Data / Medical devices / United States / Invivo Corporation
Invivo Corporation: FDA-registered device maker in Orlando, FL, United States, 5 510(k) clearances
Invivo Corporation is registered with the FDA (registration 1051786) in Orlando, FL, United States. It lists 3 device types. The FDA files 5 510(k) clearances under its listings, the newest dated September 30, 2021. 19 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.
SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.
- FDA registration number
- 1051786
- Address
- 12151 Research Pkwy, Orlando, FL, United States, 32826
- Owner or operator
- Philips Medical Systems International BV (1217116)
- Roles
- Specification developer
- First device listed
- January 27, 2010
- Registered through
- 2026
- Brand names listed
- 2-CH Patch, 40494E, Limb Clamp Electrode, ADVANCED FILTER ECG CABLE - 989803170121, Advanced Apps ECG Cable - 989803176381, Anesthetic Oxygen (O2) Sensor - 989803162051, BTP-1000P (Patch), CAB,4 LD,CV MRI ECG - 9223, CAB,4 LD,MRI ECG - 9224 and more
Devices it lists
| Device | Product code | Class | Role | Source |
|---|---|---|---|---|
| Electrode, Electrocardiograph | DRX | 2 | Specification developer | accessdata.fda.gov |
| Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) | MWI | 2 | Specification developer | accessdata.fda.gov |
| Recorder, Paper Chart | DSF | 1 | Specification developer | accessdata.fda.gov |
510(k) clearances
The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.
| Decided | 510(k) number | Device | Product code | Decision | Applicant | Source |
|---|---|---|---|---|---|---|
| September 30, 2021 | K212227 | Philips MR Patient Care Portal | MWI | Substantially Equivalent | Invivo Corporation | accessdata.fda.gov |
| December 23, 2015 | K152330 | Expression MR400 MRI Patient Monitoring System | MWI | Substantially Equivalent | Invivo Corporation | accessdata.fda.gov |
| July 15, 2013 | K131382 | EXPRESSION MR200 MRI PATIENT MONITORING SYSTEM, EXPRESSION IP5 INFORMATION PORTAL | MWI | Substantially Equivalent | Invivo Corporation | accessdata.fda.gov |
| February 22, 2013 | K124061 | MRI PATIENT MONITORING SYSTEM (MODEL 865214) | MWI | Substantially Equivalent | Invivo Corporation | accessdata.fda.gov |
| December 15, 2009 | K090785 | MRI COMPATIBLE PATIENT MONITORING SYSTEM, MODEL 865214 | MWI | Substantially Equivalent | Invivo Corp. | accessdata.fda.gov |
Recalls
Recall events where this establishment is the recalling firm, matched on its FEI.
| Initiated | Event | Class | Status | Product | Reason | Source |
|---|---|---|---|---|---|---|
| September 24, 2018 | 81092 | Class II | Completed | Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388 | The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the wave | accessdata.fda.gov |
| August 3, 2018 | 80761 | Class II | Terminated | Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388 | Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault. | accessdata.fda.gov |
| March 14, 2018 | 79611 | Class II | Terminated | PHILIPS Xper Flex Cardio Physiomonitoring System, Model Numbers: 453564241901, 453564483321, 453564621791, and 989803199561 (international only) The Xper Flex Cardio physiomonitoring system is used to facilitate invasive | The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue. | accessdata.fda.gov |
| November 20, 2017 | 80057 | Class II | Ongoing | Xper Flex Cardio Physiomonitoring System The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such. The Xper Flex C | While in Full Disclosure playback, a user may inadvertently close the Full Disclosure Control Window using the ESC key function, instead of pressing the X icon in the upper right corner of this window. | accessdata.fda.gov |
| August 18, 2017 | 78585 | Class II | Terminated | Xper Flex Cardio Physiomonitoring system. Used to facilitate invasive investigations of heart and vascular disease when non-invasive indicators warrant such. | Intermittent communication between the host system and the FlexCardio | accessdata.fda.gov |
| August 11, 2017 | 78465 | Class II | Terminated | Xper Flex Cardio PhysioMonitoring system The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular disease when non-invasive indicators warrant such. Its users, respo | Intermittent communication between the host system and the Flex Cardio | accessdata.fda.gov |
| October 19, 2015 | 73196 | Class II | Terminated | The system provides the ability to transmit patient data files for storage, viewing and analysis at distributed locations via the intranet or Internet, or may function as a stand-alone device. | Display of SpO2 and Non-Invasive Blood Pressure (NIBP) Measurements May Freeze | accessdata.fda.gov |
| October 22, 2014 | 69587 | Class II | Terminated | Philips Expression MR200 MRI Patient Monitoring System. | The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock. | accessdata.fda.gov |
| September 14, 2014 | 74296 | Class II | Terminated | Philips Expression Information Portal (IP5) Model 865471. The IP5 is intended to be used as a secondary display control unit for the Expression MRI Patient Monitoring System and the Expression MR200 MRI Patient Monitorin | Frozen Display Numerics and Disabled Menu Keys after extended run time. This customer notification was sent August 22, 2014. | accessdata.fda.gov |
| September 8, 2014 | 69192 | Class II | Terminated | Philips Expression MR200 MRI Patient Monitoring System Model 866120 | During production, excess metal (aluminum) debris was found stuck to the cling plastic that covers the device's pole assembly. The metallic debris could affect system performance due to the potential for a failure to power up or loss of power. This could result in the loss of monitoring and a delay in diagnosis of the patient. | accessdata.fda.gov |
| September 5, 2014 | 69179 | Class II | Terminated | Philips Expression MR200 MRI Patient Monitoring System | Failure to produce the Non-Invasive Blood Pressure (NIBP) Measurement. | accessdata.fda.gov |
| June 30, 2014 | 68956 | Class II | Terminated | Philips IP5 Information Portal The Expression Information Portal (Model IP5), a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professionals to remote | The wireless function of the IP5 radio could be interrupted causing loss of wireless communication between the IP5 and host monitoring system. | accessdata.fda.gov |
| December 4, 2013 | 67387 | Class II | Terminated | Invivo Corporation Expression Information Portal (Model IP5) is intended to be used as a secondary display control unit, where the device monitors the vital signs of a patient undergoing an MRI procedure from the MR cont | Invivo Corporation located at 12151 Research Parkway, Orlando, FL 32826 is recalling the Expression Information Portal (Model IP5). Upon power up, the IP5 may fail to produce audio. | accessdata.fda.gov |
| October 3, 2013 | 67424 | Class II | Terminated | Invivo Corporation Expression IP5 Information Portal Product Usage: The Expression Information Portal (Model IP5) a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained | Invivo Corporation located at 12151 Research Parkway, Orlando, FL 32826 is recalling their Expression Information Portal (Model IP5). Three different IP5 devices lost audio during use. | accessdata.fda.gov |
| October 15, 2012 | 63643 | Class II | Terminated | Invivo Corporation Expression MRI Patient Monitoring System. The Expression MRI Patient Monitoring System is intended for use by healthcare professionals to monitor vital signs for patients undergoing MRI procedures. The | Invivo Corporation located at 12151 Research Parkway, Orlando, FL 32826 is recalling the Expression MRI Patient Monitoring System because during transport of the systems cart with a docked Display Controller Unit (DCU), a caster (wheel) may become detached. If this occurs, the cart can become unbalanced and may fall. | accessdata.fda.gov |
| September 23, 2011 | 60025 | Not yet classified | Terminated | Radio used in the Precess MRI Patient Monitoring System. 3160 (Precess) and HE97B, 94044, 94045 (radio and radio kits). Label located on the back of the WPU. Product is labeled in part: "***Invivo Corporation 12601 Resea | Philips Invivo has determined that it is possible for the system's alarm limits and other device settings to spontaneously change without user intervention. This problem is associated with intermittent and frequent loss of communication between the wireless Processing Unit and the Display Controller Unit. | accessdata.fda.gov |
| August 30, 2010 | 58984 | Not yet classified | Terminated | "***Assembled for Invivo Corporation 12501 Research Parkway Orlando FL 32826***Made in U.S.A.***MRI Patient Monitoring System Model 865214***Use only with Model 989803169201 Power Adapter***REF 453564155341***SN US114016 | Philips Invivo Corporation of Orlando, FL is recalling their Expression MRI Patient Monitoring System. During transport of the system's cart with a docked Display Controller Unit (DCU), a front caster (wheel) may become loose and fall off causing the cart to become unbalanced and fall. | accessdata.fda.gov |
| August 13, 2008 | 50790 | Not yet classified | Terminated | 3.7 V Li-Polymer Battery, used in Precess MRI Compatible Patient Monitoring System and Achieva, Intera and Panorama 1.0T. REF 9065 or REF 9067. Invivo Corporation, Orlando, FL 32826 These rechargeable batteries are used | Battery Failure: The Invivo Lithium Polymer Wireless Module Battery Part Number 9065 and the Lithium Polymer Fiber Optic Module Battery Part Number 9067 is being recalled because the battery, itself, when dropped from 3 feet/1 meter onto a hard surface can produce a flame and then emit smoke. The firms investigation shows that the battery failure does not oc | accessdata.fda.gov |
| September 16, 2005 | 33936 | Not yet classified | Terminated | Product is a CALYSTO Series IV Physio-monitoring & Information System model # M-S0053. | Customer complained that a boom monitor came off the boom. No patient or user injury was associated with the incident. | accessdata.fda.gov |
Compare with other makers
- All makers of Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
- All makers of Recorder, Paper Chart
- All makers of Electrode, Electrocardiograph
Updated September 30, 2021. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.