Skip to the page

Data / Medical devices / United States / Fresenius Medical Care Renal Therapies Group, LLC

Fresenius Medical Care Renal Therapies Group, LLC: FDA-registered device maker in Waltham, MA, United States, 59 510(k) clearances

Fresenius Medical Care Renal Therapies Group, LLC is registered with the FDA (registration 1225714) in Waltham, MA, United States. It lists 13 device types. The FDA files 59 510(k) clearances under its listings, the newest dated January 29, 2026. 95 recall events name it as the recalling firm (5 Class I). Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
1225714
FEI number
3001451489
Address
920 Winter Street, Waltham, MA, United States, 02451
Owner or operator
Fresenius Medical Care Renal Therapies Group, LLC (9030510)
Roles
Complaint file establishment, Specification developer
First device listed
January 8, 2008
Registered through
2026
Brand names listed
2008K Hemodialysis Machine, 2008K2 Hemodialysis Machine, 2008K@HOME HEMODIALYSIS MACHINE, 2008K@Home HD Machine w/WWD (WETALERT), 2008K@home Hemodialysis Machine, 2008T BlueStar Hemodialysis Machine, 2008T Hemodialysis Machine, 4 LEAD ECO PREFILLED W/SNAP DISCONNECT and more

Devices it lists

DeviceProduct codeClassRoleSource
Accessories, Blood Circuit, HemodialysisKOC2Complaint file establishment, Specification developeraccessdata.fda.gov
Autonomous Extracorporeal Blood Leak Detector/AlarmODX2Complaint file establishment, Specification developeraccessdata.fda.gov
Dialysate Concentrate For Hemodialysis (Liquid Or Powder)KPO2Complaint file establishment, Specification developeraccessdata.fda.gov
Dialyzer, Capillary, Hollow FiberFJI2Complaint file establishment, Specification developeraccessdata.fda.gov
Dialyzer, High Permeability With Or Without Sealed Dialysate SystemKDI2Complaint file establishment, Specification developeraccessdata.fda.gov
Disinfectant, Dialysate Delivery SystemNII2Complaint file establishment, Specification developeraccessdata.fda.gov
Hemodialysis System For Home UseONW2Complaint file establishment, Specification developeraccessdata.fda.gov
Protector, Transducer, DialysisFIB2Complaint file establishment, Specification developeraccessdata.fda.gov
Set, Administration, For Peritoneal Dialysis, DisposableKDJ2Complaint file establishment, Specification developeraccessdata.fda.gov
Set, Tubing, Blood, With And Without Anti-Regurgitation ValveFJK2Complaint file establishment, Specification developeraccessdata.fda.gov
Subsystem, Water PurificationFIP2Complaint file establishment, Specification developeraccessdata.fda.gov
System, Dialysate Delivery, Semi-Automatic, PeritonealKPF2Complaint file establishment, Specification developeraccessdata.fda.gov
System, Peritoneal, Automatic DeliveryFKX2Complaint file establishment, Specification developeraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
January 29, 2026K253462NaturaLyte® Dry Bicarbonate Concentrate (08-4112-2)KPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
November 14, 2025K250471AquaC UNO HFIPSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
September 9, 2025K252181AquaAFIPSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
September 9, 2025K252180Bicarby Dialysate RFP-404 (RFP-404-W); Bicarby Dialysate RFP-403 (RFP-403-W); Bicarby Dialysate RFP-403 (RFP-403-G); Bicarby Dialysate RFP-453 (RFP-453-W); Bicarby Dialysate RFP-453 (RFP-453-G); Bicarby Dialysate RFP-454 (RFP-454-W); Bicarby Dialysate RFP-454 (RFP-454-G); Bicarby Dialysate RFP-456 (RFP-456-W)KPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
June 13, 2025K2432372008T HD SYS. CDX BLUESTAR (191124); 2008T HD SYS. CDX W/bibag BLUESTAR (191126); 2008T HD SYS. W/O CDX BLUESTAR (191128); 2008T HD SYS. W/O CDX W/bibag BLUESTAR (191130)KDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 30, 2025K2435055008X Hemodialysis SystemKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
April 4, 2025K243786Bicarby Dialysate RFP-402 (RFP-402-G); Bicarby Dialysate RFP-400 (RFP-400-G); Bicarby Dialysate RFP-407 (RFP-407-G); Bicarby Dialysate RFP-401 (RFP-401-G); Bicarby Dialysate RFP-404 (RFP-404-G); Bicarby Dialysate RFP-456 (RFP-456-G); Ci-Ca Dialysate 2K (RFP-457-G); Ci-Ca Dialysate 4K (RFP-458-G)KPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
April 3, 2025K250404stay•safe® catheter extension set with Safe- Lock, 12 inch; stay•safe® catheter extension set with Luer-Lock, 6 inch; stay•safe® catheter extension set with Luer-Lock, 12 inch; stay•safe® catheter extension set with Luer-Lock, 18 inch; stay•safe® to Luer-Lock Adapter, 4 inchKDJSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
June 13, 2024K232953AquaBplus; AquaB LITEFIPSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
December 6, 2023K2311252008T BlueStar Hemodialysis MachineKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 31, 2023K213992CombiSet Hemodialysis Blood Tubing SetFJKSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
October 31, 2022K222318Fresenius Liberty Select CyclerFKXSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
October 27, 2022K2229522008T BlueStar Hemodailysis MachineKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
July 27, 2022K213507AquaAFIPSubstantially EquivalentFresenius Medical Care Rental Therapies Group, LLCaccessdata.fda.gov
May 12, 2022K212620Citrasate Liquid Acid Concentrate, NaturaLyte Liquid Acid ConcentrateKPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
September 21, 2020K192707Optiflux F180NR DialyzerKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
September 18, 2020K201207CombiSet SMARTECH Hemodialysis Blood Tubing Set with attached Priming Set and Integrated Crit-Line Technology, CombiSet SMARTECH Hemodialysis Blood Tubing Set with attached Priming Set and Intergrated Crit-Line Technology, no Heparin LineKOCSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 8, 2020K192209Citrasate Dry Acid Concentrate, Granuflo Dry Acid ConcentrateKPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
April 24, 2020K192017Citrasate Liquid Acid Concentrate, NaturaLyte Liquid Acid ConcentrateKPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
February 24, 2020K191474NaturaLyte Dry Bicarbonate Concentrate (Rx-10 Bag), NaturaLyte Dry Bicarbonate Concentrate (Rx-12 Bag), NaturaLyte Dry Bicarbonate Concentrate (Carton)KPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
August 23, 2019K190459Hemoflow F3 and F4 DialyzersFJISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 24, 2018K181108Fresenius Liberty Select CyclerFKXSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 21, 2018K173593stay•safe catheter extension set with Safe-Lock, 12 inch, stay•safe catheter extension set with Luer-Lock, 6 inch, stay•safe catheter extension set with Luer-Lock, 12 inch, stay•safe catheter extension set with Luer-Lock, 18 inch, stay•safe to Luer-Lock adapter, 4 inchKDJSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 10, 2018K1739722008T BlueStar Hemodialysis MachineKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 1, 2018K173718Liberty Cycler SetFKXSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
February 14, 2018K173651Multiple Tubing Segment Set with stay safe PIN Connectors, stay safe Drain SetKDJSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
January 24, 2018K173363Cycler Drain Bag SetKDJSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
September 15, 2017K171652Fresenius Liberty Select CyclerFKXSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
May 18, 2017K162716bibagKPOSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
April 6, 2017K162488Optiflux dialyzerKDISubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
April 26, 2016K1534492008K2 Hemodialysis MachineKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
December 10, 2015K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineKDISubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
November 5, 2015K152953CRIT-LINE CLIP (CLIC) BLOOD CHAMBERKOCSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
October 19, 2015K143617AquaC UNO HFIPSubstantially EquivalentVivonic GmbHaccessdata.fda.gov
April 1, 2015K141997CRIT-LINE IV SYSTEMKOCSubstantially EquivalentFresenius Medical Careaccessdata.fda.gov
July 30, 2014K141145LIBERTY PDX CYCLERFKXSubstantially EquivalentFresenius Medical Care Renal Therapies Group, LLCaccessdata.fda.gov
July 7, 2014K133299ACETIC ACID WAND, ACID WAND ASSEMBLY, BICARBONATE WAND ASSEMBLYKDISubstantially EquivalentFresenius Medical Care, North America - Renal Theraccessdata.fda.gov
April 22, 2014K133829AQUABPLUS,AQUABPLUS & AQUABPLUS B2,AQUABPLUS&AQUABPLUS HF,AQUABPLUS& AQUABPLUS B2& AQUABLUS HFFIPSubstantially EquivalentVivonic GmbHaccessdata.fda.gov
September 19, 2013K124024APD LUER-LOCK ADAPTERKDJSubstantially EquivalentFresensius Medical Care North Americaaccessdata.fda.gov
September 9, 2013K123630FRESENIUS LIBERTY CYCLERFKXSubstantially EquivalentFresenius Medical Care North America, Design Centeaccessdata.fda.gov
August 29, 2013K131611FRESENIUS DRY ACID DISSOLUTION UNITSKPOSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
January 17, 2013K121421FRESENIUS 2008K HOME HEMODIALYSIS MACHINE WITH WIRELESS WETNESS DETECTOR (WETALERT)ONWSubstantially EquivalentFresenius Medical Care North America, Design Centeaccessdata.fda.gov
June 15, 2012K120823BVM HEMODIALYSIS BLOOD TUBING SET WITH ATTACHED PRIMING SET AND TRANSDUCER PROTECTORSFJKSubstantially EquivalentFresenius Medical Care North America, Design Centeaccessdata.fda.gov
February 3, 2011K070049FRESENIUS 2008K @HOME WITH WIRELESS WETNESS DETECTORONWSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
December 29, 2008K083542FRESENIUS STERILE STAY SAFE CAPKDJSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
March 31, 2005K043363FRESENIUS LIBERTY CYCLERFKXSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
October 30, 2002K022536\RESENIUS ACCESS FLOW RESERVING CONNECTORS, FRESENIUS STANDALONE ACCCES FLOW RESERVING CONNECTORKOCSubstantially EquivalentFresenius Medical Careaccessdata.fda.gov
April 27, 2001K010268FRESENIUS PRIMING SET WITH NEEDLES ACCESS PORT, MODEL 04-9002-9KOCSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
December 1, 2000K003445FRESENIUS VIRAL BLOCKING TRANSDUCER PROTECTOR FOR HEMODIALYSIS, MODEL 04-9500-2FIBSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
July 28, 2000K001873HEMOSAFE PATIENT CONNECTOR FOR FRESENIUS BLOODLINES, MODEL 04-9100-0FJKSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
June 23, 2000K001107FRESENIUS COMBILINES LOW VOLUME BLOOD TUBING SET, MODELS 03-2291 (POST PUMP) AND 03-2292 (PRE-PUMP)FJKSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
May 9, 2000K000451FRESENIUS COMBILLINES SINGLE NEEDLE BLOOD TUBING SET, CATALOG #03-2290FJKSubstantially EquivalentFresenius Medical Care North Americaaccessdata.fda.gov
November 1, 1996K962081COMBILINES HEMODIALYSIS BLOOD TUBING SETFJKSubstantially EquivalentNational Medical Care, Medical Products Div., Inc.accessdata.fda.gov
July 31, 1996K955762PACKPD PERITONEAL DIALYSIS UREA KINETIC MODELING SOFTWARE PROGRAMKPFSubstantially EquivalentFresenius USA, Inc.accessdata.fda.gov
July 17, 1991K911459GRANULYTE POWDER DIALYSATE CONCENTRATE AND MIXINGKPOSubstantially EquivalentFresenius USA, Inc.accessdata.fda.gov
February 26, 1990K896764UNIVERSAL CONNECTOR CATALOG NO. 48-1025-5KDJSubstantially EquivalentNational Medical Care, Medical Products Div., Inc.accessdata.fda.gov
October 13, 1982K822550SAFE LOCK TITANIUM ADAPTER SETFKXSubstantially EquivalentAmerican Medical Products, Inc.accessdata.fda.gov
October 13, 1982K822546SAFE LOCK SEALING CAPSFKXSubstantially EquivalentAmerican Medical Products, Inc.accessdata.fda.gov
August 31, 1981K811986THE AMP 80/2 PEDIATRIC CYCLER SYSTEMFKXSubstantially EquivalentAmerican Medical Products, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
August 10, 202699556Class IIOngoing5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.accessdata.fda.gov
July 29, 202699700Class IIOngoingCombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.accessdata.fda.gov
July 14, 202699543Class IIOngoingGRANUFLO Dry Acid Concentrate: Product Numbers: 0FD2251-3B; 0FD3251-3B; Dry Acid Concentrate for Bicarbonate HemodialysisComplaints of blue-green discoloration in productaccessdata.fda.gov
June 30, 202699438Class IIOngoingFresenius Medical Care, 2008T BlueStar Hemodialysis Machine, Model Numbers: 191126, 191130.Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) has received complaints regarding flow-, no water-, or temperature-related alarms with the 2008T BlueStar Hemodialysis machines.accessdata.fda.gov
May 12, 202698877Class IIOngoingBicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-GGiven the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated accessdata.fda.gov
May 11, 202699061Class IIOngoing5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HD Pre-Flush Standard Blood Tubing Set; Model Number: 03-5110-6. 2. 5008X HD/HDF Standard Blood Tubing Set; Model NumPotential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.accessdata.fda.gov
May 7, 202698892Class IIOngoingBicarby Dialysate; Model number: RFP-400-GThere is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassiumaccessdata.fda.gov
October 25, 202597934Class IIOngoing5008X CAREsystem +CLiC +CDXSeveral software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.accessdata.fda.gov
August 29, 202597543Class IIOngoingOptiflux High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, (3) 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, (4) 0500325E OPTIFLUX FTo provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.accessdata.fda.gov
April 29, 202596827Class IIOngoing5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.accessdata.fda.gov
December 5, 202495884Class IIOngoingDialyzer Optiflux 160NRePotential for internal blood leaks due to cracked polyurethaneaccessdata.fda.gov
August 8, 202495169Class IIOngoingBlood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis Machines with Part Numbers 190573, 190713, 190766, 190858, 190895, 190908, 191124Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.accessdata.fda.gov
June 3, 202494775Class IIOngoing190904 2008K@HOME HEMODIALYSIS SYSTEM W/ BIBAG - Blood Pump Rotor spare part, Model Number: F40015481 Rev AAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.accessdata.fda.gov
May 16, 202494676Class IIOngoingUS AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.accessdata.fda.gov
April 17, 202494463Class IIOngoingBlood Pump Rotor, Model Number: F40015481 Rev AAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.accessdata.fda.gov
April 8, 202494393Class IIOngoingNovalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas exchanIssue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during use of the Novalung system.accessdata.fda.gov
January 23, 202494084Class IOngoingstay-safe /Safe Lock Catheter Ext. 12 inThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.accessdata.fda.gov
October 24, 202393579Class IIIOngoingLiberty Select Cycler, Model numbers RTLR180343 and RTLR180111Affected Liberty Select cyclers may display an incorrect time stamp on the treatment data report sent to the clinic.accessdata.fda.gov
October 10, 202393396Class IIOngoingStay Safe CapUpdating the labeling to include additional warnings and contraindications for potential side effects.accessdata.fda.gov
September 6, 202392788Class IOngoing2008T Hemodialysis SYS, with CDXPotential PCBA leaching from tubing of hemodialysis machines.accessdata.fda.gov
August 8, 202392906Class IOngoing10ML SYRINGE LUER LOCK W/OUT NEEDLE, Part Number 15-R010-0, 100 per box, 12 boxes to a caseThere is a potential for blood backfill leakage and foreign material.accessdata.fda.gov
October 4, 202290947Class IIOngoingNaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case. For dialysis.The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.accessdata.fda.gov
July 27, 202291154Class IIOngoing2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM (1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".accessdata.fda.gov
June 21, 202290757Class IIOngoingAquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-USInternal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensuredaccessdata.fda.gov
June 16, 202290527Class IIOngoingAquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-USInternal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is taccessdata.fda.gov
June 1, 202290419Class IITerminatedNaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream infection.accessdata.fda.gov
May 31, 202290438Class IIOngoingNATURALYTE Liquid Acid 2.0K 2.5Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure Part Number: 08-2251-0Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for useaccessdata.fda.gov
May 4, 202290367Class IITerminatedOptiflux 200NRe Dialyzer Finished Assy., Model no. 0500320EDialyzers were distributed after being exposed to cold temperatures which exceeded the low-temperature range in the firm's standard operating procedure.accessdata.fda.gov
December 7, 202189369Class IIOngoingNovalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted communication between flow sensor and sensor box might lead to a disabled flow measurement and air bubble detection.accessdata.fda.gov
August 3, 202190411Class IITerminatedAquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installationaccessdata.fda.gov
April 28, 202187998Class IITerminated(1) 08-4078-BB BIBAG 650 GRAMS/BAG 16 BAGS/CASE (2) 08-4080-BB BIBAG 900 grams/bag 12 bags/case For patients with acute and chronic end stage renal disease going under PD in healthcare or home setting GRANUFLO & NATURALYExposed to freezing temperatures due to refrigerated truck malfunctionsaccessdata.fda.gov
March 18, 202187659Class IITerminatedAquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysisAquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installationaccessdata.fda.gov
December 17, 202087006Class IITerminatedFresenius Medical Care NaturaLyte Liquid Acid concentrate-Liquid Acid Concentrate for Bicarbonate Hemodialyis Part Number: 08-3251-9Bottles Missing Product Labelaccessdata.fda.gov
November 23, 202086939Class IIOngoingNaturalyte Liquid Acid concentrate bottles, Product Code 08-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificiaThe deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.accessdata.fda.gov
November 7, 202086855Class IIOngoingNaturaLyte Liquid Acid concentrate bottles, Model Number 08-1251-1 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificiaThe sodium and/or calcium levels in these concentrates were below the manufacturing specifications.accessdata.fda.gov
September 12, 202086537Class IICompletedLiberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may leaccessdata.fda.gov
September 1, 202086410Class IITerminatedAquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AqIncorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.accessdata.fda.gov
March 12, 202085310Class IIOngoingFresenius 2008T Hemodialysis Machine: Part Number Description 190573 2008T HEMODIAL YSIS SYS. OLC/DIASAFE PLUS; 190713 2008T HEMODIAL YSIS SYS., with CDX; 190766 2008T HEMODIAL YSIS SYSTEM W/BIBAG; 190858 2008T HEMODIAL A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machineaccessdata.fda.gov
January 29, 202084945Class IITerminatedHoneywell Pressure Transducer- Part Number: 370506-22. It is a component used in 2008T, Common Name: Hemodialysis Delivery System; 2008T HD SYS. BLUE STAR; 2008K2, Common Name: Hemodialysis Delivery Equipment; 2008K@homeThe firm is replacing pressure transducers with an improved design due to an increase in customer complaints regarding lower long-term reliability of the part.accessdata.fda.gov
May 29, 201983164Class IITerminatedLiberty Cycler Cassettes (a disposable cassette that is intended to be used to perform peritoneal dialysis (APD) in conjunction with the Liberty cycler as follows: (1) LIBERTY Line Cycler Set with Stay-safe connectors (DAdditional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.accessdata.fda.gov
April 3, 201982729Class IIITerminated2008T BlueStar Machine Software v.2.72-Hemodialysis machine Model Number: 191126 - Product Usage: The 2008T hemodialysis machine is indicated for acute and chronic dialysis therapy in a healthcare facility.Power Logic Board Fail Message on the 2008T BlueStar Machine with software version 2.72.accessdata.fda.gov
November 5, 201881598Class IITerminatedOptiflux F160NR Capillary High Flux Dialyzer Single Use Only, Product Code 0500316EPotential for external blood leaks from the dialyzer headeraccessdata.fda.gov
October 9, 201881451Class IIITerminatedFresenius Granuflo Dry Acid Concentrate- 2K 2.0 Ca 1 Mg 16.5 GAL Catalog Number: OFD2201-3B for hemodialysisDiscolored powder has confirmed the presence of a foreign substanceaccessdata.fda.gov
July 13, 201880875Class IITerminated2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature eaccessdata.fda.gov
May 2, 201880101Class IITerminatedNaturaLyte Liquid Acid Concentrate for Bicarbonate Dialysis, 3.43 Liters, Part Number: 08-3231-1 Product Usage: This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium BicarbonatThe bottles are mislabeled with an incorrect part number.accessdata.fda.gov
January 24, 201879625Class IITerminatedLiberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chronic peritoneal dialysis.The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or death.accessdata.fda.gov
December 14, 201779478Class IITerminated2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial NumbWhen the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit thaccessdata.fda.gov
October 5, 201779239Class IITerminated2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contaiaccessdata.fda.gov
September 29, 201779236Class IITerminatedNaturaLyte Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.accessdata.fda.gov
December 21, 201676037Class IITerminatedFresenius 2008T Series Hemodialysis SystemWhen the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machineaccessdata.fda.gov
May 24, 201674241Class IITerminatedCrit Line in a Clip (CLiC). CliC P/N: CL 10041001 CLiC is a continuous real-time monitor for non-invasive hematocrit oxygen saturation and percent change in blood volume measurement during hemodialysis treatment. Used inUse of the Blood Pressure Graph located on the Crit-Line screen of the 2008T Hemodialysis Machine may result in a frozen screen, possibly causing the blood pump to stop.accessdata.fda.gov
March 10, 201673487Class IITerminatedFresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System Fresenius 2008K@home Hemodialysis Machine with bibag System 2008T HemAdded Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)accessdata.fda.gov
February 26, 201673387Class IITerminatedFresenius Hemoflow F3 Dialyzer Part Number: 0520165A Indicated for Acute and Chronic HemodialysisDialyzer header leak due to possible improper torqueaccessdata.fda.gov
January 18, 201673077Class IITerminatedFresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate DialysisOn the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the final dialysate, 3.0 mEq/L.accessdata.fda.gov
December 15, 201572926Class IITerminatedFresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).accessdata.fda.gov
October 23, 201572484Class IITerminatedFresenius 132 Gallon GranuFlo Dissolution Unit - P/N G047-80101. For Use in the Preparation of acid concentrate for hemodialysis.Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.accessdata.fda.gov
July 16, 201571724Class IIOngoingNaturalyte Liquid Bicarbonate Concentrate 6.4 liter bottle A Dialysate Concentrate for Hemodialysis (liquid) Product Code: 08-4000-LBExpansion of recall to lots manufactured in 2015 due to the potential for bacterial contamination.accessdata.fda.gov
May 20, 201571160Class IIOngoingNaturaLyte Liquid Bicarbonate Concentrate (Dialysate Concentrate for Hemodialysis (liquid), 6.4 Liter Bottle Catalog Number: 08-4000-LB This concentrate is formulated to be used with a three steam hemodialysis machine whBacterial contamination.accessdata.fda.gov
April 27, 201571093Class IITerminatedCRIT-LINE Blood Chamber- an accessory to the Crit line III- Monitors in real time blood parameters Model: CL10021021Blood chamber connection leaks during the use of the CRIT-LINE Blood Chamberaccessdata.fda.gov
February 20, 201570574Class IITerminated2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD SYSTEM WITH CDX, CANADA; 190713 2008T HEMODIALYSIS SYS., with CDX; 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG; 190858 2The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.accessdata.fda.gov
December 19, 201470288Class IITerminatedFresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen saturation and percent change in blood volume calculation during hemoPotential for misinterpretation of the graphic display of the Blood Volume ( BV) slopeaccessdata.fda.gov
December 16, 201470027Class IITerminatedASSY, POWER SUPPLY 130V 50HZ. This component is used with the following machines: 2008T Hemodialysis Machine, 2008K Hemodialysis Delivery Equipment, 2008 K2 Hemodialysis Delivery Equipment 2008K@home Hemodialysis DeliverUnqualified power assemblies shipped to customers.accessdata.fda.gov
June 23, 201469158Class IITerminatedFresenius Naturalyte Liquid Bicarbonate Concentrate Product Number: 08-4000-LB The concentrate is formulated to be used with a three steam hemodialysis machine which is calibrated for acid and bicarbonate concentrates.Product was held at temperature above the labeled recommended storage temperatureaccessdata.fda.gov
June 23, 201468585Class IITerminatedBD Vacutainer Serum Tubes (RED TOP PLAIN TUBE 6.0 ML)Part Number: BD 367815 FMC Part Number: 87-7815-4 Single use tube used to collect, separate, transport or process blood specimens for laboratory determinations for in Products were held at improper storage temperatureaccessdata.fda.gov
May 1, 201468135Class IIITerminatedNaturaLyte Liquid Bicarbonate Concentrate, 6.4 liter bottle to be used with a hemodialysis unit; Part Number: 08-4000-LB. The concentrate is formulated to be used with a three steam hemodialysis machine which is calibratContainer cap may contain foreign material- 0.60% - 0.80% Manganeseaccessdata.fda.gov
April 25, 201468042Class IITerminatedFresenius 2008T Hemodialysis Machine with Revision K Actuator Test Boards and also sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 192008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pumpaccessdata.fda.gov
April 10, 201467922Class IOngoingNaturalyte Liquid Bicarbonate Concentrate, 6.4 Liter Bottle Part Number: 08-4000-LB The concentrate is formulated to be used with a three steam hemodialysis machine which is calibrated for acid and bicarbonate concentratNaturalyte Liquid Bicarbonate maybe contaminatedaccessdata.fda.gov
February 4, 201467313Class IITerminatedOptiflux F160NRe High Flux Dialyzers; Product Number: 0500316E. Single Use acute or chronic hemanalysis treatment.Dialysate Port Leak During Priming of the Dialyzersaccessdata.fda.gov
December 17, 201366953Class IITerminatedFresenius Bicarbonate Jug Adapter to the 2008 Series Hemodialysis machine Part Number: 650103. The adapter is a socket that accepts the 2008 machine spike to draw concentrate from the container.Bicarbonate Jug Adapter distributed prior to 510(k) approvalaccessdata.fda.gov
November 12, 201366622Class IITerminatedFresenius 2008 Series Hemodialysis Machines: Models: 2008H, 2008K, 2008K2, 2008k@Home, 2008T The 2008H Hemodialysis machine is indicated for acute and chronic dialysis therapy. Hemodialysis is indicated for patients with2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setupaccessdata.fda.gov
October 14, 201366339Class IITerminatedFresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitisaccessdata.fda.gov
August 29, 201365739Class IITerminated2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and hiUpdated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood lossaccessdata.fda.gov
June 3, 201365332Class IITerminatedFresenius Liberty Cycler Set Single Conn./Ext DL Product Number: 050-87216 Automated Peritoneal Dialysis System with integrated stay safe patient connector and extended drain line.Cassette may leak during set up or during peritoneal dialysis treatmentaccessdata.fda.gov
May 24, 201365090Class IITerminatedFresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008TIncorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)accessdata.fda.gov
November 26, 201263768Class IIITerminatedFresenius GranuFlo Dry Acid Concentrate For Bicarbonate Dialysis 2.0 K 2.5 Ca 1 Mg 16.5 GAL Catalog Number: OFD2251-3B Product Usage: GranuFlo is indicated in the treatment of acute and chronic renal failure during the hGranuFlo Dry Acid Concentrate may show discolorationaccessdata.fda.gov
October 12, 201264175Class IITerminatedFresenius Liberty Cycler Set Single Conn. Ext. DL. Part Number: 050-87216. Acute and chronic peritoneal dialysis.There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis.accessdata.fda.gov
March 30, 201260865Not yet classifiedTerminatedFresenius Liberty Cycler Software, User Manuals, and Handi-Guides for acute and chronic peritoneal dialysis. Software versions earlier than 2.8. The Fresenius Liberty Cycler is intended for acute or chronic peritoneal diIncreased risk of intraperitoneal volume (IIPV), referred to as Overfillaccessdata.fda.gov
March 29, 201262108Class ITerminatedFresenius NaturaLyte Liquid Acid Concentrate Product Codes: 08-0231-4, 08-1001-0, 08-1201-8, 08-1231-3, 08-1251-1, 08-1301-4, 08-2201-5, 08-2231-2, 08-2251-0, 08-2301-3,08-2351-8,08-3201-4,08-3231-1,08-3251-9,08-3301-2,0Risk of Alkalosis with acetate containing dialysis acid concentratesaccessdata.fda.gov
March 16, 201261418Not yet classifiedTerminatedFresenius Medical Naturalyte Acid Concentrate for bicarbonate Dialysis 45x, 55 gallons (208.2 Liters) Cat. No. 13-2251-0. NaturaLyte liquid acid concentrate is formulated to be used with a three stream hemodialysis machiSodium content does not meet product requirementsaccessdata.fda.gov
March 6, 201261293Not yet classifiedTerminatedFresenius Optiflux F160NRe Hemodialyzer Catalog number: 0500316E Intended for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate.Risk of internal blood leaks affecting the hemodialyzers.accessdata.fda.gov
December 27, 201160734Not yet classifiedTerminatedFresenius Optiflux F250NRe Hemodialyzer,, intended for single use in acute and chronic hemodialysis Catalog No. 0500325E The Fresenius Optiflux F250NRe Hemodialyzers are intended for patients with acute or chronic renal Optiflux F250NRe Hemodialyzers have increased risk of an internal blood leakaccessdata.fda.gov
August 11, 201159664Not yet classifiedTerminatedCombiSet True Flow" 2008K@home Bloodlines Fresenius Medical Care. Used for transferring blood during the hemodialysis treatment from the patient to the dialyzer (artificial kidney), removing extra water and waste productBlood lines coiling tabs #1 and #5 on the arterial line are not in the correct positionsaccessdata.fda.gov
May 4, 201159483Not yet classifiedTerminatedFresenius 2008K@home The Fresenius 2008K@home is indicated for acute and chronic dialysis therapy in an acute or chronic facility. The 2008K@home is also indicated for hemodialysis in the home and must be observed by a tUpgrade software to implement new on screen graphics and instructions, the addition of a bracket on the side of the machine to temporarily hold the bloodlines during set up in order to conform to approved 510(k)accessdata.fda.gov
March 23, 201158736Not yet classifiedTerminatedNaturalyte@ Sodium Bicarbonate Liquid Concentrate, Dialysate Concentrate for Hemodialysis Liquid Part number: 08-4000-LB Formulated to be used with a three stream hemodialysis machine which is calibrated for acid and bicMislabeled: incorrect lot number printed on case labelsaccessdata.fda.gov
February 16, 201157947Not yet classifiedTerminatedFresenius Optiflux F250NRe Hemodialyzer, Single Use, Catalog Number: 0500325E, Advanced Fresenius Polysulfone. Intended for use as an artificial kidney system to the treatment of patients with renal failure, fluid overloThere is an increased risk of an internal blood leak.accessdata.fda.gov
January 19, 201157851Not yet classifiedTerminatedFresenius 2008T Hemodialysis Machine Part Number 190573, 2008T Hemodialysis System, OLCIDIASAFE PLUS Part Number 190713, 2008T Hemodialysis System with CDX Part Number 190766, 2008T Hemodialysis System with biBag The FreCursor on the dialysis screen may disappear and not respond to user inputaccessdata.fda.gov
November 19, 201057293Not yet classifiedTerminatedFresenius Medical CombiSet BVM Hemodialysis Blood Tubing Set P/N: 03-2695-9Hemodialysis Blood tubing Sets may develop kinking of the arterial line.accessdata.fda.gov
September 3, 201056675Not yet classifiedTerminatedLIBERTY Automated Peritoneal Dialysis System Line Cycler Set with stay safe connectors (Dual Patient Connector) PIN: 050-87212, manufactured for Fresenius Medical Care, Waltham MA by Erika de Reynoza, Reynosa Tampa, MexiCycle set cassettes may leak due to holes/scratches in the film, contaminating dialysate and potentially resulting in peritonitis.accessdata.fda.gov
May 14, 201055771Not yet classifiedTerminatedOptiflux F180NR Dialyzers, Catalog Number: 0500318E. Manufactured by Fresenius Medical Care North America, Ogden, UT 84404. Designed for single use acute and chronic hemodialysis.Hemodialyzer may leak at the header/end capaccessdata.fda.gov
November 20, 200954546Not yet classifiedTerminatedFresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CAPower cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.accessdata.fda.gov
November 25, 200850491Not yet classifiedTerminatedCombiSet True Flow Series, Hemodialysis Blood Tubing with Priming Set and Transducer Protectors, with Access Flow Reversing Connector (Twister) Product Code: 03-2994-6 24 sets / case CombiSet Bloodlines Including split-sPort may crack and separate resulting in potential blood exposure/blood loss for the patientaccessdata.fda.gov
December 3, 200746292Not yet classifiedTerminatedOptiflux 180NR Advanced Fresenius Polysulfone, Catalog No.: 0500318E, Fresenius Medical Care North America, Waltham, MA 02451. (Hemodialyzer)Leaking: the Dialyzer may leak at the header resulting in small amounts of blood loss.accessdata.fda.gov
March 22, 200635184Not yet classifiedTerminatedNewton IQ 4 Lead Cycler Set with stay-safe patient connectors, used for acute and chronic peritoneal dialysis. Catalog Number: 050-87028Stay safe trigger body may not connect to the extension set and result in associated peritonitisaccessdata.fda.gov
November 28, 200534165Not yet classifiedTerminatedOptiflux Series Fresenius Polysulfone F160NR dialyzer High Flux Catalog Number: 0500316NMislabeled: Inner product mislabeled as F180NR (Catalog Number 0500318N) instead of F160NRaccessdata.fda.gov
September 20, 200430007Not yet classifiedTerminatedFresenius Peritoneal Dialysis Premier Luer Lock Transfer Set 8'', Sterile with Female Safe Lock Connector Luer Lock and Slide Clamp Catalog Number: 050-30034Peritoneal Dialysis Transfer Set may have occluded pathwayaccessdata.fda.gov

Compare with other makers

Updated September 2, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.