Ebi, LLC: FDA-registered device maker in Parsippany, NJ, United States, 4 PMA approvals

Ebi, LLC is registered with the FDA (registration 2242816) in Parsippany, NJ, United States. It lists 2 device types. The FDA files 4 PMA approvals under its listings, the newest dated April 30, 1987. 19 recall events name it as the recalling firm. Registered through 2026. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
2242816
Address
1 Gatehall Rd. Suite 303, Parsippany, NJ, United States, 07054
Owner or operator
EBI, LLC - AHP EOS Buyer Inc. (2242816)
Roles
Complaint file establishment, Specification developer
First device listed
February 27, 2008
Registered through
2026
Brand names listed
Biomet BHS Mini, Biomet EBI Bone Healing System, Biomet OrthoPak Non-invasive Bone Growth Stimulator System, Biomet Spinalpak Non-invasive Spine Fusion Stimulator System, EBI Bone Healing System, EBI Bone Healing System-Model 2100, EBI Osteogen Implantable Growth Sti, EBI SPF Implantable Spinal Fusion S and more

Devices it lists

DeviceProduct codeClassRoleSource
Stimulator, Bone Growth, Non-InvasiveLOF2Complaint file establishment, Specification developeraccessdata.fda.gov
Stimulator, Invasive Bone GrowthLOE3Complaint file establishment, Specification developeraccessdata.fda.gov

PMA approvals

Premarket approvals this establishment lists, with the supplements the FDA has decided since.

ApprovedPMA numberDeviceProduct codeSupplementsLatest decisionSource
April 30, 1987P850035SpF® Implantable Spinal Fusion StimulatorLOE51September 20, 2024accessdata.fda.gov
February 18, 1986P850022Biomet® OrthoPak® Non-invasive Bone Growth Stimulator System; Biomet® SpinalPak® Non-invasive Spine Fusion Stimulator SyLOF31February 29, 2024accessdata.fda.gov
January 25, 1980P790005OsteoGenTM Implantable Bone Growth StimulatorsLOE58November 23, 2022accessdata.fda.gov
November 6, 1979P790002Biomet(R) EBI BONE Healing SystemLOF35October 26, 2023accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
March 22, 202494291Class IIOngoingBiomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones, Part Number: 1067716Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatmentaccessdata.fda.gov
November 11, 201469981Class IITerminatedSolitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain uA design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.accessdata.fda.gov
October 8, 201469785Class IITerminatedCypher MIS Screw System-The system includes screws, various types and sizes of rods, and set screws. Various instruments are also available for use by the surgeon to facilitate implantation of the device. Non-cervical spInoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.accessdata.fda.gov
May 15, 201468526Class IITerminatedBiomet Spine, Cypher MIS Screw Inserter, Catalog number 14-501669, instrument for Spinal Screw Fixation System.The Cypher MIS Screw Inserter may exhibit an increased rate of instrument tip failure.accessdata.fda.gov
September 24, 201366654Class IITerminatedShortened Plug Driver; Low Volume Specialty Device; Qty:1; Rx Only; Non Sterile; Biomet Spine, Parsippany, NJ 07054 The Polaris Spinal Systems is a non-cervical spinal fixation device intended for immobilization and stabBiomet received a complaint on 5/20/2013 reporting that the tip of the instrument sheared off during a removal case.accessdata.fda.gov
August 22, 201366300Class IITerminatedBIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation systemPolaris Translation Screw Failures were recently reported to Biomet Spine. These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.accessdata.fda.gov
August 13, 201366098Class IITerminatedCypher MIS Screw System, Biomet Spine Parsippany, NJ 07054. The Cypher MIS Screw System is a non-cervical spinal fixation device. 1)Screw Tower - REF 14-501660 2) Rod Reduce - REF 14-501661 3) Counter Torque - REF 14-501Several complaints were reported to exhibit screw head splay in the Cypher MIS Screw System.accessdata.fda.gov
April 9, 201365043Class IITerminatedSolitaire-35 Peek Solitaire Anterior Spinal System. The Solitaire-35 is an intervertebral body fusion device.Several complaints were recorded regarding one of the three screws driving through the cage of the Solitaire 35 implant.accessdata.fda.gov
February 7, 201364452Class IITerminatedBiomet Spine Polaris 5.5 Button Lock Screw Inserter. Catalog Number 14-500178. The Polaris Spinal System is a non-cervical, pedicle screw spinal fixation device.Increased risk of failure at the driver tip during screw insertion or removal.accessdata.fda.gov
September 29, 201263653Class IITerminatedBiomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178Broken screw insertor hardness was measured and found to be below the acceptable limits.accessdata.fda.gov
September 16, 201056791Not yet classifiedTerminatedARRAY and POLARIS Spinal System Medium Cross Connector Non Sterile Biomet Spine, Parsippany, NJ. 07054One of the two connector ends of the medium cross connector may not be fully secured to the rod after tightening. This may not be readily apparent to the operating surgeon, resulting in inadequate torsional stability.accessdata.fda.gov
April 2, 200951730Not yet classifiedTerminatedTI Cann Partl Thd Recon Screw 6.00MM x 65MM; Trauma Fixation Systems, Rx Only; Biomet, 100 Interpace Parkway, Parsippany, NJ 07054 The screw devices are an internal fixation device intended to aid in the alignment and stSome devices could have been made with raw material lots that may have inclusions, and could result in decreased strength of the device.accessdata.fda.gov
March 27, 200951725Not yet classifiedTerminatedPhoenix Cannulated Recon Screw; Ti-Cann PARTL THD Recon Screw Trauma Fixation Systems; Rx only Biomet Trauma, 100 Interpace Parkway, Parsippany, NJ 077054 Intended to aid in the alignment and stabilization of fractures tCertain lots of raw material contain inclusions that could potentially reduce the strength of some devices.accessdata.fda.gov
May 7, 200848158Not yet classifiedTerminatedTrauma Fixation Systems - The Slotted Mallet is an auxiliary optional instrument available with the Phoenix Tibia Nail System. If desired, it may be used to seat the nail into the medullary canal. The Phoenix Tibial NailCustomer complaint received regarding the separation of the mallet handle during a medical procedure.accessdata.fda.gov
October 23, 200745502Not yet classifiedTerminatedOsteoStim Cervical Allograft (Lordotic Cervical Spacer) H. 7mm; W. 14mm; L. 10mm., Catalog Number: 45-5537 and Catalog Number : LGAL906, Distributed by EBI. (dba Biomet Spine, Biomet Trauma, Biomet Bracing, Biomet OsteobUnapproved Testing: Tissue supplier (LifeLink) utilized unapproved diagnostic testing on certain lots of distributed donor tissue instead of the FDA approved donor screening test. (Note: tissue samples were later re-tested by an approved method and found non-reactive)accessdata.fda.gov
September 6, 200636377Not yet classifiedTerminatedVHS Pediatric Hip Screw System Ped VHS, Lag Screw Ref # 200516 Stainless Steel, Rx onlyVHS Plate and lag screw components did not assemble correctly intraoperatively.accessdata.fda.gov
June 24, 200532813Not yet classifiedTerminatedEBI Trochanteric 11mm Long Nail.Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification.accessdata.fda.gov
March 16, 200531822Not yet classifiedTerminatedEBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055.The wire appears to be soft and the wire bent during insertion. This causes kinks in which the reamer could not pass over the bent/kinked wire.accessdata.fda.gov
June 17, 200326512Not yet classifiedTerminatedEBI VueCath Endoscopic Spinal SystemNon sterile device is labeled sterile.accessdata.fda.gov

Compare with other makers

Updated September 20, 2024. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.