Skip to the page

Data / Medical devices / United States / MEDTRONIC NAVIGATION, INC. (Boxborough)

MEDTRONIC NAVIGATION, INC. (Boxborough): FDA-registered device maker in Boxborough, MA, United States, 6 510(k) clearances

MEDTRONIC NAVIGATION, INC. (Boxborough) is registered with the FDA (registration 3004785967) in Boxborough, MA, United States. It lists 4 device types. The FDA files 6 510(k) clearances under its listings, the newest dated June 21, 2024. 21 recall events name it as the recalling firm. Registered through 2027. FDA data exported November 3, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3004785967
Address
200 Beaver Brook Rd, Boxborough, MA, United States, 01719
Owner or operator
Medtronic, Inc. (2112641)
Roles
Remanufacturer, Specification developer
First device listed
March 7, 2006
Registered through
2027
Brand names listed
O-Arm, O-Arm Imaging System, O-Arm O2 Imaging System, O-arm Equipment Drape, O-arm™ O2 imaging system, Surgical drape and drape accessories

Devices it lists

DeviceProduct codeClassRoleSource
Drape, Surgical, ExemptPUI2Specification developeraccessdata.fda.gov
Image-Intensified Fluoroscopic X-Ray System, MobileOXO2Remanufacturer, Specification developeraccessdata.fda.gov
Interventional Fluoroscopic X-Ray SystemOWB2Remanufacturer, Specification developeraccessdata.fda.gov
System, X-Ray, Fluoroscopic, Image-IntensifiedJAA2Remanufacturer, Specification developeraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
June 21, 2024K240465O-arm O2 Imaging SystemOWBSubstantially EquivalentMedtronic Navigation, Inc.accessdata.fda.gov
April 24, 2020K200074O-arm O2 Imaging SystemOWBSubstantially EquivalentMedtronic, Inc.accessdata.fda.gov
August 6, 2015K151000O-Arm O2 Imaging SystemOWBSubstantially EquivalentMedtronic, Inc.accessdata.fda.gov
June 17, 2010K092564O-ARM IMAGING SYSTEMOXOSubstantially EquivalentMedtronic Navigation, Inc. (Littleton)accessdata.fda.gov
March 29, 2006K060344O-ARM IMAGING SYSTEMOXOSubstantially EquivalentBreakaway Imaging, LLCaccessdata.fda.gov
May 5, 2005K050996O-ARM IMAGING SYSTEMOXOSubstantially EquivalentBreakaway Imaging, LLCaccessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
May 19, 202699041Class IIOngoingMedtronic O-arm O2 Imaging System. Model Number: BI70002000.Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.accessdata.fda.gov
May 27, 202597045Class IIOngoingO-arm O2 Imaging System. Mobile X-Ray System.A mechanical component that supports the O2 gantry to the support structure in affected devices may have been damaged during manufacture.accessdata.fda.gov
April 26, 202494566Class IIOngoingO-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000Potential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.accessdata.fda.gov
July 25, 202392778Class IIOngoingThe O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is inGround cable installed incorrectly.accessdata.fda.gov
November 15, 201984441Class IITerminatedO-arm" O2 Imaging System, Product Number/CFN: BI-700-02000Due to a manufacturing issue affecting particular O-arm" O2 Imaging Systems, Medtronic has determined that the gantry tractor motor drive belt may loosen over an extended number of uses at a higher frequency in these systems. If the belt loosens, it may result in rotation of 3D images about the gantry isocenter. While the 3D image is anatomically accurate wiaccessdata.fda.gov
May 1, 201982991Class IITerminatedMedtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the system voltage): Product Name Model#/ CFN OARM ASSY 8170000027 SYSTEM PRODUCT BI70000027 OARM ASSY BI70000027R SYS PThe 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affecaccessdata.fda.gov
September 17, 201881355Class IITerminatedMedtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000RO-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuitaccessdata.fda.gov
August 30, 201881153Class IITerminatedO-arm 1000 2nd Edition Imaging System, OARM ASSY Bl70000027 SYSTEM PRODUCT, Model Number Bl70000027 Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is inteSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.accessdata.fda.gov
January 25, 201879275Class IITerminatedO-arm 1000 mobile image-intensified fluoroscopic x-ray system; (1) Model: BASE OARM Bl70000028100 SYS 100V, Product Number: Bl70000028100 (UDI: 00643169354418); (2) Model: BASE OARM Bl70000028120 SYS 120V, Product NumberSoftware update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose display/report, Gantry motion, Network communication, System shutdown, and Early termination of 3D spin.accessdata.fda.gov
September 30, 201779272Class IITerminatedO-arm O2 interventional fluoroscopic x-ray system; Catalog Number: B1-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighting During servicing of certain internal components of the O-Arm O2 Surgical Imaging System, it is possible that the high-power cable connector may come in contact with surrounding metal during the disconnecting/reconnecting process which may result in an electrical short and a shock to the field service technician.accessdata.fda.gov
September 22, 201675279Class IITerminatedMedtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spiO-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated imagesaccessdata.fda.gov
March 17, 201674158Class IITerminatedMedtronic Navigation O-arm 02 surgical x-ray imaging systemReview of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).accessdata.fda.gov
March 17, 201674153Class IITerminatedMedtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.accessdata.fda.gov
January 29, 201673303Class IITerminatedMedtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image qualityaccessdata.fda.gov
July 25, 201367363Class IITerminatedMedtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where aReview of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)accessdata.fda.gov
January 31, 201364256Class IITerminatedMedtronic O-arm Imaging System Product Usage: image-intensified fluoroscopic x-ray system, mobilePotential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical directionaccessdata.fda.gov
July 11, 201262571Class IITerminatedMedtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.accessdata.fda.gov
October 15, 201057048Not yet classifiedTerminatedMedtronic Navigation O-Arm Imaging System, Door Assembly (BI-700-00006) used in Product Catalog Numbers: 9732719,9733346,81-700-00027, 81-700-00027R, 81700- 00027-100, 81-700-00027-1 OOR, 81-700-00027-120, 81-700-00027-1Safety Notice: Door Cable Periodic Maintenance Schedule and Cable Wear for the O-ARM Imaging System Breakage of the door cable during opening and closing may occur and cause physical harm or delay of therapyaccessdata.fda.gov
October 12, 201061741Class IITerminatedMedtronic 0-arm Imaging System with Software Version 3.1.1, Catalog Numbers: BI-700-00027-120R and BI-700-00027-120. The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging.Software Version 3.1.1 does not meet navigational accuracy.accessdata.fda.gov
May 10, 201055675Not yet classifiedTerminatedO-Arm Imaging System The O-Arm imaging system is a mobile X-Ray system designed for medical applications in surgical theaters, focusing on 3D imaging but also offering 2D fluoroscopic Imaging.The possibility that the real time image display can freeze up during use of a 2D continuous fluoroscopy.accessdata.fda.gov
June 30, 200849501Not yet classifiedTerminatedBI-700-00027 0-arm 1000 Imaging System Mobile X-ray System Generating 2D and 3D images of human anatomy for surgical applications.Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii)accessdata.fda.gov

Compare with other makers

Updated June 15, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.