Skip to the page

Data / Medical devices / Spain / Sedecal S.A.

Sedecal S.A.: FDA-registered device maker in ALGETE Madrid(Province), Spain, 18 510(k) clearances

Sedecal S.A. is registered with the FDA (registration 9617251) in ALGETE Madrid(Province), Spain. It lists 9 device types. The FDA files 18 510(k) clearances under its listings, the newest dated August 21, 2023. 11 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
9617251
FEI number
3002495042
Address
Pelaya, 9 Pol. Ind. Rio De Janeiro, ALGETE Madrid(Province), Spain, 28110
Owner or operator
SEDECAL S.A. (9039772)
Roles
Contract manufacturer, Manufacturer
First device listed
November 30, 1999
Registered through
2026
Brand names listed
Aero DR TX m01, BRS, Challenge-X, Chiropro, FLEXI DT, FM Plus, MILL-WBS, MILLENNIUM and more

Devices it lists

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
August 21, 2023K232185SM-IVIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
November 21, 2022K222951SM-VIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
July 18, 2022K221803PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress.IZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
February 1, 2021K203537Proteus XR/f, Models ST and ET, Multirad and Multirad NET Radiographic SystemsKPRSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
October 7, 2020K202293Radiographic System Challenge XKPRSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
November 15, 2019K192936Soltus 500IZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
September 25, 2019K192011PhoeniXIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
August 2, 2019K191813MobileDiagnost wDR 2.2IZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
January 26, 2018K173299DRAGON X SPSL4HC, DRAGON X SPSL8HCIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
March 21, 2014K133782NOVA FA DR SYSTEMMQBSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
August 2, 2012K121410DRAGON LWIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
January 26, 2011K103522SEDECALIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
March 17, 2009K090322EASY MOVING PLUS, SM-XXHF-YYIZLSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
March 11, 2009K090279SEDECAL, MILLENNIUM PLUSKPRSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
February 27, 2009K090238SEDECAL X-PLUS LP PLUS DIGITAL DIAGNOSTIC X-RAY SYSTEMSKPRSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
September 14, 2006K062335SEDECAL X PLUS LP PLUSKPRSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
November 18, 2004K042876URS LP DIGITAL, MILLENNIUM, RADPROKPRSubstantially EquivalentSedecal., Sa.accessdata.fda.gov
November 2, 2001K012546OPTIMA URS UNIVERSAL RADIOGRAPHIC SYSTEMKPRSubstantially EquivalentSedecal USA, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
June 10, 202597075Class IIOngoingSedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)Sedecal sent an to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.accessdata.fda.gov
January 22, 202494013Class IIOngoingBattery Mobile X-ray Unit EASY MOVING Model with telescopic column 3.1, model numbers: a. SM-20HF-Batt; b. SM-32HF-Batt; c. SM-40HF-Batt; d. SM-50HF-Batt; e. SM-40-HF-B-D-VIR; f. SM-20HF-B-D-KM (AeroDR X30); g. SM-32-HF-Potential interference with the anti-fall system.accessdata.fda.gov
December 13, 202393614Class IIOngoingSedecal X Optima URS digital mobile diagnostic x-ray system. A a stationary x-ray unit with a universal swivel arm. It allows one to take exposures of patients in standing, sitting, or laying position.Some parts have fallen off the equipment resulting in a safety risk to patients and/or users.accessdata.fda.gov
May 23, 202392464Class IIOngoingwDR 2.2 Mobile Digital Diagnostic X-Ray SystemThere is a software login in issue that may prevent the user from logging in.accessdata.fda.gov
February 6, 202391656Class IIOngoingPhoenix mKDR, digital mobile diagnostic x-ray systemSedecal was informed of 3 incidents of erratic movements related to X-ray mobile systems.accessdata.fda.gov
August 12, 202290736Class IIOngoingSedecal Soltus Battery Mobile X-Ray System REF Soltus 450 Battery Mobile X-Ray System REF Soltus 500Due to un-commanded movement of mobile x-ray unit.accessdata.fda.gov
October 14, 202193651Class IIOngoingMobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.accessdata.fda.gov
July 18, 201881377Class IIOngoingSedecal NOVA FA DR System. for diagnostic radiography.A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.accessdata.fda.gov
January 18, 201673428Class IITerminatedMobile wDR motorized portable diagnostic X-ray systems. Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinaThis is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we have observed that static electricity can also elicit the unwanted behavior.accessdata.fda.gov
July 13, 201571714Class IITerminatedRadPro 40kW Digital X-Ray Panel and Mobile/Portable Diagnostic X-Ray. Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the Electrostatic energy may be stored in Systems that can cause a short-cicuit which, in turn, can generate an Electomagnetic Pulse which might disturb the electronics of the Digital Motion Control Board producing uncontrolled movements and such uncontrolled movements could cause the system to bump into the user or ppatient.accessdata.fda.gov
August 6, 201470237Class IITerminatedMobileDiagnost wDR motorized portable diagnostic X-ray systems.Reports of unexpected movement.accessdata.fda.gov

Compare with other makers

Updated October 23, 2025. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.