Skip to the page

Data / Medical devices / France / Imactis, SAS

Imactis, SAS: FDA-registered device maker in Saint Martin D'Heres, France, 1 510(k) clearance

Imactis, SAS is registered with the FDA (registration 3015060224) in Saint Martin D'Heres, France. It lists 2 device types. The FDA files 1 510(k) clearance under its listings, the newest dated April 24, 2018. 3 recall events name it as the recalling firm. Registered through 2027. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3015060224
Address
20 rue du Tour de l'eau, Saint Martin D'Heres, France, 38400
Owner or operator
GE HealthCare Technologies Inc. (9912004)
Roles
Manufacturer
First device listed
December 17, 2024
Registered through
2027
Brand names listed
IMACTIS CT-Navigation system, Imactis Mobile Cart, USMAN MAGNETIC NAVIGATION STATION

Devices it lists

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
April 24, 2018K162314IMACTIS CT-Navigation systemJAKSubstantially EquivalentImactis, Sasaccessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
February 11, 202187391Class IITerminatedImpactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs oaccessdata.fda.gov
August 24, 202085258Class IIOngoingImactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the ImactisCT-Navigation Workstation (REF J02000/J00180)The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.accessdata.fda.gov
August 1, 201984460Class IIOngoingImactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameteThe firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a levelaccessdata.fda.gov

Compare with other makers

Updated April 24, 2018. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.