Data / Medical devices / United States / Radiometer America, Inc.

Radiometer America, Inc.: FDA-registered device maker in Middleburg Heights, OH, United States, 1 510(k) clearance

Radiometer America, Inc. is registered with the FDA (registration 1523456) in Middleburg Heights, OH, United States. It lists 2 device types. The FDA files 1 510(k) clearance under its listings, the newest dated January 28, 2016. 44 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
1523456
Address
6836 Engle Rd, Middleburg Heights, OH, United States, 44130
Owner or operator
RADIOMETER MEDICAL APS (8020503)
Roles
Complaint file establishment, Repackager or relabeler
First device listed
April 28, 1986
Registered through
2026
Brand names listed
SAFEPICO

Devices it lists

DeviceProduct codeClassRoleSource
Arterial Blood Sampling KitCBT1Complaint file establishment, Repackager or relabeleraccessdata.fda.gov
Multi-Analyte Controls, All Kinds (Assayed)JJY1Complaint file establishment, Repackager or relabeleraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
January 28, 2016K153712VK-3 Verification Kit, VK-4 Verification Kit, VK-R5 Verification Kit, VK-R7 Verification Kit, VK-Crea Verification KitJJYSubstantially EquivalentRadiometer America, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
December 8, 201470103Class IITerminatedABL90 FLEX analyzer; Model number. A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.The ABL90 analyzer does not always use the most recent calibration data to calculate patient results. This can in some cases lead to a biased patient result.accessdata.fda.gov
October 29, 201469813Class IITerminatedCrea A and B membranes, packaged 4 per box Used in conjunction with the ABL800 FLEX analyzers which are intended for: In Vitro Testing of samples of whole blood for the parameters pCO2, cK+,cNa+, cCa2+,cCl-,cGlu, cLac, cNegative drift on QC during in-use lifetime of CREA membranes can be observed.accessdata.fda.gov
October 20, 201469663Class IITerminatedHemoCue Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201accessdata.fda.gov
July 28, 201468982Class IITerminatedRADIANCE v. 2.4X, 2.5X and 2.6X Product Usage: RADIANCE is a data processing application for clinical use intended to monitor, collect, store, retrieve and process laboratory data. RADIANCE is also intended to control claccessdata.fda.gov
October 15, 201366875Class IITerminatedABL90 FLEX Analyzer Radiometer Medical ApS, kandevej 21 Brnshj, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.accessdata.fda.gov
July 5, 201367066Class IITerminatedpCO2 D788 Membranes. Distributed by Radiometer America, Inc., Westlake, OH. An automatic or semi-automatic instrument used to identify and quantify two or more gases, and sometimes electrolytes, in whole blood using multRADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.accessdata.fda.gov
June 10, 201367044Class IITerminatedpO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibrataccessdata.fda.gov
November 27, 201265843Class IITerminatedTCM4 series Base Units, and spare part Power Supply, Radiometer Medical ApS, Akandevej 21, Brnshj, Denmark, www.radiometer.com. The devices are intended for continuous monitoring of carbon dioxide and oxygen partial presRADIOMETER became aware that the TCM4 series Base Unit does not comply with IEC 60601-1 Sub Clause 15B.accessdata.fda.gov
October 8, 201266108Class IITerminatedLactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.accessdata.fda.gov
July 5, 201262503Class IITerminatedThe ABL software version 6.12 for the ABL800 analyzer. The analyzers are intended for the in Vitro testing of: -Samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl, cGlu, cLac, cCrea, ctBil, anA customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixeaccessdata.fda.gov
December 16, 201160929Not yet classifiedTerminatedS1827 Crea Calibration solution 1,, Lot DV-02 Radiometer Medical, Bronshoj Denmark Product Usage: The S1827 Crea Calibration Solution ) Lot DV-02 is a solution used by the ABL 8x7 series creatine analyzers to establish cWhen the user activates the dosicap (containing the crea additive) and mounts it onto the bottle, the threads will not seal correctly causing approximately 10% of the solution to leak out when the bottle is placed horizontally for dissolving the additive. This will cause the concentration of creatinin and creatin in the calibration solution to be wrong, and accessdata.fda.gov
September 27, 201160153Not yet classifiedTerminatedTC pCO2 Electrodes, Model/Part #945-655-E5260, 510(k) #K043003. Product Usage: The device is intended for used with the firm's TCM series monitor. The E260mpCO2 electrode is a non-invasive device, used with the TCM serieTranscutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all calibrations and never get ready for use.accessdata.fda.gov
April 7, 201158657Not yet classifiedTerminatedCa++ Membrane Units, Model #, and Model # used on ABL505/555, EML105, ABL600, ABL700 Series, and ABL800 Analyzers. The membranes are packaged in a plastic box which contains four (4) membranes per box. Calcium membrane jBased on internal testing and user/customer complaint reports received, the firm determined that wrinkles may arise on the outer membrane of the calcium membrane units upon installation. Retention of fluids between the wrinkles can cause a positive bias resulting in calcium measuring errors. This situation poses a risk to the safety of patients, particularlyaccessdata.fda.gov
January 12, 201158273Not yet classifiedTerminatedABL90 FLEX Portable Blood Gas Analyzer System, Part Number:, 510 (k) #K092686. All units equipped with software versions: 2.4.1680.25 and all previous versions are subject to recall. The ABL90 FLEX is a portable, automatFollowing the receipt of two (2) reported customer complaints, the firm determined that the current operating software for their ABL 90 Analyzer does not prevent the further use of the sensor cassette when there is the possibility of iron rich plaque build-up in the reference electrode. If the device operator/user ignores the error message and continues to oaccessdata.fda.gov
November 1, 201057275Not yet classifiedTerminatedABL90 FLEX Blood Analyzer Part Number:. Intended for use by trained technologists, nurses, physicians and therapists and for use in a laboratory environment, near patient or point-of-care setting. Also, Intended for in vDue to a software error, a patient mix up can occur on the ABL90. If a sample is pre-registered and the sample "fails" the next result will inherit the patient data from the previous result.accessdata.fda.gov
April 16, 200951986Not yet classifiedTerminatedWaste bottles Part Numbers/Article Numbers: /D512 and /D513. The waste bottles are used to collect waste from the ABL7XX and the ABL8XX blood gas analyzer after rinsing.When removing the waste bottle from the analyzer waste material may leak both from the analyzer and from the waste bottle.accessdata.fda.gov
March 26, 200951983Not yet classifiedTerminatedReference Electrode Membranes for use with blood gas analyzer ABL 800 FLEX, when used on ABL 827 and ABL 837 with Creatinine modules (8x7). The device is packaged four (4) units per box.Analysis of customer data, (for ABL 8X7 customers) has shown that the reference membranes are out of analytical specifications when replaced as specified in the device labeling. These errors in analytical specifications could lead to inadequate or clinical interventions (therapy) or misleading interpretation of patient status in clinical settings.accessdata.fda.gov
September 5, 200849468Not yet classifiedTerminatedABL7XX Calibration 1 Solution S1720. P/N S1720, Intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oxThe firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the calibration curves for these parameters and thereby the measured results for patient samples and quality controls will be biased.accessdata.fda.gov
January 18, 200846686Not yet classifiedTerminatedABL700 Series analyzer - Human Blood Analyzer (Laboratory Device).Reported calibration errors are not displayed on the analyzers parameter bar screen or on the patient results when the calibration number is between 32768 - 65535 and 98303 - 131070. This may cause the operator to perform measurements and obtain questionable results.accessdata.fda.gov
June 22, 200738511Not yet classifiedTerminatedPump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.accessdata.fda.gov
January 16, 200737167Not yet classifiedTerminatedABL800 FLEX Series analyzer with FLEXQ module-equipped with software version 5.2 to 5.27In some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood sampler associated with a specific patient is pre-registered in the analyzer by scanning a barcode.accessdata.fda.gov
November 13, 200636812Not yet classifiedTerminatedABL 700 Series blood gas analyzers equipped with Software Version 3.833.Software anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes of blood gas analyzers. This condition can influence analyzer results. When the protein removal program is run too frequently, more than once accessdata.fda.gov
July 18, 200636009Not yet classifiedTerminatedSafePICO Samplers, self-filling arterial blood samplers. The syringe blood samplers are individually packed in plastic pouches and distributed in cartons. Each carton contains ten (10) boxes and each box contains 100 pouLeaking Tip Caps-Tip caps may leak; after air has been expelled from the syringe through the vented safe tip caps.accessdata.fda.gov
July 6, 200636539Not yet classifiedTerminatedRadiance Data Management System, Software Version 2.60, modification for the ABL800 Flex Blood Gas AnalyzerIncorrect FI02 values: programming issue can cause incorrect values to be transmitted to the LIS when : 1)- The FIO2 result is edited in the manual sample processing mode--2)- An existing result is opened and FI02 is then edited and sent.....In both cases RADIANCE will transmit the original FI02 value, not the value that was edited.accessdata.fda.gov
June 29, 200636007Not yet classifiedTerminatedSafeClinitubes, Capillary caps (purple),. The defective capillary caps may be included in capillary tube kits as follows: Clinitubes Kits: (D941P-240-85x1), Lot Code: R0048; Clinitubes Kits: (D957P-70-100x1), Lot Codes: Leaking Capillary Caps- The dimensions and shapes of the capillary caps do not fit the capillary tubes correctly. This can cause the capillary tubes to leak.accessdata.fda.gov
June 22, 200637526Not yet classifiedTerminatedSafeClinitubes, REF, D957P-70-100x1, contained in kits containing 250 SafeClinitubes, insert, mixing wires and end caps. Capillary samplers for drawing small amounts of blood for blood gas analysis.The clinitubes are not able to be filled due to contamination with silicone during production.accessdata.fda.gov
June 6, 200637530Class IIITerminatedQualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF. The solution is used for quality check of blood gas analyzers.The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.accessdata.fda.gov
May 22, 200635695Not yet classifiedTerminatedModel # (pump tube for waste) & Model # (pump tube for solution), Waste Tubing used on ABL 7xx and ABL8xx, component of the ABL700/800 Blood Gas Analyzers, also included in service kits part no.An increased failure rate was observed with the yellow rubber tubing. The tubing was found to be leaking blood and solutions after distribution.accessdata.fda.gov
May 1, 200635698Not yet classifiedTerminatedDri-Tek BG-OX Cartridge, Model #D7120, Cat. # Cartridges used in the NPT7 Analyzer, an in vitro diagnostic device.Air bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.accessdata.fda.gov
April 7, 200635696Not yet classifiedTerminatedTCM4 Series Monitoring System (Base Unit) (affected device) CompactFlash cards: (Defective Device Component)System shut down-When the TCM Monitor is turned on and the booting process begins, the device stops after the memory count and will not proceed further.accessdata.fda.gov
March 22, 200635775Not yet classifiedTerminatedCapillary (end) caps, accessory for use with blood gas analyzers included in Radiometer''s CLINITUBES kits. The end caps are packaged in bags. These bags are placed in the capillary tube kit which also contains: capillarCapilliary end caps, an accessory for ABL blood gas analyzers, leak blood from capillary tubes during transport.accessdata.fda.gov
November 1, 200535789Not yet classifiedTerminatedD826 Accessory kit for single tcpO2 Electrodes for use with the TCM400 Transcutaneous Monitor (Cutaneous Oxygen Monitor)Excessive Drift-the membrane units of the device cause the electrode to exceed performance standards for drift in the first two or more calibrationsaccessdata.fda.gov
October 31, 200535796Not yet classifiedTerminatedABL800 Series Blood Gas Analyzer equipped with Software Version 5.21.ABL800 Series Blood Gas Analyzer became inoperable when the 'restore default setup' button was touched.accessdata.fda.gov
October 31, 200535663Not yet classifiedTerminatedS1730 Calibration Solution 2, 200 mL, packed in single unit plastic bottles, REF, used with ABL700 Series Blood Gas Analyzers.Calibration solution for ABL700 Blood Gas Analyzers is labeled with an incorrect bar code. The bar code identified on the Cal Solution 2 product is actually the bar code for the rinse solution.accessdata.fda.gov
September 29, 200534963Not yet classifiedTerminatedBlood Gas and Co-Ox2 Electrolyte & Metabolyte Analyzers, ABL-700 Series, manufactured from 8/1998 through 11/2004.Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).accessdata.fda.gov
September 19, 200535788Not yet classifiedTerminatedReference Electrodes for ABL 700/800 Blood Gas Analyzers. Electrode Part Number:, distributed one (1) electrode per box.Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.accessdata.fda.gov
June 17, 200535665Not yet classifiedTerminatedD711, Reference Membranes, Part #, used with the ABL835 Blood Gas Analyzers.The recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.accessdata.fda.gov
June 17, 200534953Not yet classifiedTerminatedBlood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-700 series.Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.accessdata.fda.gov
June 9, 200535774Not yet classifiedTerminatedABL 5 Capillary Adaptor (a connection that fits the capillary tube on to the aspirator on the ABL 5 & BPH 5 Analyzers), in vitro diagnostic accessory, Part Number:.The dimensions and shape of the analyzer adaptor do not fit the analyzer tubes or probe, causing air to be aspirated with the blood sample and causing a deviation in the pO2 sample results.accessdata.fda.gov
May 24, 200535773Not yet classifiedTerminatedPackage Insert which accompanies the Hematocrit and Metabolite QUALICHECK LEVEL 2 Quality Control of Hematocrits and Metabolites, used with the ABL 77/70/555 Blood Gas & Electrolyte Analyzers. The QC reagents are packageThe package insert, which accompanied the QC reagents used with ABL blood gas analyzers, referenced incorrect control ranges for Hct (hematocrit) testing. The range stated on the package insert was too low to generate accurate Hct test results (The control ranges were acceptable for the glucose and lactate testing).accessdata.fda.gov
March 2, 200535699Not yet classifiedTerminatedRadiance Data Management System Software Version 2.42 (Stand-alone software package) for the Radiance STAT Analyzer Management System, in vitro diagnostic.Software anomaly. Data generated by blood gas equipment did not correlate with data shown in the Radiance Data Management System.accessdata.fda.gov
February 2, 200535700Not yet classifiedTerminatedTC Monitor - Transcutaneous Monitor for p02 and pC02, Model #TCM4, (Cutaneous Oxygen Monitor)Alarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped.accessdata.fda.gov
February 1, 200535649Not yet classifiedTerminatedD 744 Chloride membrane, CI, REF Model #, accessory to the ABL700 blood analyzer.The CI membrane units generate too low values for status and sensitivity during calibration.accessdata.fda.gov
January 12, 200535664Not yet classifiedTerminatedD7120, BG-OX Cartridge, D7120 REF, and, cuvettes for the NPT7 series of blood gas analyzers.QC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH.accessdata.fda.gov

Compare with other makers

Updated January 28, 2016. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.