Skip to the page

Data / Medical devices / United States / Cenorin, LLC

Cenorin, LLC: FDA-registered device maker in Kent, WA, United States, 2 510(k) clearances

Cenorin, LLC is registered with the FDA (registration 3004153171) in Kent, WA, United States. It lists 2 device types. The FDA files 2 510(k) clearances under its listings, the newest dated September 9, 2019. 6 recall events name it as the recalling firm. Registered through 2027. FDA data exported November 2, 2026.

SpecStream is the live web data engine for teams and agents. It reads the FDA registration and listing, 510(k), PMA, recall and enforcement files from openFDA every week, and this page is that reading, joined by the FDA's own numbers. Ask for the makers of any device type in any country and new clearances, approvals and recalls arrive in your sheet, CRM, Slack or agent as the FDA publishes them.

FDA registration number
3004153171
Address
6324 S 199th Pl Ste 107, Kent, WA, United States, 98032
Owner or operator
CENORIN (9057829)
Roles
Manufacturer
First device listed
October 8, 2004
Registered through
2027
Brand names listed
50 SERIES: MODEL 150 DRYER, 50 SERIES: MODEL 150-1 DRYER, 50 SERIES: MODEL 150-2 DRYER, 50 SERIES: MODEL 350 DRYER, 50 SERIES: MODEL 350-1 DRYER, 50 SERIES: MODEL 350-2 DRYER, Cenorin 610 Washer-Pasteurizer/High Level Disinfector, Cenorin 610HT Washer-Pasteurizer/High Level Disinfector and more

Devices it lists

DeviceProduct codeClassRoleSource
Device, Pasteurization, Hot WaterLDS2Manufactureraccessdata.fda.gov
Unit, DryingJRJ1Manufactureraccessdata.fda.gov

510(k) clearances

The 510(k) numbers this establishment lists, with the FDA decision on each. The applicant is the company that holds the clearance.

Decided510(k) numberDeviceProduct codeDecisionApplicantSource
September 9, 2019K183444Cenorin 610 Washer-Pasteurizers/High Level Disinfector and Cenorin 610HT Washer-Pasteurizers/High Level DisinfectorLDSSubstantially EquivalentCenorin, LLCaccessdata.fda.gov
April 17, 1981K810311STERI-VERSLDSSubstantially EquivalentThe Hilrad Co.accessdata.fda.gov

Recalls

Recall events where this establishment is the recalling firm, matched on its FEI.

InitiatedEventClassStatusProductReasonSource
April 9, 201880295Class IITerminatedHLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurizaThere is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.accessdata.fda.gov
June 7, 201777637Class IITerminatedThermaSure(TM), a CENORIN(TM) product, SERIES 130 MEDICAL DEVICE DRYER: (a) Model 135; (b) Model 136 Forced air dryer for medical devicesDryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.accessdata.fda.gov
April 24, 201468237Class IITerminatedHLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height:Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.accessdata.fda.gov
January 24, 201368580Class IITerminatedHLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.accessdata.fda.gov
June 7, 201055802Not yet classifiedTerminatedHLD SYSTEM 520 is a two-tank Medical Device Cleaning and High Level Disinfection Systems (Washer/Pasteurizers). Each unit has a metal serial tag that is riveted to the enclosure of the unit and it is labeled in part: "HLThe water level sensor in the Medical Device Cleaning and High Level Disinfection (Washer/Pasteurizer) is sensitive to externally induced vibration noise, and it produces a false signal that the water level in the tank is full while it is only 75% full. It could present a health risk of biological hazard because the medical devices are not being adequately caccessdata.fda.gov
December 21, 200637018Not yet classifiedTerminatedThermaSure Medical Device Drying Cabinet, series 130Potential for user to receive an electrical shock.accessdata.fda.gov

Compare with other makers

Updated September 9, 2019. Read from openFDA bulk downloads (device registration and listing, 510(k), PMA, recall and enforcement files). CC0 1.0, public domain (openFDA terms of service). The files are joined on FDA identifiers: the 510(k) and PMA numbers a registration lists, and the recalling firm's FEI for recalls. Recall class comes from the FDA enforcement report. The FDA states that openFDA data is unvalidated and is not for decisions about medical care. Free to reuse with credit to SpecStream under CC BY 4.0. Spot a wrong value? Write to [email protected] with the link to this page.