Data / Medical devices / United States / Illumina, Inc.

Illumina, Inc.: FDA-registered device maker in San Diego, CA, United States, 3 510(k) clearances and 1 PMA approval

Illumina, Inc. is registered with the FDA (registration 3003218906) in San Diego, CA, United States. It lists 6 device types. The FDA files 3 510(k) clearances and 1 PMA approval under its listings, the newest dated August 21, 2024. 9 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3003218906
Address
5200 Illumina Way, San Diego, CA, United States, 92122
Owner or operator
Illumina, Inc. (10030352)
Roles
Complaint file establishment, Manufacturer, Specification developer
First device listed
January 10, 2014
Registered through
2026
Brand names listed
DRAGEN™ for Illumina DNA Prep with Enrichment Dx Application for NextSeq 550Dx, DRAGEN™ for Illumina DNA Prep with Enrichment Dx Application for NovaSeq 6000Dx, Illumina DRAGEN Server for NextSeq 550Dx, Illumina® DNA Prep with Enrichment Dx with UD Indexes Set A (16 samples), Illumina® DNA Prep with Enrichment Dx with UD Indexes Set A (96 samples), Illumina® DNA Prep with Enrichment Dx with UD Indexes Set B (16 samples), Illumina® DNA Prep with Enrichment Dx with UD Indexes Set B (96 samples), Local Run Manager CF 139 Variant Analysis Module and more

Devices it lists

DeviceProduct codeClassRoleSource
Dna And Rna Oncology ControlsSCU2Complaint file establishment, Manufactureraccessdata.fda.gov
High Throughput Dna Sequence AnalyzerPFF2Complaint file establishment, Manufacturer, Specification developeraccessdata.fda.gov
Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection SystemPQP3Complaint file establishment, Manufactureraccessdata.fda.gov
Reagents For Molecular Diagnostic Test SystemsPFT1Complaint file establishment, Manufacturer, Specification developeraccessdata.fda.gov
System, Cystic Fibrosis Transmembrane Conductance Regulator Gene, Mutations & Variants Panel Sequencing DetectionPFR2Complaint file establishment, Manufactureraccessdata.fda.gov
System, Cystic Fibrosis Transmembrane Conductance Regulator Gene, Variant Gene Sequence DetectionPFS2Complaint file establishment, Manufactureraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
November 19, 2013K132750ILLUMINA MISEQDX CYSTIC FIBROSIS CLINICAL SEQUENCING ASSAYPFSSubstantially EquivalentIllumina, Inc.accessdata.fda.gov
November 19, 2013K124006ILLUMINA MISEQDX CYSTIC FIBROSIS 139-VARIANT ASSAYPFRSubstantially EquivalentIllumina, Inc.accessdata.fda.gov
November 19, 2013DEN130011ILLUMINA MISEQDX PLATFORMPFFUnknownIllumina, Inc.accessdata.fda.gov

PMA approvals

Premarket approvals this establishment lists, with the supplements the FDA has decided since.

ApprovedPMA numberDeviceProduct codeSupplementsLatest decisionSource
August 21, 2024P230011TruSight Oncology ComprehensivePQP8September 4, 2026accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
October 28, 202597913Class IIOngoingReagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75 Cycles, REF: 20031098; NextSeq 550DAn instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance expaccessdata.fda.gov
September 17, 202499585Class IIOngoingNovaSeq 6000Dx including DRAGEN Server, Catalog Number: 20068232; Illumina DRAGEN Server for NextSeq 550Dx, Catalog Number: 20086130Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic assays has a server with a security vulnerability.accessdata.fda.gov
May 8, 202499957Not yet classifiedOpen, ClassifiedNextseq 550Dx Instrument REF: 20005715, FOR IN VITRO DIAGNOSTIC USE(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.accessdata.fda.gov
April 5, 202392233Class IICompletedNextSeq 550Dx Instrument, Catalog # 20005715, contaning Universal Copy Service softwareCybersecurity vulnerability concerning the software used for sequencing instruments.accessdata.fda.gov
May 3, 202290160Class IICompletedillumina REF DX-410-1001 Model: MiSeq Dx illumina REF 15036706 Model: MiSeq Dx illumina REF 20014053 Model: MiSeq Dxcybersecurity vulnerabilityaccessdata.fda.gov
February 22, 202289699Class IICompletedIllumina NextSeq 550Dx, REF: 20005715, CE, IVDThere is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.accessdata.fda.gov
November 13, 201469766Class IITerminatedIllumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.accessdata.fda.gov
November 6, 201469715Class IIITerminatedIllumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument.accessdata.fda.gov
July 28, 201468959Class IITerminatedIllumina Worklist Manager (IWM) (software v1.0.15), a component of Illumina MiSeqDx Platform. Product Usage: The Illumina MiSeqDx is a sequencing instrument that measures fluorescence signals of labeled nucleotides throuA software component of the MiSeqDx instrument, called the Illumina Worklist Manager (IWM) v1.0.15, fails to perform as intendedaccessdata.fda.gov

Compare with other makers

Updated September 4, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.