Data / Medical devices / United States / Foundation Medicine, Inc.

Foundation Medicine, Inc.: FDA-registered device maker in Cambridge, MA, United States, 4 PMA approvals

Foundation Medicine, Inc. is registered with the FDA (registration 3010679023) in Cambridge, MA, United States. It lists 1 device type. The FDA files 4 PMA approvals under its listings, the newest dated November 6, 2020. 9 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3010679023
Address
150 Second St 1st Floor, Cambridge, MA, United States, 02141
Owner or operator
Foundation Medicine, Inc. (10053707)
Roles
Complaint file establishment, Manufacturer
First device listed
December 15, 2017
Registered through
2026
Brand names listed
FoundationOne CDx, FoundationOne® Liquid CDx

Devices it lists

DeviceProduct codeClassRoleSource
Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection SystemPQP3Complaint file establishment, Manufactureraccessdata.fda.gov

PMA approvals

Premarket approvals this establishment lists, with the supplements the FDA has decided since.

ApprovedPMA numberDeviceProduct codeSupplementsLatest decisionSource
November 6, 2020P200016FoundationOne Liquid CDxPQP2March 8, 2023accessdata.fda.gov
October 26, 2020P200006FoundationOne Liquid CDx (F1 Liquid CDx)PQP4March 8, 2023accessdata.fda.gov
August 26, 2020P190032FoundationOne Liquid CDx (F1 Liquid CDx)PQP27September 18, 2026accessdata.fda.gov
November 30, 2017P170019FoundationOne CDxPQP61September 16, 2026accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
June 10, 202699268Class IIOngoingHamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.Software issue that can cause an increase in spontaneous software closures.accessdata.fda.gov
March 24, 202698640Class IIOngoingProtocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995.Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.accessdata.fda.gov
March 20, 202698625Class IIOngoingSequencing Agent SEQ0067, component of FoundationOne CDx (F1CDx)Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variantsaccessdata.fda.gov
April 9, 202596621Class IIOngoingFoundationOne Companion Diagnostic (F1CDx)Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The remaining three (3) with tumor profiling result changes did not impact CDx indications.accessdata.fda.gov
June 18, 202496802Class IIOngoingFoundationOne Companion Diagnostic (F1CDx)Two (2) reports were sent to customers with an incorrect negative claim on the claims page; the device variant information was displayed correctly in the tumor profiling section of both FDA-approved test reports.accessdata.fda.gov
February 15, 202396804Class IIOngoingFoundationOne Companion Diagnostic (F1CDx)An incorrect negative claim was identified on the claims page; the device variant information was displayed correctly in the tumor profiling section of the FDA-approved test report.accessdata.fda.gov
November 8, 202297854Class IIOngoingFoundationOne CDx (F1CDx), RAL-0003 version 31.0Reports were sent to customers without the companion diagnostic (CDx) Claims Page. Claims pages were distributed via amended reports within 7 days of the issue occurrence.accessdata.fda.gov
August 8, 201983729Class IITerminatedFoundationOne CDx test reportIdentified potential false positive MSI-H on the test reports provided to the physicians.accessdata.fda.gov
October 31, 201881614Class IITerminatedFoundationOne CDx Test ResultsAn incorrect test result for a single analyte may have been reported for up to 800 physician reportsaccessdata.fda.gov

Compare with other makers

Updated August 14, 2026. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.