Skip to the page

Data / Medical devices / United States / Fresenius Kabi USA, LLC

Fresenius Kabi USA, LLC: FDA-registered device maker in North Andover, MA, United States, 2 510(k) clearances

Fresenius Kabi USA, LLC is registered with the FDA (registration 3014732157) in North Andover, MA, United States. It lists 3 device types. The FDA files 2 510(k) clearances under its listings, the newest dated November 10, 2021. 22 recall events name it as the recalling firm (13 Class I). Registered through 2026. FDA data exported November 3, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3014732157
Address
50 High Street Suite 50, North Andover, MA, United States, 01845
Owner or operator
Fresenius Kabi AG (9027285)
Roles
Manufacturer, Specification developer
First device listed
May 4, 2022
Registered through
2026
Brand names listed
Ivenix Infusion System (IIS)

Devices it lists

DeviceProduct codeClassRoleSource
Infusion Safety Management SoftwarePHC2Manufacturer, Specification developeraccessdata.fda.gov
Pump, InfusionFRN2Manufacturer, Specification developeraccessdata.fda.gov
Set, Administration, IntravascularFPA2Manufacturer, Specification developeraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
November 10, 2021K213089Ivenix Infusion System, LVP Epidural Administraion Set NRFit ConnectorFPASubstantially EquivalentIvenix, Inc.accessdata.fda.gov
June 7, 2019K183311Ivenix Infusion System (IIS)FRNSubstantially EquivalentIvenix, Inc.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
May 6, 202698945Class IOngoingIvenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in uninaccessdata.fda.gov
May 6, 202698929Class IOngoingLVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.accessdata.fda.gov
February 4, 202698699Class IIOngoingThe Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.Potential for the Cassette Loading Lever to break.accessdata.fda.gov
November 21, 202598106Class IIOngoingIvenix Infusion System (IIS), LVP Software: LVP-SW-0005.Emphasizing instructions for LVP duration programming located in the IFU.accessdata.fda.gov
November 14, 202598009Class IOngoingIvenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.accessdata.fda.gov
November 3, 202597941Class IOngoingIVENIX INFUSION SYSTEM, LVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Large Volume Infusion Pump Administration Set.Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.accessdata.fda.gov
November 3, 202597939Class IIOngoingIvenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.accessdata.fda.gov
September 3, 202597554Class IIOngoingIvenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.Potential issue that can cause the device to register "phantom" touches in the lower-left corner of the touchscreen during operation.accessdata.fda.gov
June 24, 202597165Class IIOngoingLVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25). Product Code: SET-0013-25.Potential for external cassette leaksaccessdata.fda.gov
May 12, 202596880Class IOngoingIVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.accessdata.fda.gov
January 10, 202596085Class IOngoingIvenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administrationLarge Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primaryaccessdata.fda.gov
December 6, 202495891Class IOngoingIvenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.accessdata.fda.gov
August 30, 202495240Class IOngoingIvenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.accessdata.fda.gov
August 1, 202495113Class IOngoingLVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25), Product Code: SET-0013-25. for intravenous infusion.A defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).accessdata.fda.gov
March 7, 202494228Class IOngoingLVP Software of the Ivenix Infusion System (IIS), Version 5.8.0.Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.accessdata.fda.gov
November 29, 202393474Class IOngoingIvenix Infusion System (IIS), Large Volume Pump LVP-0004The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.accessdata.fda.gov
May 10, 202393968Class IICompletedLVP software of the Ivenix Infusion System (IIS); LVP-SW-0004Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds accessdata.fda.gov
March 10, 202391783Class IOngoingIvenix Infusion System (IIS), Large Volume Pump, Model No. LVP-0004Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.accessdata.fda.gov
September 16, 202293958Class IICompletedLVP software of the Ivenix Infusion System (IIS), LVP-SW-0004Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolveaccessdata.fda.gov
September 8, 202290845Class ITerminatedIvenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate inteaccessdata.fda.gov
April 22, 202292973Class IIOngoingIvenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids thStart-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration setaccessdata.fda.gov
April 22, 202290314Class IITerminatedIvenix Infusion System (IIS)Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.accessdata.fda.gov

Compare with other makers

Updated May 29, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.