Gentherm Medical, LLC is registered with the FDA (registration 1516825) in Cincinnati, OH, United States. It lists 3 device types. The FDA files 6 510(k) clearances under its listings, the newest dated October 24, 2023. 28 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
1516825
Address
12011 Mosteller RD, Cincinnati, OH, United States, 45241
Owner or operator
Gentherm Medical LLC (1516825)
Roles
Manufacturer, Manufacturer for export, Specification developer
First device listed
December 13, 2010
Registered through
2026
Brand names listed
400CE, 414CE, FilteredFlo Warming Blankets FF-145, FilteredFlo Warming Blankets FF-243, FilteredFlo Warming Blankets FF-244, FilteredFlo Warming Blankets FF-246, FilteredFlo Warming Blankets FF-247, FilteredFlo Warming Blankets FF-248 and more

Devices it lists

DeviceProduct codeClassRoleSource
Controller, Temperature, Cardiopulmonary BypassDWC2Manufacturer, Manufacturer for exportaccessdata.fda.gov
Pack, Hot Or Cold, Water CirculatingILO2Manufactureraccessdata.fda.gov
System, Thermal RegulatingDWJ2Manufacturer, Specification developeraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
October 24, 2023K232638FilteredFlo Warming BlanketsDWJSubstantially EquivalentIob Medical, Inc.accessdata.fda.gov
June 30, 2023K231596IOB Temperature Management SystemDWJSubstantially EquivalentIob Medical, Inc.accessdata.fda.gov
January 17, 2013K123946WARMAIR MODEL 135 HYPERTHERMIA SYSTEM INCLUDING FILTEREDFLO BLANKETSDWJSubstantially EquivalentCincinnati Sub-Zero Products, Inc.accessdata.fda.gov
December 18, 2012K122813HEMOTHERM MODEL 400CE COOLER/HEATER SYSTEMDWCSubstantially EquivalentCincinnati Sub-Zero Products, Inc.accessdata.fda.gov
February 17, 2012K120081NORM-O-TEMP HYPERTHERMIA SYSTEM INCLIDING BLANKETS/PADSDWJSubstantially EquivalentCincinnati Sub-Zero Products, Inc.accessdata.fda.gov
September 20, 2010K101589BLANKETROL III, MODEL 233 AND COOLBLUE, MODEL 25-01 HYPER-HYPOTHERMIA SYSTEMSDWJSubstantially EquivalentCincinnati Sub-Zero Products, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
July 13, 202699478Class IIOngoingBlanketrol II Model 222S; Thermal regulating system.Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down.accessdata.fda.gov
November 22, 202189193Class IITerminatedElectri-Cool II Portable Cold Therapy Unit, Model Number 767, Part Number 86193This recall has been initiated to provide a labeling update with revised instruction how to operate the unit with Cold Therapy pads.accessdata.fda.gov
June 17, 202188187Class IITerminatedHemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used to cool or warm blood during cardiopulmonary bypass proceduresThere is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgaccessdata.fda.gov
July 19, 201983387Class IIITerminatedMICRO-TEMP LT System, Model 749- Localized Heat Therapy System is intended to warm a patient's body part through conductive heat transfer.Updated Manual-Warnigs have been added stating that exceeding 40'C for extended periods may cause tissue damageaccessdata.fda.gov
July 2, 201983464Class IITerminatedBlanketrol III Hyper-Hypothermia System- body hypothermia system temperature management Model 233 - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modesaccessdata.fda.gov
June 20, 201983376Class IITerminatedNorm-O-Temp model 111ZThe manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.accessdata.fda.gov
December 19, 201881961Class IITerminatedHemotherm CE Dual Reservoir Cooler/Heater, Model No. 400CE Product Usage: The HEMOTHERM Model 400CE Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood OEvidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating a host spot that led to discoloration on the board due to overheating and shut down of the device.accessdata.fda.gov
September 4, 201880941Class IITerminatedHemotherm 400CE, Catalog Numbers 86022 & 86023 The Hemotherm Dual Reservoir Cooler/Heater is used to lower, maintain, or raise the temperature of the water flowing through a Blood Oxygenator/Heat Exchanger that is used tPotential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended change to control settings.accessdata.fda.gov
November 27, 201778173Class IITerminatedHemotherm Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulating device that supplies temperature controlled water to a heat exchanger. The device, in conjunctionDue to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.accessdata.fda.gov
March 9, 201776718Class IITerminatedHemotherm CE, model 400CE, Heater-Cooling DeviceImproperly rated fuses were used in the manufacture of the device.accessdata.fda.gov
November 7, 201675641Class IITerminatedThe Hemotherm 400CE 115v, containing power board p/n 39945The device may not maintain water flow or temperature control.accessdata.fda.gov
January 7, 201570207Class IITerminatedCold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.accessdata.fda.gov
January 8, 201468356Class IITerminated115 V Blanketrol II, Model 222S Hyper-Hypothermia System. Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual accessdata.fda.gov
September 3, 201366155Class IITerminatedBlanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patientDuring a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.accessdata.fda.gov
May 15, 201366632Class IITerminatedBlanketrol III Data Export Software Version 1.10 The Blanketrol III Data Export Software is intended for use in the electronic transfer of medical device data from the Blanketrol III to a computer, without controlling orDuring a customer inquiry regarding a beta version of the Bill Data Export Software, Model USB-127, it was discovered that the wrong software version was released to a customer.accessdata.fda.gov
February 3, 201261754Not yet classifiedTerminatedECMO Heaters Model 333W, 115V - Part #86135, and 230V & 240V - Part #86136, Cardiopulmonary by-pass temperature controllers. The ECMO Heaters are used in conjunction with a heat exchanger to warm the fluid temperature. TDesign error. The probe jack contacts (terminals on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with one another. If this situation occurs, a Service Technician, Operator/User, and/or Patient could be exposed to electric shock during device saccessdata.fda.gov
October 27, 201160721Not yet classifiedTerminatedBlanketrol II 48C Hyper-Hypothermia System, Model Numbers: 222 and Model Number: 222R. The Blanketrol II 48C System provides a flow of temperature-controlled water to a heat exchanger. The device, in conjunction with a hThe suspect hyper-hypothermia device systems sold to physicians for research and/or the treatment of AIDS or Cancer patients, were not appropriately cleared by the FDA prior to distribution.accessdata.fda.gov
September 26, 201160307Not yet classifiedTerminatedCincinnati Sub Zero, Steri-Probe Temperature Probe, Esophageal/Rectal Temperature Probe, Part Number: 39622, Catalog #491B. TA disposable temperature probe used in conjunction with a measuring/monitoring device to continReports that the Rectal/ Esophageal Temperature Probes are not reading correctly. This situation causes the Blanketrol device, which the probes connect to, to shut down operation and display a 'check probe' error.accessdata.fda.gov
July 20, 201159691Not yet classifiedTerminatedOperations and Technical Manuals for the firm's WarmAir Model 135 Hyperthermia System (115V, 100V, and 230/240V), 510 (k) #K101148, Serial Numbers: 994-1350001 through 104-1356894. The WarmAir 135 Hyperthermia System is Cincinnati Sub-Zero is conducting a field correction due to changes that have been made to the WarmAir Model 135 devices Operation and Technical Manuals in order to comply with FDA device labeling regulations.accessdata.fda.gov
June 1, 201159023Not yet classifiedTerminatedWarmAir Model 135 Convective Warming Device, Cincinnati Sub-Zero Products, Inc. Used to prevent hypothermia and/or reduce cold discomfort before, during, and after surgical procedures through convective heat transfer froDue to its design the circuit breaker on the device could trip causing it to shut down and not provide heated air to the blanket.accessdata.fda.gov
May 19, 201159018Not yet classifiedTerminatedModel 134, WarmAir Convective Warming Device, 115V, Part #86180, Device Listing #D111865, The WarmAir 134 is used in conjunction with a single-patient use blanket to warm a patient's body temperature. The device draws inA component defect. The recalling firm determined that one of the connectors attached to the firm's WarmAir 134 device's Printed Circuit Board (PCB) can overheat. Should this event occur, the excess heat build up within the device may result in damage to the device's external case and/or melting of some of the internal components of the unit.accessdata.fda.gov
May 2, 201158971Not yet classifiedTerminatedECMO Heater Model 333, Cardiopulmonary bypass temperature controller. This device is intended to provide temperature-controlled warm water flow to a blood heat exchanger in order to maintain a patient's blood temperatureThe probe jack contacts (terminals) on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with each other. If this situation occurs, the service technician, user, and/or patient could be exposed to electric shock during servicing and/or use.accessdata.fda.gov
October 18, 201057004Not yet classifiedTerminated115V Blanketrol III Model 233 Hyper/Hypothermia System, Part number: 86107; and 115V CoolBlue Model 25-01 Hyper/Hypothermia System, Part number: 86000. Used to lower or to raise a patient's temperature and/or maintain a The firm initiated this recall due to changes that have been made to the Blanketrol III Model 233 device's Operation and Operation/Technical Service Manuals, in order to stay in compliance with labeling regulations.accessdata.fda.gov
November 14, 200953947Not yet classifiedTerminatedFilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for patient warming).Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be assured.accessdata.fda.gov
February 16, 200951228Not yet classifiedTerminatedWarmAir 135 Thermal Regulating System. Catalog Numbers: 86186-115 V, 06186-115 V (sold in the U.S.) and #86185-100 V ( not sold in the U.S.). Intended for warming patients before, during and after surgery, and other specThe blower on the WarmAir 135 Warming Unit may overheat. Resistance in a connector between circuit board and wiring harness results in excessive heat that may melt connector or deform the outer casing of the blower unit.accessdata.fda.gov
January 18, 200530829Not yet classifiedTerminatedMicro-temp II heat therapy unit, model #747.There is the potential that the power switch assembly can dislodge from the cabinet during use.accessdata.fda.gov
January 18, 200530820Not yet classifiedTerminatedElectri-Cool IIThere is the potential that the power switch assembly can dislodge from the cabinet during use.accessdata.fda.gov
November 5, 200430414Not yet classifiedTerminatedWarm Air Model 135 Warming Units.Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device.accessdata.fda.gov

Compare with other makers

Updated August 14, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.