Skip to the page

Data / Medical devices / United States / Icu Medical, Inc.

Icu Medical, Inc.: FDA-registered device maker in Lake Forest, IL, United States, 114 510(k) clearances

Icu Medical, Inc. is registered with the FDA (registration 3013319212) in Lake Forest, IL, United States. It lists 32 device types. The FDA files 114 510(k) clearances under its listings, the newest dated April 2, 2026. 88 recall events name it as the recalling firm (13 Class I). Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3013319212
Address
600 N Field Dr, Lake Forest, IL, United States, 60045
Owner or operator
ICU MEDICAL, INC. (2025816)
Roles
Complaint file establishment, Specification developer
First device listed
August 4, 1999
Registered through
2026
Brand names listed
ADAPTOR, CATHETER, Accessories, pump, infusion, Additive Hinge Cap, Administration Set, Administration Sets: Primary Sets, Extension Sets, Burette Sets, Blood Sets, Advanced Sensor Catheter, Blood Set, Burette Set and more

Devices it lists

DeviceProduct codeClassRoleSource
Accessories, Pump, InfusionMRZ2Complaint file establishment, Specification developeraccessdata.fda.gov
Adaptor, CatheterGCE1Complaint file establishmentaccessdata.fda.gov
Aid, TransferIKX1Complaint file establishment, Specification developeraccessdata.fda.gov
Blood Pressure CuffDXQ2Complaint file establishmentaccessdata.fda.gov
Cap, Device DisinfectantQBP2Complaint file establishment, Specification developeraccessdata.fda.gov
Catheter, Flow DirectedDYG2Complaint file establishmentaccessdata.fda.gov
Catheter, Intravascular, DiagnosticDQO2Complaint file establishmentaccessdata.fda.gov
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 DaysFOZ2Complaint file establishmentaccessdata.fda.gov
Catheter, IrrigationGBX1Complaint file establishment, Specification developeraccessdata.fda.gov
Catheter, Oximeter, Fiber-OpticDQE2Complaint file establishmentaccessdata.fda.gov
Closed Antineoplastic And Hazardous Drug Reconstitution And Transfer SystemONB2Complaint file establishmentaccessdata.fda.gov
Container, I.V.KPE2Complaint file establishment, Specification developeraccessdata.fda.gov
Device, Warming. Blood And PlasmaKZL2Complaint file establishment, Specification developeraccessdata.fda.gov
Endoscopic Central Control UnitODA2Complaint file establishment, Specification developeraccessdata.fda.gov
Hemodialysis Catheter Luer End CapPEH2Complaint file establishmentaccessdata.fda.gov
Infusion Safety Management SoftwarePHC2Complaint file establishment, Specification developeraccessdata.fda.gov
Medical Device Data SystemOUG1Complaint file establishment, Specification developeraccessdata.fda.gov
OximeterDQA2Complaint file establishmentaccessdata.fda.gov
Probe, ThermodilutionKRB2Complaint file establishmentaccessdata.fda.gov
Pump, InfusionFRN2Complaint file establishment, Specification developeraccessdata.fda.gov
Pump, Infusion, PcaMEA2Complaint file establishment, Specification developeraccessdata.fda.gov
Set, Administration, IntravascularFPA2Complaint file establishment, Specification developeraccessdata.fda.gov
Set, I.V. Fluid TransferLHI2Complaint file establishment, Specification developeraccessdata.fda.gov
Stand, InfusionFOX1Complaint file establishment, Specification developeraccessdata.fda.gov
System, Thermal RegulatingDWJ2Complaint file establishment, Specification developeraccessdata.fda.gov
System/Device, Pharmacy CompoundingNEP2Complaint file establishmentaccessdata.fda.gov
Transducer, Blood-Pressure, ExtravascularDRS2Complaint file establishmentaccessdata.fda.gov
Transmitters And Receivers, Physiological Signal, RadiofrequencyDRG2Complaint file establishmentaccessdata.fda.gov
Tube, Aspirating, Flexible, ConnectingBYY2Complaint file establishment, Specification developeraccessdata.fda.gov
Tubing, Fluid DeliveryFPK2Complaint file establishment, Specification developeraccessdata.fda.gov
Warmer, Blood, Non-Electromagnetic RadiationBSB2Complaint file establishment, Specification developeraccessdata.fda.gov
Warmer, Thermal, Infusion FluidLGZ2Complaint file establishment, Specification developeraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance. Newest 100 of 114.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
April 2, 2026K252130LifeShield Infusion Safety Software SuitePHCSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
April 2, 2025K242117LifeShield Infusion Safety Software SuitePHCSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
April 2, 2025K242115Plum Duo™ Precision IV PumpFRNSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
April 2, 2025K242114Plum Solo™ Precision IV PumpFRNSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
January 22, 2025K243985Rio™ Drug Reconstitution Transfer DeviceLHISubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
August 24, 2023K223607Plum Duo™ Infusion SystemFRNSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
August 24, 2023K223606LifeShield™ Infusion Safety Software SuitePHCSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
April 20, 2020K192154Rio Vial-to-Bag Drug Reconstitution DeviceLHISubstantially EquivalentIcu Medicalaccessdata.fda.gov
March 6, 2020K190918SwabTip Male Disinfectant CapQBPSubstantially EquivalentIcu Medicalaccessdata.fda.gov
August 10, 2018K170110Diana Medication Transfer SetONBSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
May 25, 2018K180111ClearGuard HD Antimicrobial Barrier CapPEHSubstantially EquivalentPursuit Vascular, Inc.accessdata.fda.gov
August 25, 2017K171346Sapphire SetsMRZSubstantially EquivalentIcu Medicalaccessdata.fda.gov
March 28, 2017K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleFRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 6, 2017K161036Hospira Extension Set, Hospira Primary SetFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
December 6, 2016K160492Hospira Sapphire SetsMRZSubstantially EquivalentHospira, Inc.accessdata.fda.gov
June 1, 2016K160870Hospira Administration SetsFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
May 5, 2016K152006Cogent Hemodynamic Monitoring SystemDXQSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
April 8, 2016K143612LifeCare PCA Infusion SystemMEASubstantially EquivalentHospira, Inc.accessdata.fda.gov
March 11, 2016K16037420G x 1.25 NovaCath Secure IV Catheter System, 24G x 0.56 NovaCath Secure IV Catheter SystemFOZSubstantially EquivalentTangent Medical Technologies, Inc.accessdata.fda.gov
January 19, 2016K151969Hospira Extension SetsFPASubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
March 4, 2015K143087Hospira Blood SetFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
February 25, 2015K143015Hospira Primary SetsFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 16, 2015K141789PLUM 360 INFUSION SYSTEM WITH HOSPIRA MEDNET, SMART CARD PLUG N PLAY CE 3.0 MODULE FOR PLUM A+ INFUSION SYSTEM UPGRADEFRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 15, 2015K142622IV AdministrationFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 9, 2015K142974Extension SetsFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
December 11, 2014K141389SAPPHIRE INFUSION PUMPFRNSubstantially EquivalentQ Core Medical , Ltd.accessdata.fda.gov
December 8, 2014K142433Hospira Extension SetFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
November 25, 2014K142367Hospira Primary SetsFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
October 28, 2014K141686EQUATOR CONVECTIVE WARMER, SNUGGLE WARM ADULT FULL BODY CONVECTIVE WARMING BLANKET, SNUGGLE WARM PEDIATRICFULL BODYDWJSubstantially EquivalentSmith Medical Asd, Inc.accessdata.fda.gov
August 5, 2014K141834SAPPHIRE ADMINISTRATION SETFPASubstantially EquivalentQ Core Medical , Ltd.accessdata.fda.gov
July 25, 2014K141193MEDFUSION 4000 PUMP DEVICE SPECIFIC REPORTS SOLFWARE FOR PHARMGUARD SERVER VERSION 2.0PHCSubstantially EquivalentSmiths Medical Asd, Inc.accessdata.fda.gov
July 8, 2014K141102HOSPIRA MEDNET MEDICATION MANAGEMENT SUITEFRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
July 2, 2014K140129PICCOX PERIPHERALLY INSERTED PRESSURE INJECTABLE, CENTRAL VENOUS OXIMETRY CATHETERDQESubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
January 24, 2014K133602ADD-VANTAGE ADDAPTORLHISubstantially EquivalentHospira, Inc.accessdata.fda.gov
December 24, 2013K130975SWABCAP AND SWABFLUSHQBPSubstantially EquivalentExcelsior Medical Corp.accessdata.fda.gov
December 9, 2013K130394CADD -SOLIS AMBULATORY INFUSION PUMP, VERSION 3.0MEASubstantially EquivalentSmiths Medical Asd, Inc.accessdata.fda.gov
October 17, 2013K123049SAPPHIRE INFUSION PUMPFRNSubstantially EquivalentQ Core Medical , Ltd.accessdata.fda.gov
July 26, 2013K131549CHEMOLOCK, CHEMOLOCK PORTONBSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
July 20, 2012K120839NOVACATH SECURE IV CATHETER SYSTEM-18GFOZSubstantially EquivalentTangent Medical Technologies, Inc.accessdata.fda.gov
March 5, 2012K110901SYMBIQ INFUSION AND ADMINSTRATION SETFRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 12, 2012K113683LIFESHIELD VISION INFUSION SET WITH PRE-PIERCED RESEALFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 7, 2011K103224HOSPIRA PLUM INFUSION SET; HOSPIRA INFUSION SET WITH YELLOW STRIPE TUBINGFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
December 10, 2010K103344INFUSION SETSFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
August 11, 2010K101677HOSPIRA INFUSION BLOOD SETSFPASubstantially EquivalentHospira, Inc.accessdata.fda.gov
September 18, 2009K091268HOSPIRA LATEX-FREE CRITICAL CARE AND ADVANCED SENSOR CATHETERSDQESubstantially EquivalentHospira, Inc.accessdata.fda.gov
June 12, 2009K091181LOCKBOX FOR USE WITH MEDFUSION 3000 SERIES PUMPSMRZSubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
March 13, 2009K090610HOSPIRA VITAL SIGNS WIRELESS MONITORING SYSTEMDRGSubstantially EquivalentHospira, Inc.accessdata.fda.gov
December 17, 2008K082783CADD -SOLIS MEDICATION SAFETY SOFTWARE: ADMINISTRATOR AND POINT-OF-CAREMRZSubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
September 11, 2008K080579EFFECTIV (TM) CAPQBPSubstantially EquivalentHospira, Inc.accessdata.fda.gov
July 18, 2008K081156CADD MEDICATION CASSETTE RESERVOIRFPASubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
June 2, 2008K072080NORMOFLO IRRIGATION FLUID WARMER, MODELS H-1100, H-1129, NORMOFLO IRRIGATION WARMING SET, MODELS, IR-40, IR-500, IR-600,LGZSubstantially EquivalentSmiths Medical Asd, Inc.accessdata.fda.gov
April 25, 2008K080743LOCKBOX FOR USE WITH CADD SOLIS INFUSION PUMPFRNSubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
March 7, 2008K072144CADD -SOLIS AMBULATORY INFUSION PUMP AND CADD -SOLIS MEDICATION SAFETY SOFTWARE - ADMINISTRATORMEASubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
April 24, 2007K070398HOSPIRA PLUM A+ INFUSION PUMP, MODEL 12391FRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
August 4, 2006K061450HOSPIRA LATEX-FREE ADVANCED SENSOR CATHETERS; HOSPIRA LATEX-FREE CRITICAL CARE CATHETERSDQESubstantially EquivalentHospira, Inc.accessdata.fda.gov
August 1, 2006K061513LEVEL 1 SNUGGLE WARM PEDIATRIC AND LARGE PEDIATRIC UNDERBODY BLANKETDWJSubstantially EquivalentSmiths Medical Asd, Inc.accessdata.fda.gov
April 21, 2006K060806HOSPIRA GEMSTAR INFUSION PUMP SYSTEM, 7 THERAPY, MODEL 13087; 6 THERAPY, MODEL 13086; PAIN MANAGEMENT, MODEL 13088FRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
February 17, 2006K052828DISPOSABLE TRANSPAC III INTEGRATED TRANSDUCER (IT)DRSSubstantially EquivalentHospira, Inc.accessdata.fda.gov
December 22, 2005K052865HOSPIRA CATH LAB CONVENIENCE KITSDQOSubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
November 17, 2005K051568CADD-MS 3 AMBULATORY INFUSION PUMPFRNSubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
August 24, 2005K052052HOSPIRA PLUM A+ INFUSION PUMP SYSTEM V11.5, HOSPIRA PLUM A+ 3 INFUSION PUMP SYSTEM, V11.5FRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
January 4, 2005K042432PHARMGUARD TOOLBOX, MODEL G6000782FRNSubstantially EquivalentMedex, Inc.accessdata.fda.gov
October 18, 2004K042800LIFECARE PCA INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWAREMEASubstantially EquivalentHospira, Inc.accessdata.fda.gov
September 1, 2004K042182Y CLAVE CONNECTOR WITH INTEGRATED CHECK VALVE, MODEL XI-3432FPASubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
August 24, 2004K042081PLUM A+ INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE , AND PLUM A+3 INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWAREFRNSubstantially EquivalentHospira, Inc.accessdata.fda.gov
May 19, 2004K040636CADD ADMINISTRATION SET WITH FLOW STOP, MEDICATION CASSETTE RESERVOIR WITH FLOW STOP AND EXTENSION SETFPASubstantially EquivalentSmiths Medical MD, Inc.accessdata.fda.gov
April 30, 2004K040899MEDEX 3000 SERIES MR SYRINGE INFUSION PUMPFRNSubstantially EquivalentMedex, Inc.accessdata.fda.gov
April 27, 2004K040632SNUGGLE WARM 4000 CONVECTIVE WARMING SYSTEM, MODEL SW4000DWJSubstantially EquivalentSmiths Medical Asd, Inc.accessdata.fda.gov
December 4, 2003K033576LIFESHIELD LATEX-FREE PRIMARY IV PUMP SET DISTAL MICROBORE PATIENT LINE, CONVERTIBLE PIN, 72 INCH, WITH 2 PRESSUREFPASubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
May 23, 2003K031361CADD HIGH-VOLUME ADMINISTRATION SETS WITH FLOWSTOPFPASubstantially EquivalentDeltec, Inc.accessdata.fda.gov
December 18, 2001K011907SNUGGLE WARM 4000 CONVECTIVE WARMING SYSTEMDWJSubstantially EquivalentSims Level 1, Inc.accessdata.fda.gov
June 5, 2001K011442ABBOTT PLUM A+INFUSION PUMP, MODEL 11971FRNSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
April 6, 2001K010924MODIFICATION TO PLUM XL INFUSION PUMPFRNSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
August 9, 2000K001764HOTLINE 2 FLUID WARMERLGZSubstantially EquivalentSims Level 1, Inc.accessdata.fda.gov
June 8, 2000K000842CADD-PRIZM PCS II MODEL 6101 AMBULATORY INFUSION SYSTEMFRNSubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
April 27, 2000K000942CADD CHECKVALVE ADMINISTRATION SET, MODEL 21-7090FPASubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
June 15, 1999K991599CADD ADMINISTRATION SETFPASubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
June 9, 1999K991301CADD-PRIZM HIGH VOLUME ADMINISTRATION SET WITH 1.2 U FILTER AND ADD-ON ANTI-SIPHON VALVE, CADD ADMINISTRATION SET WITH 1FPASubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
June 9, 1999K991255CADD-TPN ADMINISTRATION SET, MODEL 21-7071FPASubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
February 12, 1999K990083CADD ADMINISTRATION SET, CADD 250ML FLEXIBLE MEDICATION RESERVOIR, MODIFIED SECURIT SHELL, SECURITY SHELL ADAPTERFPASubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
January 12, 1999K982159ABBOTT PLUM A+ INFUSION PUMPFRNSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
November 4, 1998K982838CADD-LEGACY 1 AMBULATORY INFUSION SYSTEM MODEL NUMBER 6400FRNSubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
November 2, 1998K982839CADD-LEGACY PCA AMBULATORY INFUSION SYSTEM, MODEL 6300MEASubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
November 2, 1998K982836CADD-LEGACY PLUS AMBULATORY INFUSION SYSTEMFRNSubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
January 9, 1998K973917CADD-DIPLOMAT PC COMMUNICATIONS SYSTEM AND THE CADD- AMBASSADOR PUMP COMMUNICATIONS SYSTEMFRNSubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
June 27, 1997K971293LIFESHIELD PRIMARY I.V. SET WITH BACKCHECK VALVEFPASubstantially EquivalentAbbott Mfg., Inc.accessdata.fda.gov
June 24, 1997K970855CLAVE CONNECTORFPASubstantially EquivalentIcu Medical, Inc.accessdata.fda.gov
February 10, 1997K961552CVP POLYURETHANE CATHETERFOZSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
August 23, 1996K960826CADD-PRIZM MODEL 6100 PUMPFRNSubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
June 4, 1996K960957EXTERNAL POWER SOURCE (EPS) SYSTEM, POWER PACKFRNSubstantially EquivalentSims Deltec, Inc.accessdata.fda.gov
January 18, 1996K952799ABBOTT XL3M INFUSION PUMPFRNSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
February 21, 1995K943310MODEL 6100 AMBULATORY INFUSION SYSTEMMEASubstantially Equivalent - Subject to Tracking Reg.Pharmacia Deltec, Inc.accessdata.fda.gov
October 3, 1994K941214LIFESHIELD EXTENSION SETFPKSubstantially EquivalentAbbott Mfg., Inc.accessdata.fda.gov
June 13, 1994K941254DYE MANAGEMENT SYSTEMDQOSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
February 9, 1994K933390REMOTE RESERVOIR ADAPTER(TM) CASSETTE WITH BAGFPASubstantially EquivalentPharmacia Deltec, Inc.accessdata.fda.gov
April 6, 1992K914988500 INFUSERFRNSubstantially EquivalentExcel Medical, Inc.accessdata.fda.gov
October 17, 1991K911383HOTLINE FLUID WARMERLGZSubstantially EquivalentLevel 1 Technologies, Inc.accessdata.fda.gov
June 11, 1991K911791CDD-PCA(TM) MODEL 5800R AMBULATORY INFUSION PUMPMEASubstantially EquivalentPharmacia Deltec, Inc.accessdata.fda.gov
May 28, 1991K911401ABBOTT LIFECARE 5000 PUMPFRNSubstantially EquivalentAbbott Laboratoriesaccessdata.fda.gov
September 14, 1990K902374CADD-TPN MODEL 5700 AMBULATORY INFUSION SYSTEMFRNSubstantially EquivalentPharmacia Deltec, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
August 20, 202699648Class IIOngoingCADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-2Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.accessdata.fda.gov
June 10, 202699192Class IIOngoingICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pumpIn a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.accessdata.fda.gov
April 29, 202698872Class IIOngoingPlum Solo Precision IV Pump, 40001-0401Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.accessdata.fda.gov
December 15, 202598129Class IIOngoingPlum Duo Infusion System, List Number 40002-04-01ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.accessdata.fda.gov
December 15, 202598128Class IIOngoingICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Haaccessdata.fda.gov
November 24, 202598022Class IIOngoingLifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-deliveraccessdata.fda.gov
November 24, 202598000Class IIOngoingLifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.accessdata.fda.gov
July 11, 202597186Class IOngoingPlum Duo Infusion System, List Number: 400020401ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.accessdata.fda.gov
May 2, 202596771Class IIOngoingLifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for accessdata.fda.gov
April 22, 202596777Class IIOngoingICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001, and 3) 400021003; infusion pumpInternal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight oaccessdata.fda.gov
October 22, 202495586Class IOngoingAllegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infusion Systems, Pump List Numbers: 11005, 1197ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum Infusion Systems. Preliminary reports suggest that thesaccessdata.fda.gov
May 7, 202494690Class IOngoingReplacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following descriptions: Bomba de infusin Plum 360 comDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.accessdata.fda.gov
July 13, 202392630Class IIOngoingPlum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453, 300100481, 300100483, 300100613, 300101013, 300101113, 300101114, 300101313, 3001Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.accessdata.fda.gov
March 22, 202391874Class IOngoingReplacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.accessdata.fda.gov
July 29, 201983549Class IITerminatedInfusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as High-Pressure Filter PRIMARY I.V. PLUMSET Convetible Pin, 104 Inch with Orange Polyethylene-Lined*/Light ResistaPotential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.accessdata.fda.gov
November 8, 201881615Class IITerminatedSAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter, PAV, 123 Inch, List Number 16373-02 Sapphire Epidural Sets are indicated for the delivery of fluids from a container to a patient's epidural space. Sets There have been reports of distal occlusion alarms during the use of SAPPHIRE Epidural Set Yellow-Striped administration sets.accessdata.fda.gov
August 1, 201883480Class IITerminatedPlum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.accessdata.fda.gov
October 30, 201778418Class IITerminatedICU Mednet(TM) Medication Management Suite 6.1 and 6.21, Product Code/List Numbers: (a) 16037-64-01; MedNet 6.1 b) 16037-64-02; MedNet 6.1 c) 16037-64-03; MedNet 6.1 d) 16037-64-04; MedNet 6.21 The ICU Medical MedNet MedIssue 1: The MedNet Meds 6.1 and 6.21 programs, under certain conditions, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a medication may be piggybacked and (b) whether an infusion may be interrupted by a piggyback. This only occurs when: 1. Medication entries are built with the same accessdata.fda.gov
October 30, 201778312Class IITerminatedPlum 360 Infusion System, List number 30010.(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the battery is discharged or is disconnected; and (2) when the user accesses the Preventive Maintenance Screen in Service/Biomed Mode with a Totalaccessdata.fda.gov
June 15, 201777624Class IITerminatedPlum 360 Infusion SystemPotential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an infusion.accessdata.fda.gov
December 30, 201676253Class IITerminatedPlum 360 Infusion Pump, Software Version 15.02. The infusion pump is capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusionsDepleted Battery alarm shows instead of Replace Battery. On battery power, ongoing therapy stops, alarms show and sound, pump shuts down after 3 minutes. On AC power, the pump reboots (takes about 25 seconds). Therapy can resume. AC power interruption of 3-7 seconds prompts an incorrect E323 alarm, ongoing therapy stop, and pump reboot. Longer power loss resaccessdata.fda.gov
July 28, 201674903Class IITerminatedHospira MedNet Medication Management Suite software, version 6.1, List Numbers 16037-64-02 and 16037-64-03, in combination with the SapphirePlus 13.1x infusion pump Hospira MedNet Medication Management Suite is intended Hospira MedNet 6.1 software, in combination with the SapphirePlus 13.1x infusion pump, may result in incorrect bolus amount calculations for drugs whose concentrations are listed in Million Units.accessdata.fda.gov
March 25, 201673799Class IITerminatedThe Plum 360 is a large volume infuser capable of delivering fluids for a variety of therapies such as parenteral, enteral, or epidural infusions. The Plum 360 infuser can deliver fluids over a broad range of infusion raFirm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.accessdata.fda.gov
December 22, 201573104Class IITerminatedThe Symbiq infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. It is intended primarily for use in the hospital setting and can be used in other acute and non-acPotential for the Symbiq pole clamp assemblies to be missing the rubber stop pad.accessdata.fda.gov
July 2, 201469336Class ITerminatedThe Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.accessdata.fda.gov
June 11, 201468672Class IITerminatedPlum A+ Infusion System with Hospira MedNet List 20792-04-25, Pole Clamp Assembly Part Number 820-95443-008. A cassette based multi-function infusion system. Each system includes a pumping module and an assortment of disPlum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to tighten and loosen the grip of the pole clamp to the pole.accessdata.fda.gov
June 6, 201468658Class IITerminatedThe Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and HospHospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" oaccessdata.fda.gov
March 25, 201468129Class IITerminatedLATEX-FREE; EXTENSION SET; For Bladder Irrigation or Urinary Drainage, Compatible with T-U-R FLOW-POUCH RESERVOIR, list 6542, 71 inch (180 cm) Nominal Length; List No. 4693-01; single use; Hospira Product Usage: Bladder The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral Resection; List No. 06544-01 Cystoscopy / Irrigation Set; List No. 06599-01, Large Bore Y-Irrigation Set (Lataccessdata.fda.gov
February 25, 201467722Class IITerminatedLifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series InfHospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenaccessdata.fda.gov
February 18, 201467802Class ITerminatedGemStar Docking Station, used with GemStar infusion pump, in combination with other GemStar accessories such as the GemStar Lock Box and patient bolus cord. The docking station is also used for hosting the GemStar pump fThere are two situations that may occur when using the GemStar Docking Station, List Number 13075-XX-XX, in conjunction with the GemStar infusion pump: 1) when the Docking Station is used in conjunction with a GemStar Phase 3 pump (List 13000-XX, 13100-XX, or 13150-XX) the potential exists for the GemStar Phase 3 pump to fail to power up while connected to taccessdata.fda.gov
February 3, 201468047Class ITerminatedThe GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an outpThe GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.accessdata.fda.gov
January 31, 201468200Class IITerminatedGemStar infusion pump, List numbers 13000 and 13100, GemStar Infusion Pump (pain management) List number 13150. Used for intravenous, arterial, subcutaneous, short-term epidural infusion and parenteral administration of The connection between the beeper subassembly and the pump may fail. The GemStar infusion pump will identify this failure during the "self-test" while powering up which will result in a Beeper Error ("code 10/001/000"). This Beeper Error (code 10/001/000) is a service alarm that places the pump in an inoperable mode and requires service before it can be retuaccessdata.fda.gov
January 24, 201467453Class IITerminatedLifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids by intravenous (I.V.) administration.There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a caccessdata.fda.gov
October 31, 201367591Class ITerminatedAbbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers.Hospira has received customer reports of broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.accessdata.fda.gov
October 25, 201367038Class IITerminated1) List number 126970465 labeled in part: 12697-65 LifeShield LATEX-FREE HEMA; Y-TYPE BLOOD SET, Nonvented, 88 Inch with 210 Micron Filter Pump, Prepierced Injection Site and Secure Lock; Made in Costa Rica; Product InquSeveral lots of the LifeShield Latex-Free HEMA Y-Type Blood Set IV Administration Set were assembled incorrectly.accessdata.fda.gov
September 10, 201366448Class IITerminatedHospira RECEPTAL CANISTERS: 1L(1000mL), 1.5L(1500mL), 2L(2000mL) Canister, ATS, 2000, with Valve 1) List Numbers 43449-01 and 43449-11; LIST NO. 43449 RECEPTAL CANISTER; CANISTER FOR REUSE, THROW AWAY LINER ONLY; DO NOT Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 1.5 L liner, 2L liner or 3L liner with a 1L canister. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent.accessdata.fda.gov
September 9, 201366372Class IITerminated1) Product labeled in part: List No. 14212-28; LIFESHIELD; PLUM Y-TYPE BLOOD SET; 200 Micron Filter, CLAVE Port, Non-Vented, 110 Inch, Non-DEHP; *** 2) Product labeled in part: List No. 14254-28; LIFESHIELD; PRIMARY PLUMIntravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.accessdata.fda.gov
August 28, 201366263Class IITerminated1) SYMBIQ One-Channel Infuser; *** 2) SYMBIQ Two-Channel Infuser; *** E09 batteries can be identified by the "E09" that is printed beneath the bar code and at the end of the serial number along the bottom of the label foHospira shipped the incorrect replaceable battery pack (Part No: HSP3566-E10) instead of the correct E09 battery pack (Part No. HSP3566-E09) for the Symbiq Single and Dual Channel Infusers.accessdata.fda.gov
July 17, 201366431Class IITerminatedProduct is labeled in part: LIST NO. 43449; RECEPTAL CANISTER; DO NOT FILL 900 ABOVE THIS LINE; Product of China; Hospira, Inc.,; Lake Forest, IL 60045 USA. Used to isolate suction waste during surgeries, or to clear patHospira has recieved reports that the vacuum needed to create suction with the 1L Receptal System cannot be achieved due to the canister and liner not properly seating.accessdata.fda.gov
June 18, 201365647Class IITerminatedLifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the administration of I.V.Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve. Preliminary investigation findings found that there is a potential for the ball valve to not seal the drip chamber as intended, leading to air in the line of the administration set.accessdata.fda.gov
April 1, 201365689Class ITerminatedList No. 14200-28; SECONDARY BLOOD SET; 200 Micron Filter, 36 Inch, Non-DEHP; For Blood Product Administration; Made in Costa Rica; Product Inquiries should be directed to Hospira., Inc., Lake Forrest, IL 60045 USA HospiIt was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.accessdata.fda.gov
March 21, 201364841Class IITerminatedThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor,During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.accessdata.fda.gov
March 19, 201369505Class IITerminatedGemStar infusion pump, list numbers 13000 and 13100, GemStar Infusion Pump (pain management) list number 13150. Product Usage The pump is suitable for intravenous, arterial, subcutaneous, short-term epidural infusion andSome patients are obtaining access to the dosage reset codes required to change the settings on GemStar infusion pumps and are subsequently changing the infusion rate set by their doctors. Patients are obtaining access to these codes via website sponsors by downloading the GemStar User/Service Manual featured on the sponsors' websites. The GemStar User/Serviaccessdata.fda.gov
March 18, 201364839Class ITerminatedGemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar PuBattery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.accessdata.fda.gov
March 18, 201364837Class ITerminatedThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor,Pump shutting off during use without warning.accessdata.fda.gov
March 15, 201364832Class ITerminatedThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor,The proximal and distal pressure sensor calibration can drift resulting in the pump failing the Proximal or Distal Occlusion Operational Test, as described in the GemStar Technical Service Manual, or reporting one of the following errors during device setup or infusion: 1) Cassette Check - D; 2) Cassette Check - P; 3) Proximal Occlusion; 4) Distal Occlusion;accessdata.fda.gov
March 8, 201368318Class IITerminatedThe LifeCare PCA Plus II infusion pump system is designed for safe, effective, analgesic delivery to patients through the intravenous or epidural routes in a wide range of clinical settings, including ICU, CCU, SICU, traHospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.accessdata.fda.gov
March 8, 201364230Class IITerminatedLifeCare PCA Infusion System with Hospira MedNet Software, PCA Serial List Number 20709 Allows for clinician administration or self-administration of analgesic medications.Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.accessdata.fda.gov
February 27, 201364225Class IITerminatedThe PCA 3 infusion pump allows clinicians, or patients, to (self) Administer analgesia safely and effectively within clinician programmed limits.The infusion pump has a clear plastic door that, due to its design, allows for excessive flexing. This in turn permits unauthorized access to the vial of medication.accessdata.fda.gov
February 19, 201364220Class IITerminatedMinibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device.Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.accessdata.fda.gov
February 6, 201364261Class IITerminatedPlum A+ Hyperbaric Infusion Pump. List Number 11005. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. TThe fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.accessdata.fda.gov
February 5, 201363993Class IITerminatedSymbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Product List Number 16026Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply Controller (PSC) Boards distributed prior to September 2012. On pumps where incorrect installation of components has occurred, the backup battery power discharges and subsequenaccessdata.fda.gov
February 4, 201365610Class IITerminatedThe Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneousThe devices secondary audio signal fails to activate.accessdata.fda.gov
February 4, 201363994Class IITerminatedSymbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytSymbiq Infusers have the potential to experience a white screen during titration of a Keep Vein Open (KVO) delivery. This can allow the clinician to select the "NEXT" button prior to completion of the programming. If this error occurs, a delay in therapy could occur and potentially result in significant injury.accessdata.fda.gov
February 4, 201363990Class IITerminatedSymbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytFrequent Proximal Occlusion Alarms (POAs) occurring on Symbiq devices with list numbers 16026 and 16027. When there are false alarms, the pump will continue to alarm even though there is no occlusion present. False POAs could result in a delay/interruption in therapy, and require the clinician to reset the alarm multiple times or to replace the pump.accessdata.fda.gov
February 1, 201364542Class IITerminatedThe Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self contained, mA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.accessdata.fda.gov
February 1, 201364263Class IITerminatedThe Plum A+ Hyperbaric Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+ Hyperbaric IPlum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspectiaccessdata.fda.gov
December 7, 201263772Class IITerminatedPlum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971; b) list 11973 - software version 10.3; c) list 12391 - software version The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.accessdata.fda.gov
October 26, 201263518Class IITerminatedPlum A+ Single Channel Infusion Pumps with Plug-and-Play Modules that provide wired Ethernet and wireless local area networking; Hospira, Inc., Lake Forest, IL 60045; pump list number 11971, module list numbers 12101, 12The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knaccessdata.fda.gov
August 29, 201263056Class ITerminatedSymbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; List Number 16026-01 Device is a general purpose infusion pump The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a delay in therapy or result in over delivery or under delivery of medication if the user does not confirm the programmed vaccessdata.fda.gov
August 28, 201263078Class IITerminatedEmpty Evacuated Container 250 mL and 1000 mL. Empty Evacuated containers are used for mixing compatible medications or blood collections.Embedded iron oxide glass defect which may have the potential to break off into the solution.accessdata.fda.gov
August 2, 201262909Class IITerminatedSterile Empty Vial and Injector (30mL PCA vial), Rx only, One/List No. 6021-03, Hospira, Inc., Lake Forest, IL 60045. For use with a compatible Hospira PCA pump set with injector and a compatible Hospira infusion device.Reports of leaking during filling and administration.accessdata.fda.gov
July 13, 201262582Class IITerminatedHospira LifeShield Latex-Free 100 mL Burette Set, Convertible Pin, 77 inch with 2 Prepierced Injection Sites and Option-Lok Microdrip Soluset with Calibrated Burette and Precision Drip Chamber; an Rx sterile medical deviThe float valve in the burette sticks to the burette wall and does not open or close properly.accessdata.fda.gov
April 25, 201261680Not yet classifiedTerminatedLifeShield Primary Plumset, CLAVE port, CLAVE Y-Site, 104 inch, 15 drops/mL, 48 sets per case; a sterile Rx device; Hospira Inc, Lake Forest, IL USA. Made in Costa Rica; product 120300412, list number 12030-12. Product UThe Lifeshield Primary Plumsets may cause a cassette test failure alarm when the set is loaded onto the infusion pump. A set which results in a cassette test failure alarm may result in delay in therapy, requiring medical intervention.accessdata.fda.gov
March 28, 201261472Not yet classifiedTerminatedGemStar Bolus Cord; a remote bolus cord and switch for use with GemStar 7 Therapy and Pain Management Pumps, list 13000, 13150, 13087 and 13088; Hospira, Inc., Lake Forest, IL 60045; List Number 13027; An accessory for tHospira has received customer reports of bolus delivery failures when using the GemStar bolus cord. Possible failures may be a result of cracked or damaged bolus cords which may result in an open or short circuit of the Bolus Cord.accessdata.fda.gov
February 5, 201264091Class IITerminatedPlum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapHospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion aaccessdata.fda.gov
January 12, 201260869Not yet classifiedTerminatedHospira Blood Set, with 200 Micron Filter, 80 Inch, Non-DEHP; an Rx sterile medical device used to administer blood products; 48 individually wrapped sets per case; Made in Costa Rica, Hospira, Inc., Lake Forest, IL 6004Silicone applied to the piercing pin of the blood set is reducing retention force of the piercing pin while in use. This could cause the piercing pin to slip out of the administration container.accessdata.fda.gov
November 18, 201160504Not yet classifiedTerminated0.9% Sodium Chloride Injection, USP, pre-filled flush solution, for IV Flush only, sterile fluid path; 10 mL Single-Use Syringe with Male Luer Lock, 100 syringes per pack, 4 packs per case; Made in Costa Rica, Hospira, IThere is particulate matter in the fluid pathway, which has been identified as the same material as the rubber tip of the syringe plunger.accessdata.fda.gov
November 10, 201160413Not yet classifiedTerminatedHospira LifeShield Latex-Free 150 mL Burette Set, with Calibrated Burette and 2 Clave Ports, Convertible Pin, 77 inch, Soluset Microdrip with Option-Lok; an Rx sterile medical device used to administer fluids; 20 individThe float valve in the burette sticks to the burette wall and does not open or close properly.accessdata.fda.gov
September 16, 201159956Not yet classifiedTerminatedPlum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971 b) list 11973 - software version 10.3 c) list 12391 - software version 11Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could cause a delay in therapy if the clinician is unable to start the device due to the continuous recycling and/accessdata.fda.gov
September 16, 201159953Not yet classifiedTerminatedPlum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971 b) list 11973 - software version 10.3 c) list 12391 - software version 11Hospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged the clamp prior to opening the cassette door, unrestricted flow may occur.accessdata.fda.gov
August 1, 201159470Not yet classifiedTerminatedGemStar Li-Ion Docking Station; a lithium ion battery powered docking station for use with GemStar Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; List Numbers 13075-04-01, 13075-04-03, 13075-04-05, 13075-04-07, 13Hospira has received reports of sparking, smoking, charring and electrical shock when using the GemStar Docking Station due to fluid ingress into the docking station.accessdata.fda.gov
March 28, 201158276Not yet classifiedTerminatedHospira Latex-Free Blood Secondary Set, 26 Inch with Secure Lock; a sterile Rx tubing set for the administration of fluids from a blood bag in series with an I.V. solution bottle and I.V. set; Made in Costa Rica; HospiraThere is a potential for the air filter at the distal end of the set to become detached. This could result in the contamination of the sterile fluid path inside the tubing.accessdata.fda.gov
March 23, 201158202Not yet classifiedTerminatedGemStar SP Infusion Suite Database, Version 1.0; an Rx PC software application; Hospira, Inc., Lake Forest, IL 60045; list 13092-04-01 A stand-alone PC application for administration of medications specific to the GemStaThe software enables programming of the GemStar device for a Bolus Lower Lockout Limit outside of the hard limits entered into the GemStar SP Infusion Suite Database PC application.accessdata.fda.gov
March 16, 201164005Class IITerminatedSymbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16026. Note that Hospira retired this pump June 30, 2015. The Hospira Symbiq Infusion System is intended for the May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay oaccessdata.fda.gov
February 14, 201157724Not yet classifiedTerminatedPlum A+ Single Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 11971 b) list 11973 - software version 10.3 c) list 12391 - software version 11Hospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.accessdata.fda.gov
February 14, 201157697Not yet classifiedTerminatedHospira Plum XL Infusion Pump, Single Channel; list number 11555Hospira has received reports of the Plum XL, XLM and XLD Infusion Pumps with no audible alarm at the low audio level setting.accessdata.fda.gov
August 12, 201056505Not yet classifiedTerminatedGemstar Pump Set, Latex-Free Extension Set with Integral Pressure-Activated Anti-Siphon Valve and Y-Extension with Backcheck Valve-SL, Nonvented, 110 inch; Rx, sterile, 24 individually packaged sets per case; Hospira, InUnderdelivery of infusion during clinical use at low rate settings (less than 10 mL/hour). Patient may not receive the intended amount of medication.accessdata.fda.gov
July 26, 201056730Not yet classifiedTerminatedEndoTool Drug Dose Calculator-Software Version 7.2.1800.3 and 7.2.1825.5The EndoTool software did not always provide an audible alert when a blood glucose measurement was due. This could result in a patient not receiving an appropriate dose of insulin.accessdata.fda.gov
January 8, 201054195Not yet classifiedTerminatedLifeShield Latex-Free Primary SYMBIQ Set, Piggyback with Backcheck Valve, 3 Prepierced Y-Sites, 106 inch, Non-DEHP, 15 drops/mL; an Rx sterile single use fluid path way for administration of I.V. fluids using the Symbiq There is a potential for fluid to leak from the backcheck valve which could result in inaccurate delivery and occupational exposure.accessdata.fda.gov
November 24, 200954080Not yet classifiedTerminatedLatex-Free Symbiq Pump Set, convertible pin, 106 inch piggyback with backcheck valve, 2 CLAVE ports, distal microbore tubing and OPTION-LOK, MICRODRIP, non-DEHP 60 drops/mL; a sterile single patient use Rx device; HospirCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.accessdata.fda.gov
November 24, 200954079Not yet classifiedTerminatedLifeshield, latex-free 100mL Burette Set, convertible pin, 77 inch with 2 pre-pierced injection sites and OPTION-LOK MICRODRIP SOLUSET, 60 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.accessdata.fda.gov
January 3, 200846846Not yet classifiedTerminatedHospira LifeCare PCA Label Utility, Version 1.0 CD. The installation/user guide is entitled "LifeCare PCA Barcode Label Utility Program." This software program is designed to allow entry of each drug concentration into aIncorrect dosage labels created: Customers have experienced independent changes in drug concentration (mg/mL to mcg/mL and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.accessdata.fda.gov
December 20, 200637014Not yet classifiedTerminatedLifeShield Latex-Free Primary I.V. Pump Set with Distal Microbore Patient Line and Gravity Flow Prevention Valve, Convertible Pin, 110 Inch with High-Pressure Filter, Orange Polyethylene-Lined/Light Resistant Tubing and The tubing can separate from the set at the filter inlet post.accessdata.fda.gov
January 9, 200634354Not yet classifiedTerminatedHospira Latex Free Blood Secondary Set, 26 inch with Secure Lock; Made in Costa Rica, Hospira, Inc., Lake Forest, IL 60045 U.S.A.The I.V. administrations sets were stolen in-transit between the manufacturing site and the sterilizer, resulting in non-sterilized product possibly being on the market.accessdata.fda.gov
September 29, 200533659Not yet classifiedTerminatedLifeShield Latex-Free HEMA Blood PlumSet; Nonvented, 100 inch, with 210 Micron Filter, Prepierced Injection Site and Option-Lok, Dual Channel Capped Secondary Port; Hospira, Inc., Lake Forest, IL 60045 USA, Made in CostaInlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.accessdata.fda.gov
May 6, 200531990Not yet classifiedTerminatedLifeShield Latex-Free HEMA Blood PlumSet, Nonvented, 110 Inch with 210 Micron Filter, Prepierced Injection Site and OPTION-LOK, Dual Channel, Capped Secondary Port; a sterile Rx blood administration set for use with the Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.accessdata.fda.gov
November 2, 200430478Not yet classifiedTerminatedAngiographic Syringe, 10 cc Thumb Ring with Rotator Reservoir.Syringes could possibly allow air to enter the patient's vascular system.accessdata.fda.gov

Compare with other makers

Updated September 17, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.