SensorOYC3Complaint file establishment, Specification developeraccessdata.fda.gov
Sensor, Glucose, InvasiveMDS3Complaint file establishment, Specification developeraccessdata.fda.gov
Sensor, Glucose, Invasive, Non-AdjunctivePQF3Complaint file establishment, Manufacturer, Remanufacturer, Repackager or relabeler, Specification developeraccessdata.fda.gov
Tape And Bandage, AdhesiveKGX1Manufactureraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
April 24, 2026K260429Control-IQ+ technologyQJISubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
November 6, 2025K253074Tandem Mobi insulin pump with interoperable technologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
September 24, 2025K250792t:slim X2 insulin pump with interoperable technologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
May 21, 2025K250798Control-IQ+ technologyQJISubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
July 26, 2024K241078Tandem Mobi Insulin Pump with interoperable technologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
March 21, 2024K240309Tandem Mobi insulin pump with interoperable technologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
November 3, 2023K232382Control-IQ TechnologyQJISubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
November 3, 2023K232380t:slim X2 Insulin Pump with Interoperable TechnologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
July 10, 2023K223213Tandem Mobi Insulin Pump with Interoperable TechnologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
February 15, 2022K203234t:slim X2 Insulin Pump with Interoperable Technology (with t:connect mobile app)QFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
June 16, 2020K200467Control-IQ TechnologyQJISubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
June 4, 2020K201214t:slim X2 Insulin Pump with Interoperable TechnologyQFGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
February 28, 2020K193483Basal-IQ TechnologyQJSSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
December 13, 2019DEN190034Control-IQ TechnologyQJIUnknownTandem Diabetes Care, Inc.accessdata.fda.gov
February 14, 2019DEN180058t:slim X2 insulin pump with interoperable technologyQFGUnknownTandem Diabetes Care, Inc.accessdata.fda.gov
October 25, 2016K162080t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device UpdaterLZGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
July 25, 2016K160056t:slim Insulin Delivery SystemLZGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
January 9, 2015K143189t:flex Insulin Delivery SystemLZGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
October 31, 2014K141758T:SLIM INSULIN DELIVERY SYSTEMLZGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
March 24, 2014K133593T:SLIM INSULIN DELIVERY SYSTEMLZGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov
November 9, 2011K111210T:SLIM (TM) INSULIN DELIVERY SYSTEMLZGSubstantially EquivalentTandem Diabetes Care, Inc.accessdata.fda.gov

PMA approvals

Premarket approvals this establishment lists, with the supplements the FDA has decided since.

ApprovedPMA numberDeviceProduct codeSupplementsLatest decisionSource
June 21, 2018P180008t:slim X2 Insulin Pump with Basal-IQ TechnologyOZO6December 19, 2019accessdata.fda.gov
September 8, 2015P140015T:SLIM G4 INSULIN PUMP WITH DEXCOM G4 PLATINUM CGMOYC29December 19, 2019accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
September 23, 202699851Not yet classifiedOpen, ClassifiedTandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below thaccessdata.fda.gov
July 16, 202699589Class IIOngoingBrand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: NoA beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM aaccessdata.fda.gov
January 30, 202698478Class IIOngoingTandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivaccessdata.fda.gov
October 6, 202597662Class IOngoingTandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin fInsulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could accessdata.fda.gov
August 5, 202597403Class IIOngoingt:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, CA software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.accessdata.fda.gov
July 21, 202597003Class IOngoingBattery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical UsInsulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.accessdata.fda.gov
July 3, 202597264Class IIOngoingt:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)an app defect that occurs when the is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.accessdata.fda.gov
February 27, 202596417Class IOngoingt:slim X2 Insulin Pump with Interoperable TechnologyA software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous gaccessdata.fda.gov
March 5, 202494312Class IOngoingt:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technologyDuring normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspeaccessdata.fda.gov
May 24, 202290628Class IIOngoingt:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ TechnologyInsulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.accessdata.fda.gov
September 10, 202090113Class IIOngoingt:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ TechnologyInsulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stopaccessdata.fda.gov
April 23, 201879887Class IIOngoingt:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alaraccessdata.fda.gov
January 10, 201467264Class ITerminatedT:slim Insulin Delivery System Insulin Infusion PumpSpecific lots of insulin cartridges that are used with the T:slim Insulin Pump may be at risk for leaking. A cartridge leak could result in the device delivering too much or too little insulin, which can lead to a serious adverse event.accessdata.fda.gov

Compare with other makers

Updated October 30, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.