Skip to the page

Data / Medical devices / Denmark / NRT X-Ray A/S

NRT X-Ray A/S: FDA-registered device maker in HASSELAGER Midtjylland, Denmark, 3 510(k) clearances

NRT X-Ray A/S is registered with the FDA (registration 1000188474) in HASSELAGER Midtjylland, Denmark. It lists 2 device types. The FDA files 3 510(k) clearances under its listings, the newest dated December 23, 2024. 4 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
1000188474
Address
Birkegaardsvej 16, HASSELAGER Midtjylland, Denmark, 8361
Owner or operator
NRT X-RAY A/S (9059076)
Roles
Complaint file establishment, Manufacturer
First device listed
October 14, 2007
Registered through
2026
Brand names listed
Adora DRFi, GE PRECISION MPI, Intelli C

Devices it lists

DeviceProduct codeClassRoleSource
Interventional Fluoroscopic X-Ray SystemOWB2Complaint file establishment, Manufactureraccessdata.fda.gov
System, X-Ray, Fluoroscopic, Image-IntensifiedJAA2Complaint file establishment, Manufactureraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
December 23, 2024K242948Adora DRFi (04550010)JAASubstantially EquivalentNrt X-Ray A/Saccessdata.fda.gov
July 11, 2018K173631Intelli-COWBSubstantially EquivalentNrt X-Ray A/Saccessdata.fda.gov
November 19, 2003K033486GE PRECISION MPIOWBSubstantially EquivalentNrt-Nordisk Rontgen Teknik A/Saccessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
April 17, 202494363Class IIOngoingIntelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray systemX-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting taccessdata.fda.gov
April 27, 202392230Class IIOngoingIntelli-C, Right side suspended Tabletop, REF: 03400000There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.accessdata.fda.gov
January 21, 202289413Class IIOngoing1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.accessdata.fda.gov
May 1, 200952173Not yet classifiedTerminatedCelex/GE Precision MPi, REF no.: 02980000 Is an all-digital multipurpose tilt-C X-ray system, intended for a multitude of diagnostic procedures, including: R&F, radiology, fluoroscopy, interventional procedures, vascularThe GE Precision MPi is in violation with 21 CFR Part 1020.32 requiring a signal audible to the fluoroscopist to sound for each passage of 5 minutes of fluoroscopic irradiation time during an examination or procedure.accessdata.fda.gov

Compare with other makers

Updated December 23, 2024. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.