Data / Medical devices / United States / Luminex Corporation
Luminex Corporation: FDA-registered device maker in Austin, TX, United States, 7 510(k) clearances
Luminex Corporation is registered with the FDA (registration 1650733) in Austin, TX, United States. It lists 10 device types. The FDA files 7 510(k) clearances under its listings, the newest dated April 12, 2016. 14 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 3, 2026.
SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.
- FDA registration number
- 1650733
- FEI number
- 3002524000
- Address
- 12212 Technology Blvd, Austin, TX, United States, 78727
- Owner or operator
- Luminex Corporation (9032042)
- Roles
- Complaint file establishment, Contract manufacturer, Manufacturer, Remanufacturer
- First device listed
- April 16, 2008
- Registered through
- 2026
- Brand names listed
- FLEXMAP 3D, Gastrointestinal Pathogen Panel (GPP), LIAISON PLEX System, Luminex 100, Luminex 100/200, Luminex 200, MAGPIX, MAGPIX® Drive Fluid PLUS (4PK) and more
Devices it lists
510(k) clearances
The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.
| Decided | 510(k) number | Device | Product code | Decision | Applicant | Source |
|---|---|---|---|---|---|---|
| April 12, 2016 | K160517 | ARIES System | OOI | Substantially Equivalent | Luminex Corporation | accessdata.fda.gov |
| December 17, 2015 | K152386 | NxTAG Respiratory Pathogen Panel, NxTAG Respiratory Pathogen Panel Files, SYNCT Software | OCC | Substantially Equivalent | Luminex Molecular Diagnostics, Inc. | accessdata.fda.gov |
| September 4, 2015 | K143653 | Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex) | OCC | Substantially Equivalent | Nanosphere, Inc. | accessdata.fda.gov |
| January 10, 2014 | K133302 | FLEXMAP 3D | NSU | Substantially Equivalent | Luminex Corp. | accessdata.fda.gov |
| March 21, 2013 | K121894 | XTAG GASTROINTESTINAL PATHOGEN PANEL(GPP) XTAG DATA ANALYSIS SOFTWARE FOR GPP (TDAS GPP) LUMINEX MAGPIX | PCH | Substantially Equivalent | Luminex Molecular Diagnostics, Inc. | accessdata.fda.gov |
| January 9, 2013 | K121399 | LUMINEX FLEXMAP 3D INSTRUMENT SYSTEM, LUMINEX XPONENT 4.0 SPI SOFTWARE, FLEXMAP 3D IVD CALIBRATION KIT, FLEXMAP 3D IVD P | NSU | Substantially Equivalent | Luminex Corp. | accessdata.fda.gov |
| March 7, 2008 | K073506 | LUMINEX LX 100/200 INSTRUMENT | NSU | Substantially Equivalent | Luminex Corp. | accessdata.fda.gov |
Recalls
Recall events where this establishment is the recalling firm, matched on its FEI.
| Initiated | Event | Class | Status | Product | Reason | Source |
|---|---|---|---|---|---|---|
| September 14, 2022 | 90937 | Class II | Ongoing | ARIES SARS-CoV-2 Assay, REF: 50-10047 | A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result. | accessdata.fda.gov |
| July 14, 2022 | 90698 | Class II | Ongoing | ARIES SARS-CoV-2 Assay | SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative result would be generated. | accessdata.fda.gov |
| April 5, 2022 | 90023 | Class II | Ongoing | Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system | Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results. | accessdata.fda.gov |
| November 17, 2021 | 89244 | Class II | Terminated | ARIES HSV 1&2 Assay, Catalog/REF# 50-10017 | Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid. | accessdata.fda.gov |
| April 16, 2021 | 89303 | Class II | Terminated | ARIES HSV 1&2 Assay, Part Number/REF 50-10017 | Impacted lot may leak inside the ARIES instrument. | accessdata.fda.gov |
| June 15, 2020 | 85872 | Class II | Terminated | Verigene Processor SP with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge | An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call. | accessdata.fda.gov |
| February 20, 2019 | 83470 | Class II | Ongoing | Aries Bordetella Assay, REF 50-10037, UDI # 00840487101452 | Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay. | accessdata.fda.gov |
| December 26, 2018 | 81890 | Class III | Terminated | Vergene Amplification Tray - 700uL, BC-GP REF 20-011-018 The Verigene(R) Gram-Positive Blood Culture Nucleic Acid Test (BC-GP) performed using the sample-to-result Verigene System is a qualitative, multiplexed in vitro d | On lot of VERIGENE(R) BC-GP Utility Kits (20-011-018) were labeled with the VERIGENE(R) Amplification Tray label. | accessdata.fda.gov |
| December 10, 2018 | 86358 | Class II | Ongoing | VERIGENE Processor SP Model No. 10-0000-07. An in vitro diagnostic device intended for processing and identifying nucleic acids | No results/incorrect results due to failure of the hybridization heater. | accessdata.fda.gov |
| September 14, 2018 | 81220 | Class II | Terminated | VERIGENE Blood Culture Gram-Positive Nucleic Acid Kits (packaged as five-pack carriers) Product Usage: In vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive ba | Potential to lead to a false negative results. | accessdata.fda.gov |
| June 29, 2018 | 81420 | Class II | Terminated | Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017 | An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order. | accessdata.fda.gov |
| May 8, 2018 | 80517 | Class II | Terminated | Verigene Enteric Pathogens Nucleic Acid Test (EP) for detecting and identification of pathogenic enteric bacteria, viruses, and genetic virulence markers from liquid or soft stool preserved in Cary-Blair medium, collecte | Elevated false positive results Yersinia enterocolitica (Yersinia) from customers using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit (20-005-023). | accessdata.fda.gov |
| April 19, 2017 | 77052 | Class II | Terminated | ARIES System and ARIES M1 System | The hand-held barcode scanner that is used as an accessory to enter ARIES cassette and specimen identifications on the ARIES Systems populates patient sample IDs with a previously scanned sample ID when entering orders into an ARIES instrument. | accessdata.fda.gov |
| May 24, 2011 | 59345 | Not yet classified | Terminated | xTAG CYP2D6 Kit v3 IVD For use with Luminex 100/200 Instrument Luminex Molecular Diagnostics, Inc. 439 University Ave. Toronto, Ontario Canada Catalog # I030B0373 The xTAG CYP2D6 Kit v3 is a device used to simultaneously | Inaccurate test results, due to possible defective primers, which could lead to incorrect dosage on manufactured drug assays. | accessdata.fda.gov |
Compare with other makers
- All makers of Instrumentation For Clinical Multiplex Test Systems
- All makers of Influenza A And Influenza B Multiplex Nucleic Acid Assay
- All makers of Influenza A Virus Subtype Differentiation Nucleic Acid Assay
- All makers of Human Metapneumovirus (Hmpv) Rna Assay System
- All makers of Real Time Nucleic Acid Amplification System
- All makers of Bordetella Pertussis Dna Assay System
Updated October 26, 2022. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.